Can magnetic resonance imaging after cranioplasty using titanium mesh detect brain tumors?

被引:1
作者
Takatsu, Yasuo [1 ,2 ]
Yoshida, Rei [3 ]
Yamatani, Yuya [4 ]
Kanno, Mikihisa [5 ]
Miyati, Tosiaki [2 ]
机构
[1] Fujita Hlth Univ, Sch Med Sci, Mol Imaging, 1-98 Dengakugakubo, Toyoake, Aichi 4701192, Japan
[2] Kanazawa Univ, Grad Sch Med Sci, Div Hlth Sci, 5-11-80 Kodatsuno,Kutsukake cho, Kanazawa 9200942, Japan
[3] Kurihara Cent Hosp, Dept Radiol, 3-3-1 Miyano chuo, Kurihara city, Miyagi 9872205, Japan
[4] Nara Med Univ Hosp, Div Cent Radiol, 840 Shijo Cho, Kashihara, Nara 6348522, Japan
[5] Minami Matsuyama Hosp, Dept Radiol Technol, 3-10 Asodamachi, Matsuyama, Ehime 7908534, Japan
关键词
Magnetic resonance imaging; Brain tumor; Titanium mesh; RF-shielding effect; Phantom; GRADIENT-ECHO; GADOLINIUM; CONTRAST; RECONSTRUCTION; DETECTABILITY; STENTS;
D O I
10.1007/s13246-022-01200-9
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This study determined the dependence of the concentration and position of contrast-enhanced tumors on the radio frequency (RF)-shielding effect of titanium mesh using the contrast-to-noise ratio (CNR) in magnetic resonance imaging (MRI). A phantom was constructed by filling a plastic container with manganese chloride tetrahydrate and agar. Four cellophane cylindrical containers were arranged from the end of the plastic container, and the brain tumor model was filled with gadobutrol diluted with NaCl, with molarity values of 0.2-1.0 mmol/L. The titanium mesh board was set on the left side of the phantom. Images were acquired using a 1.5-T MRI as well as two-dimensional spin-echo (2D SE) and three-dimensional fast spoiled gradient echo (3D FSPGR) sequences. CNR was calculated using the signal intensity values of the tumor model, surrounding area of the brain model, and background noise. Furthermore, the fractional change in CNR was calculated using values of CNR with and without the mesh. Moreover, a profile of CNR was created. The fractional change in CNR decreased at the brain tumor positions present near the mesh and at a contrast medium concentration of approximately & LE; 0.5 mmol/L in 2D SE and & LE; 0.25 mmol/L in 3D FSPGR. According to the CNR profiles, directly under the mesh, almost all contrast concentrations in 2D SE was unrecognizable; however, at a concentration of & GE; 0.5 mmol/L in 3D FSPGR was recognizable.
引用
收藏
页码:109 / 118
页数:10
相关论文
共 36 条
[1]   Improved lumen visualization in metallic vascular implants by reducing RF artifacts [J].
Bartels, LW ;
Bakker, CJG ;
Viergever, MA .
MAGNETIC RESONANCE IN MEDICINE, 2002, 47 (01) :171-180
[2]   Nephrogenic systemic fibrosis associated with gadolinium based contrast agents: A summary of the medical literature reporting [J].
Broome, Dale R. .
EUROPEAN JOURNAL OF RADIOLOGY, 2008, 66 (02) :230-234
[3]   Long-term results following titanium cranioplasty of large skull defects [J].
Cabraja, Mario ;
Klein, Martin ;
Lehmann, Thomas-Nikolas .
NEUROSURGICAL FOCUS, 2009, 26 (06) :1-7
[4]   IMAGING AFTER TITANIUM CRANIOPLASTY [J].
CHANDLER, CL ;
UTTLEY, D ;
ARCHER, DJ ;
MACVICAR, D .
BRITISH JOURNAL OF NEUROSURGERY, 1994, 8 (04) :409-414
[5]  
Chiriac A, 2013, DIG J NANOMATER BIOS, V8, P729
[6]   TOTAL INHOMOGENEITY CORRECTION INCLUDING CHEMICAL-SHIFTS AND SUSCEPTIBILITY BY VIEW ANGLE TILTING [J].
CHO, ZH ;
KIM, DJ ;
KIM, YK .
MEDICAL PHYSICS, 1988, 15 (01) :7-11
[7]   CONTRAST, RESOLUTION, AND DETECTABILITY IN MR IMAGING [J].
CONSTABLE, RT ;
HENKELMAN, RM .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1991, 15 (02) :297-303
[8]   Influence of multichannel combination, parallel imaging and other reconstruction techniques on MRI noise characteristics [J].
Dietrich, Olaf ;
Raya, Jose G. ;
Reeder, Scott B. ;
Ingrisch, Michael ;
Reiser, Maximilian F. ;
Schoenberg, Stefan O. .
MAGNETIC RESONANCE IMAGING, 2008, 26 (06) :754-762
[9]  
Elephterios B., 2010, ROMANIAN NEUROSURGER, V17, P456
[10]   Estimation of Gadolinium-based Contrast Agent Concentration Using Quantitative Synthetic MRI and Its Application to Brain Metastases: A Feasibility Study [J].
Fujita, Shohei ;
Nakazawa, Misaki ;
Hagiwara, Akifumi ;
Ueda, Ryo ;
Horita, Moeko ;
Maekawa, Tomoko ;
Irie, Ryusuke ;
Andica, Christina ;
Kumamaru, Kanako Kunishima ;
Hori, Masaaki ;
Aoki, Shigeki .