Device-device interaction between cardiac implantable electronic devices and continuous-flow left ventricular assist devices

被引:7
作者
Tanawuttiwat, Tanyanan [1 ,2 ]
Das, Mithilesh K. [1 ]
Miller, John M. [1 ]
Guglin, Maya E. [1 ]
机构
[1] Indiana Univ, Div Cardiovasc Med, Indianapolis, IN 46202 USA
[2] Indiana Univ, 1800 N Capitol Ave,E300B, Indian, IN 46202 USA
关键词
Cardiac implantable electronic device; Continuous-flow left ventricular assist device; Implantable cardioverter-defibrillator; Leadless pacemaker; Subcutaneous implantable cardioverter-defi-brillator; ELECTROMAGNETIC-INTERFERENCE; CARDIOVERTER-DEFIBRILLATOR; PATIENT; SYSTEM; LVAD;
D O I
10.1016/j.hrthm.2023.02.020
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The current design of an innovative left ventricular assist device (LVAD) makes use of magnetic levitation technology, which enables the rotors of the device to be completely suspended by magnetic force, reducing friction and blood or plasma damage. However, this electromagnetic field can result in electromagnetic interference (EMI), which can interfere with proper functioning of another car-diac implantable electronic device (CIED) in its direct proximity. Approximately 80% of patients with an LVAD have a CIED, most frequently an implantable cardioverter-defibrillator (ICD). Several device-device interactions have been reported, including EMI -induced inappropriate shocks, inability to establish telemetry connection, EMI-induced premature battery depletion, undersens-ing by the device, and other CIED malfunctions. Unfortunately, additional procedures, including generator exchange, lead adjust-ment, and system extraction, are frequently required because of these interactions. In some circumstances, the additional procedure might be preventable or avoidable with appropriate solutions. In this article, we describe how EMI from the LVAD impacts the func-tionality of the CIED and provide possible management options, including manufacturer-specific information, for the current CIEDs (eg, transvenous and leadless pacemakers, transvenous and subcu-taneous ICDs, and transvenous cardiac resynchronization therapy pacemakers and ICDs).
引用
收藏
页码:918 / 926
页数:9
相关论文
共 53 条
  • [1] Multicenter Experience of Concomitant Use of Left Ventricular Assist Devices and Subcutaneous Implantable Cardioverter-Defibrillators
    Afzal, Muhammad R.
    Ahmed, Asim
    Prutkin, Jordan M.
    Saba, Samir
    Cha, Yong-Mei
    Friedman, Paul
    Knight, Bradley P.
    Kanwar, Manreet
    Lampert, Brent C.
    Weiss, Raul
    [J]. JACC-CLINICAL ELECTROPHYSIOLOGY, 2018, 4 (09) : 1261 - 1262
  • [2] Troubleshooting electromagnetic interference in a patient with centrifugal flow left ventricular assist device and subcutaneous implantable cardioverter defibrillator
    Ahmed, Asim S.
    Patel, Parin J.
    Bagga, Shiv
    Gilge, Jasen L.
    Schleeter, Thomas
    Lakhani, Baqir A.
    Ravichandran, Ashwin K.
    Donnelley, Steve
    Allavatam, Venu
    Prystowsky, Eric N.
    Padanilam, Benzy J.
    [J]. JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2018, 29 (03) : 477 - 481
  • [3] Electromagnetic interaction between an axial left ventricular assist device and an implantable cardioverter defibrillator
    Bakhtiary, Farhad
    Therapidis, Panagiotis
    Scherer, Mirela
    Dzemali, Omer
    Moritz, Anton
    Kleine, Peter
    [J]. JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2008, 136 (05) : 1380 - 1381
  • [4] Simple Method to Eliminate Interference between Ventricular Assist Devices and Cardiac Rhythm Devices
    Baran, David A.
    Martin, Tony
    Blicharz, Donna
    Carr, Candace
    Zucker, Mark J.
    McBride, Lawrence
    Camacho, Margarita
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2009, 32 (09): : 1142 - 1145
  • [5] Batul Syeda A, 2017, HeartRhythm Case Rep, V3, P78, DOI 10.1016/j.hrcr.2016.09.010
  • [6] Overcoming Electromagnetic Interference by LVADs on ICD Function by Shielding the ICD Programmer Wand and Extension Cable
    Biviano, Angelo
    Mancini, Donna
    Naka, Yoshifumi
    Arellano, Jerry
    Garan, Hasan
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2009, 32 (07): : 945 - 948
  • [7] Subcutaneous implantable cardioverter-defibrillator troubleshooting in patients with a left ventricular assist device: A case series and systematic review
    Black-Maier, Eric
    Lewis, Robert K.
    Barnett, Adam S.
    Pokorney, Sean D.
    Sun, Albert Y.
    Koontz, Jason I.
    Daubert, James P.
    Piccini, Jonathan P.
    [J]. HEART RHYTHM, 2020, 17 (09) : 1536 - 1544
  • [8] Clinical implications of left ventricular assist device implantation in patients with an implantable cardioverter-defibrillator
    Boudghene-Stambouli, Fanny
    Boule, Stephane
    Goeminne, Celine
    Botcherby, Edward
    Marquie, Christelle
    Kouakam, Claude
    Guedon-Moreau, Laurence
    Schurtz, Guillaume
    de Groote, Pascal
    Lamblin, Nicolas
    Fertin, Marie
    Robin, Emmanuel
    Brigadeau, Francois
    Klug, Didier
    Lacroix, Dominique
    Meurice, Jonathan
    Wissocque, Ludivine
    Vincentelli, Andre
    Kacet, Salem
    [J]. JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2014, 39 (02) : 177 - 184
  • [9] Electromagnetic interference induced ventricular tachycardia in left ventricular assist device
    Chatta, Payush
    Grewal, Dennis
    Ali, Saif
    Garg, Jalaj
    Mandapati, Ravi
    Contractor, Tahmeed
    Bhardwaj, Rahul
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2022, 45 (12): : 1379 - 1382
  • [10] The role of implantable cardioverter defibrillators in patients bridged to transplantation with a continuous-flow left ventricular assist device: A propensity score matched analysis
    Clerkin, Kevin J.
    Topkara, Veli K.
    Mancini, Donna M.
    Yuzefpolskaya, Melana
    Demmer, Ryan T.
    Dizon, Jose M.
    Takeda, Koji
    Takayama, Hiroo
    Naka, Yoshifumi
    Colombo, Paolo C.
    Garan, A. Reshad
    [J]. JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2017, 36 (06) : 633 - 639