Detective quantum efficiency of a double-layered detector for dual-energy x-ray imaging

被引:0
|
作者
Shin, Hubeom [1 ]
Yoo, Seungjun [1 ]
Oh, Seokwon [1 ]
Lee, Junho [1 ]
Kyung, Ho [1 ,2 ]
机构
[1] Pusan Natl Univ, Sch Mech Engn, Busandaehak Ro 63beon Gil, Busan 46241, South Korea
[2] Pusan Natl Univ, Ctr Adv Med Engn Res, Busandaehak Ro 63beon Gil, Busan 46241, South Korea
基金
新加坡国家研究基金会;
关键词
Medical-image reconstruction methods and algorithms; computer-aided diagnosis; Inspection with x-rays; X-ray radiography and digital radiography (DR); CASCADED SYSTEMS-ANALYSIS; OPTIMIZATION;
D O I
10.1088/1748-0221/18/11/C11005
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A sandwich-like double-layered detector can perform dual-energy imaging (DEI) using a single x-ray exposure without object motion artifacts. The energy separation between measurements obtained from the front and rear-detector layers can be tuned by introducing an x-ray beam-attenuating material between them. However, the design of the interdetector filter significantly influences dose efficiency by altering the number of x-ray photons reaching the rear-detector layer within the sandwich detector. Since the sandwich detector typically incorporates phosphors of differing thicknesses for its two detector layers, it exhibits a unique spatial resolution characteristic in the reconstructed dual energy (DE) images. To comprehensively assess detector performance in terms of design (filter) and operation (reconstruction), we established a framework that describes the dual-energy detective quantum efficiency (DE-DQE) using linear-systems theory. The developed DE-DQE model was validated through comparison with measurements. The agreement between the modulation-transfer functions was reasonable, and the correspondence between noise-power spectra was excellent. This proposed DE-DQE concept is universally applicable to any linearly operating DE system and holds a significant value in enhancing the performance of sandwich detectors or ensuring their optimal operation.
引用
收藏
页数:12
相关论文
共 50 条
  • [21] Dual-energy X-ray absorptiometry
    Braillon, Pierre
    Peretti, Noel
    NUTRITION CLINIQUE ET METABOLISME, 2011, 25 (02): : 91 - 96
  • [22] Dual-energy X-ray Absorptiometry
    Jain, Rajesh K.
    Vokes, Tamara
    JOURNAL OF CLINICAL DENSITOMETRY, 2017, 20 (03) : 291 - 303
  • [23] Dual-energy X-ray diagnostics
    Coumans, J.J.H.
    Philips Technical Review, 1986, 42 (8-9): : 274 - 285
  • [24] Detective quantum efficiency of television X-ray detectors
    Kalata, Kenneth
    Stanton, Martin
    Phillips, Walter
    Journal of X-Ray Science and Technology, 1992, 3 (03)
  • [25] Linear modeling of single-shot dual-energy x-ray imaging using a sandwich detector
    Kim, J.
    Kim, D. W.
    Kim, S. H.
    Yun, S.
    Youn, H.
    Jeon, H.
    Kim, H. K.
    JOURNAL OF INSTRUMENTATION, 2017, 12
  • [26] Material discrimination by high-energy x-ray dual-energy imaging
    Wang Xue-Wu
    Li Jian-Min
    Kang Ke-Jun
    Tang Chuan-Xiang
    Mang Li
    Chen Zhi-Qiang
    Li Yuan-Jing
    Zhong Hua-Qiang
    HIGH ENERGY PHYSICS AND NUCLEAR PHYSICS-CHINESE EDITION, 2007, 31 (11): : 1076 - 1081
  • [27] Investigation of dual-energy X-ray photon counting using a cadmium telluride detector with dual-energy selection electronics
    Sato, Eiichi
    Kosuge, Yoshiyuki
    Yamanome, Hayato
    Mikata, Aldko
    Miura, Tatsuya
    Oda, Yasuyuki
    Ishii, Tomotaka
    Hagiwara, Osahiko
    Matsukiyo, Hiroshi
    Watanabe, Manabu
    Kusachi, Shinya
    RADIATION PHYSICS AND CHEMISTRY, 2017, 130 : 385 - 390
  • [28] Robust Material Classification on Dual-Energy X-Ray Imaging Devices
    Cigla, Cevahir
    Kucukates, Busra
    Yalcin, Ozan
    Ak, Duygu Selin
    Isikoglu, Sukrucan Taylan
    ANOMALY DETECTION AND IMAGING WITH X-RAYS (ADIX) VII, 2022, 12104
  • [29] Evaluation of a dichromatic X-ray source for dual-energy imaging in mammography
    Tuffanelli, A
    Fabbri, S
    Sarnelli, A
    Taibi, A
    Gambaccini, M
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2002, 489 (1-3): : 509 - 518
  • [30] Theoretical Feasibility of Dual-Energy X-Ray Imaging of Respiratory Disease
    Basharat, F.
    Tanguay, J.
    MEDICAL PHYSICS, 2020, 47 (06) : E558 - E558