Accuracy of Advanced Active Robot for Total Knee Arthroplasty: A Cadaveric Study

被引:3
作者
Lee, Han-Jun [1 ]
Park, Kwan Kyu [2 ]
Park, Yong-Beom [3 ,6 ]
Choi, Sung Wook [4 ]
Kim, Bong-Oh [5 ]
Kim, Seong Hwan [1 ]
机构
[1] Chung Ang Univ, Coll Med, Chung Ang Univ Hosp, Dept Orthoped Surg, Seoul, South Korea
[2] Yonsei Univ, Coll Med, Severance Hosp, Dept Orthoped Surg, Seoul, South Korea
[3] Chung Ang Univ, Gwangmyeong Hosp, Coll Med, Dept Orthoped Surg, Gwangmyeong Si, Gyeonggi Do, South Korea
[4] Jeju Natl Univ, Sch Med, Jeju Natl Univ Hosp, Dept Orthoped Surg, Jeju, South Korea
[5] Curexo Inc, R&D Ctr, Seoul, South Korea
[6] Chung Ang Univ, Gwangmyeong Hosp, Dept Orthoped Surg, 110 Deokan Ro, Gwangmyeong Si 14353, Gyeonggi Do, South Korea
关键词
Knee; arthroplasty; robot; accuracy; cadaver; NAVIGATION SYSTEM; CURRENT STATE;
D O I
10.1055/s-0042-1760391
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Purpose Although the accuracy of other types of robotic systems for total knee arthroplasty (TKA) has been assessed in cadaveric studies, no investigations have been performed to evaluate this newly advanced active robotic system. Therefore, the authors aimed to analyze the accuracy of bone resection in terms of thickness and alignment in a cadaveric study.Methods Three cadaveric specimens (six knees) and an active robotic system (CUVIS Joint, CUREXO) were used in the study. Three surgeons with different experiences in robotic TKAs performed this cadaveric study using the same robotic protocol with two different implant designs. The thickness and angle of bone resection planes obtained from the optical tracking system and the difference between resection planes and the planning data were assessed to determine accuracy.Results With respect to the overall resection accuracy compared to the plan, the cutting depth accuracy was within 1.0 mm mean of root mean square (RMS), and the resection angle accuracy in terms of sagittal, coronal, and axial planes was within 1.0 degree mean RMS. In contrast, no significant differences were observed between the planned and measured values in terms of the resection angles and cutting thickness. The hip-knee-ankle angle at postoperative evaluation was 0.7 degrees +/- 0.7 degrees (RMS 1.0 degrees).Conclusion This in vivo study suggests that the use of this newly advanced active robotic system for TKA demonstrates a high degree of accuracy in terms of resection thickness and alignment. This finding supports the clinical application of this advanced robotic system.
引用
收藏
页码:135 / 141
页数:7
相关论文
共 29 条
[1]   Concepts and techniques of a new robotically assisted technique for total knee arthroplasty: the ROSA knee system [J].
Batailler, Cecile ;
Hannouche, Didier ;
Benazzo, Francesco ;
Parratte, Sebastien .
ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2021, 141 (12) :2049-2058
[2]   Robotics in Total Knee Arthroplasty [J].
Bautista, Maria ;
Manrique, Jorge ;
Hozack, William J. .
JOURNAL OF KNEE SURGERY, 2019, 32 (07) :600-606
[3]   Accuracy assessment of a novel image-free handheld robot for Total Knee Arthroplasty in a cadaveric study [J].
Casper, Michael ;
Mitra, Riddhit ;
Khare, Rahul ;
Jaramaz, Branislav ;
Hamlin, Brian ;
McGinley, Brian ;
Mayman, David ;
Headrick, Jeff ;
Urish, Kenneth ;
Gittins, Mark ;
Incavo, Stephen ;
Neginhal, Vivek .
COMPUTER ASSISTED SURGERY, 2018, 23 (01) :14-20
[4]  
Clark T C., 2013, ISRN Orthop, V2013, P1, DOI DOI 10.1155/2013/794827
[5]   Patient Dissatisfaction Following Total Knee Arthroplasty: A Systematic Review of the Literature [J].
Gunaratne, Rajitha ;
Pratt, Dylan N. ;
Banda, Joseph ;
Fick, Daniel P. ;
Khan, Riaz J. K. ;
Robertson, Brett W. .
JOURNAL OF ARTHROPLASTY, 2017, 32 (12) :3854-3860
[6]   Robotic-Arm Assisted Total Knee Arthroplasty Demonstrated Greater Accuracy and Precision to Plan Compared with Manual Techniques [J].
Hampp, Emily L. ;
Chughtai, Morad ;
Scholl, Laura Y. ;
Sodhi, Nipun ;
Bhowmik-Stoker, Manoshi ;
Jacofsky, David J. ;
Mont, Michael A. .
JOURNAL OF KNEE SURGERY, 2019, 32 (03) :239-250
[7]   Robotics in orthopaedic surgery: why, what and how? [J].
Innocenti, Bernardo ;
Bori, Edoardo .
ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2021, 141 (12) :2035-2042
[8]   Which orally administered antithrombotic agent is most effective for preventing venous thromboembolism after total knee arthroplasty? A propensity score-matching analysis [J].
Jang, Seonpyo ;
Shin, Woo Cheol ;
Song, Min Ku ;
Han, Hyuk-Soo ;
Lee, Myung Chul ;
Ro, Du Hyun .
KNEE SURGERY & RELATED RESEARCH, 2021, 33 (01)
[9]   Appropriateness of the Use of Navigation System in Total Knee Arthroplasty [J].
Jung, Suk Han ;
Cho, Myung Rae ;
Song, Suk-Kyoon .
CLINICS IN ORTHOPEDIC SURGERY, 2020, 12 (03) :324-329
[10]   Robotic-arm assisted total knee arthroplasty is associated with improved early functional recovery and reduced time to hospital discharge compared with conventional jig-based total knee arthroplasty A PROSPECTIVE COHORT STUDY [J].
Kayani, B. ;
Konan, S. ;
Tahmassebi, J. ;
Pietrzak, J. R. T. ;
Haddad, F. S. .
BONE & JOINT JOURNAL, 2018, 100B (07) :930-937