A novel artificial intelligence based algorithm to reduce wearable cardioverter-defibrillator alarms

被引:1
作者
Arkles, Jeffrey [1 ]
Delaughter, Craig [2 ]
D'Souza, Benjamin [1 ,3 ]
机构
[1] Univ Penn, Cardiovasc Div, Electrophysiol Sect, Perelman Sch Med, Philadelphia, PA 19104 USA
[2] Baylor Heart & Vasc Hosp, Ft Worth, TX USA
[3] Penn Presbyterian Med Ctr, Heart & Vasc Pavil,51 N 39th St,4th floor, Philadelphia, PA 19104 USA
关键词
Cardiomyopathy; Sudden cardiac death; Defibrillation; Wearable cardioverter-defibrillator; Ventricular fibrillation; Ventricular tachycardia; Internal cardioverter-defibrillator; SUDDEN CARDIAC DEATH; HIGH-RISK; EXPERIENCE;
D O I
10.1007/s10840-023-01497-w
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BackgroundThe wearable cardioverter defibrillator (WCD) is an important tool in mitigating sudden cardiac death (SCD). The WCD provides patient alarms for detected arrhythmias or electrical noise/artifact. Some patients experience frequent alarms for artifact. We sought to evaluate the effects of a novel artificial intelligence algorithm to reduce alarms related to electrical noise or artifact (advanced arrhythmia discrimination algorithm, AArD).MethodsA retrospective review of a large commercial database of prescribed WCD. Patients prescribed the WCD during the years 2017 (discrimination algorithm, DA group) or during 2019 (advanced arrhythmia discrimination algorithm, AArD) were analyzed. A total of 96,000 patients were sampled, 48,000 in the control group (using standard direct algorithm, DA) and compared to 48,000 in the intervention AArD (4000 per group per month) for 12 months. The AArD further discriminates ECG signals based on a machine-learning algorithm utilizing intensity and frequency beyond the standard DA. Outcomes regarding alarms, arrhythmias, and safety were analyzed.ResultsThe AArD algorithm was associated with a significant decrease in frequency of alarms over the course of WCD use; 54% of patients in the AARD versus 27% of DA had 0 alarms (P < .001). In the entire cohort, there was a 56% relative reduction in alarms with the use of AArD. Appropriate arrhythmia treatment time was not significantly different between the groups (44 s DA vs 45 s AArD [p = ns]).ConclusionsA novel artificial intelligence algorithm reduces alarms without delaying appropriate therapy for VT/VF. These changes may improve compliance and quality of life in patients with a WCD.
引用
收藏
页码:1723 / 1728
页数:6
相关论文
共 10 条
[1]   Aggregate National Experience With the Wearable Cardioverter-Defibrillator Event Rates, Compliance, and Survival [J].
Chung, Mina K. ;
Szymkiewicz, Steven J. ;
Shao, Mingyuan ;
Zishiri, Edwin ;
Niebauer, Mark J. ;
Lindsay, Bruce D. ;
Tchou, Patrick J. .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2010, 56 (03) :194-203
[2]   Use of a wearable defibrillator in terminating tachyarrhythmias in patients at high risk for sudden death: Results of WEARIT/BIROAD [J].
Feldman, AM ;
Klein, H ;
Tchou, P ;
Murali, S ;
Hall, WJ ;
Mancini, D ;
Boehmer, J ;
Harvey, M ;
Hellman, MS ;
Szymkiewicz, SJ ;
Moss, AJ .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2004, 27 (01) :4-9
[3]   Wearable cardioverter-defibrillator in patients with a transient risk of sudden cardiac death: the WEARIT-France cohort study [J].
Garcia, Rodrigue ;
Combes, Nicolas ;
Defaye, Pascal ;
Narayanan, Kumar ;
Guedon-Moreau, Laurence ;
Boveda, Serge ;
Blangy, Hugues ;
Bouet, Jerome ;
Briand, Florent ;
Chevalier, Philippe ;
Cottin, Yves ;
Da Costa, Antoine ;
Degand, Bruno ;
Deharo, Jean-Claude ;
Eschalier, Romain ;
Extramiana, Fabrice ;
Goralski, Marc ;
Guy-Moyat, Benoit ;
Guyomar, Yves ;
Hermida, Jean-Sylvain ;
Jourda, Francois ;
Lellouche, Nicolas ;
Mahfoud, Mohanad ;
Manenti, Vladimir ;
Mansourati, Jacques ;
Martin, Angeline ;
Pasquie, Jean-Luc ;
Ritter, Philippe ;
Rollin, Anne ;
Tibi, Thierry ;
Yalioua, Arab ;
Gras, Daniel ;
Sadoul, Nicolas ;
Piot, Olivier ;
Leclercq, Christophe ;
Marijon, Eloi .
EUROPACE, 2021, 23 (01) :73-81
[4]   One-year follow-up of the prospective registry of patients using the wearable defibrillator (WEARIT-II Registry) [J].
Kutyifa, Valentina ;
Moss, Arthur J. ;
Klein, Helmut U. ;
McNitt, Scott ;
Zareba, Wojciech ;
Goldenberg, Ilan .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2018, 41 (10) :1307-1313
[5]   Use of the Wearable Cardioverter Defibrillator in High-Risk Cardiac Patients Data From the Prospective Registry of Patients Using the Wearable Cardioverter Defibrillator (WEARIT-II Registry) [J].
Kutyifa, Valentina ;
Moss, Arthur J. ;
Klein, Helmut ;
Biton, Yitschak ;
McNitt, Scott ;
MacKecknie, Bonnie ;
Zareba, Wojciech ;
Goldenberg, Ilan .
CIRCULATION, 2015, 132 (17) :1613-1619
[6]   Impact of wearable cardioverter-defibrillator compliance on outcomes in the VEST trial: As-treated and per-protocol analyses [J].
Olgin, Jeffrey E. ;
Lee, Byron K. ;
Vittinghoff, Eric ;
Morin, Daniel P. ;
Zweibel, Steven ;
Rashba, Eric ;
Chung, Eugene H. ;
Borggrefe, Martin ;
Hulley, Stephen ;
Lin, Feng ;
Hue, Trisha F. ;
Pletcher, Mark J. .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2020, 31 (05) :1009-1018
[7]   Wearable Cardioverter-Defibrillator after Myocardial Infarction [J].
Olgin, Jeffrey E. ;
Pletcher, Mark J. ;
Vittinghoff, Eric ;
Wranicz, Jerzy ;
Malik, Rajesh ;
Morin, Daniel P. ;
Zweibel, Steven ;
Buxton, Alfred E. ;
Elayi, Claude S. ;
Chung, Eugene H. ;
Rashba, Eric ;
Borggrefe, Martin ;
Hue, Trisha F. ;
Maguire, Carol ;
Lin, Feng ;
Simon, Joel A. ;
Hulley, Stephen ;
Lee, Byron K. .
NEW ENGLAND JOURNAL OF MEDICINE, 2018, 379 (13) :1205-1215
[8]   Incidence of out-of-hospital cardiac arrest [J].
Rea, TD ;
Pearce, RM ;
Raghunathan, TE ;
Lemaitre, RN ;
Sotoodehnia, N ;
Jouven, X ;
Siscovick, DS .
AMERICAN JOURNAL OF CARDIOLOGY, 2004, 93 (12) :1455-1460
[9]   Experience With the Wearable Cardioverter-Defibrillator in Patients at High Risk for Sudden Cardiac Death [J].
Waessnig, Nadine K. ;
Guenther, Michael ;
Quick, Silvio ;
Pfluecke, Christian ;
Rottstaedt, Fabian ;
Szymkiewicz, Steven J. ;
Ringquist, Steven ;
Strasser, Ruth H. ;
Speiser, Uwe .
CIRCULATION, 2016, 134 (09) :635-643
[10]   Sudden cardiac death in the United States, 1989 to 1998 [J].
Zheng, ZJ ;
Croft, JB ;
Giles, WH ;
Mensah, GA .
CIRCULATION, 2001, 104 (18) :2158-2163