Device-device interaction between cardiac implantable electronic devices and continuous-flow left ventricular assist devices

被引:7
作者
Tanawuttiwat, Tanyanan [1 ,2 ]
Das, Mithilesh K. [1 ]
Miller, John M. [1 ]
Guglin, Maya E. [1 ]
机构
[1] Indiana Univ, Div Cardiovasc Med, Indianapolis, IN 46202 USA
[2] Indiana Univ, 1800 N Capitol Ave,E300B, Indian, IN 46202 USA
关键词
Cardiac implantable electronic device; Continuous-flow left ventricular assist device; Implantable cardioverter-defibrillator; Leadless pacemaker; Subcutaneous implantable cardioverter-defi-brillator; ELECTROMAGNETIC-INTERFERENCE; CARDIOVERTER-DEFIBRILLATOR; PATIENT; SYSTEM; LVAD;
D O I
10.1016/j.hrthm.2023.02.020
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The current design of an innovative left ventricular assist device (LVAD) makes use of magnetic levitation technology, which enables the rotors of the device to be completely suspended by magnetic force, reducing friction and blood or plasma damage. However, this electromagnetic field can result in electromagnetic interference (EMI), which can interfere with proper functioning of another car-diac implantable electronic device (CIED) in its direct proximity. Approximately 80% of patients with an LVAD have a CIED, most frequently an implantable cardioverter-defibrillator (ICD). Several device-device interactions have been reported, including EMI -induced inappropriate shocks, inability to establish telemetry connection, EMI-induced premature battery depletion, undersens-ing by the device, and other CIED malfunctions. Unfortunately, additional procedures, including generator exchange, lead adjust-ment, and system extraction, are frequently required because of these interactions. In some circumstances, the additional procedure might be preventable or avoidable with appropriate solutions. In this article, we describe how EMI from the LVAD impacts the func-tionality of the CIED and provide possible management options, including manufacturer-specific information, for the current CIEDs (eg, transvenous and leadless pacemakers, transvenous and subcu-taneous ICDs, and transvenous cardiac resynchronization therapy pacemakers and ICDs).
引用
收藏
页码:918 / 926
页数:9
相关论文
共 53 条
  • [11] Electric smog: telemetry interference between ICD and LVAD; [Elektrosmog: Telemetrie-Interferenzen zwischen ICD und LVAD]
    Duncker D.
    König T.
    Müller-Leisse J.
    Michalski R.
    Oswald H.
    Schmitto J.D.
    Bauersachs J.
    Veltmann C.
    [J]. Herzschrittmachertherapie + Elektrophysiologie, 2017, 28 (3) : 257 - 259
  • [12] Eiringhaus Jorg, 2021, HeartRhythm Case Rep, V7, P691, DOI 10.1016/j.hrcr.2021.07.006
  • [13] Electromagnetic Interference from Left Ventricular Assist Device (LVAD) Inhibiting the Pacing Function of an Implantable Cardioverter-Defibrillator (ICD) Device
    Erqou, Sebhat
    Kormos, Robert L.
    Wang, Norman C.
    McNamara, Dennis M.
    Bazaz, Raveen
    [J]. CASE REPORTS IN CARDIOLOGY, 2018, 2018
  • [14] A compact and miniaturized implantable antenna for ISM band in wireless cardiac pacemaker system
    Feng, Yang
    Li, Zhaonan
    Qi, Lin
    Shen, Wanting
    Li, Gaosheng
    [J]. SCIENTIFIC REPORTS, 2022, 12 (01)
  • [15] Left Ventricular Mechanical Assist Devices and Cardiac Device Interactions: An Observational Case Series
    Foo, David
    Walker, Bruce D.
    Kuchar, Dennis L.
    Thorburn, Charles W.
    Tay, Andre
    Hayward, Christopher S.
    Macdonald, Peter
    Keogh, Anne
    Kotlyar, Eugene
    Spratt, Philip
    Jansz, Paul
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2009, 32 (07): : 879 - 887
  • [16] Implantable Cardioverter-Defibrillator Interference After LVAD Implantation
    Galand, Vincent
    Leclercq, Christophe
    Boule, Stephane
    Mondoly, Pierre
    Ploux, Sylvain
    Defaye, Pascal
    Verdonk, Constance
    Pozzi, Matteo
    Lellouche, Nicolas
    Martins, Raphael P.
    [J]. JACC-CLINICAL ELECTROPHYSIOLOGY, 2020, 6 (10) : 1315 - 1317
  • [17] Implantable cardiac defibrillator leads dysfunction after LVAD implantation
    Galand, Vincent
    Leclercq, Christophe
    Bourenane, Hamed
    Boule, Stephane
    Vincentelli, Andre
    Maury, Philippe
    Mondoly, Pierre
    Picard, Francois
    Welte, Nicolas
    Kindo, Michel
    Cardi, Thomas
    Pasquie, Jean-Luc
    Gaudard, Philippe
    Gourraud, Jean-Baptiste
    Probst, Vincent
    Defaye, Pascal
    Boignard, Aude
    Para, Marylou
    Algalarrondo, Vincent
    Pelce, Edeline
    Gariboldi, Vlad
    Pozzi, Matteo
    Obadia, Jean-Francois
    Anselme, Frederic
    Litzler, Pierre-Yves
    Blanchart, Katrien
    Babatasi, Gerard
    Garnier, Fabien
    Bielefeld, Marie
    Hamon, David
    Lellouche, Nicolas
    Bourguignon, Thierry
    Pierre, Bertrand
    Eschalier, Romain
    D'Ostrevy, Nicolas
    Varlet, Emilie
    Marijon, Eloi
    Blangy, Hugues
    Sadoul, Nicolas
    Flecher, Erwan
    Martins, Raphael P.
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2020, 43 (11): : 1309 - 1317
  • [18] Gupta A, 2014, TEX HEART I J, V42, P140
  • [19] Electromagnetic interference from left ventricular assist devices in patients with subcutaneous implantable cardioverter-defibrillators
    Ishida, Yuji
    Payne, Joshua E.
    Field, Michael E.
    Gold, Michael R.
    [J]. JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2020, 31 (05) : 1195 - 1201
  • [20] Review of Medical Implant Communication System (MICS) band and network
    Islam, Mohd Noor
    Yuce, Mehmet R.
    [J]. ICT EXPRESS, 2016, 2 (04): : 188 - 194