Performance of a triple GEM detector equipped with Al-GEM foils for X-rays detection

被引:7
|
作者
Caruggi, Federico [1 ,2 ]
Cancelli, Stephanie [1 ,3 ]
Celora, Agostino [1 ]
Guiotto, Federico [1 ]
Croci, Gabriele [1 ,2 ,3 ]
Tardocchi, Marco [2 ,3 ]
Murtas, Fabrizio [4 ]
de Oliveira, Rui [5 ]
Cippo, Enrico Perelli [2 ]
Gorini, Giuseppe [1 ,2 ,3 ]
Grosso, Giovanni [2 ]
Muraro, Andrea [2 ]
机构
[1] Univ Milano Bicocca, Dipartimento Fis G Occhialini, Piazza Sci 3, I-20125 Milan, Italy
[2] CNR, Ist Sci & Tecnol Plasmi, Via Cozzi 53, I-20125 Milan, Italy
[3] Ist Nazl Fis Nucl, Sez Milano Bicocca, Piazza Sci 3, I-20125 Milan, Italy
[4] Ist Nazl Fis Nucl, Lab Nazl Frascati, Via Enr Fermi 40, I-00044 Frascati, Italy
[5] European Org Nucl Res CERN, Geneva, Switzerland
关键词
Micropattern Gaseous detectors; Gas electron multiplier; X-ray detectors; SYSTEM;
D O I
10.1016/j.nima.2022.167855
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The study of Soft X-ray emission can be a source of fundamental information, particularly for what concerns tokamaks and plasma diagnostics, but also in general in the fields of high energy and nuclear physics. Detection systems based on Gas Electron Multipliers (GEM) technology can be of particular use in the context of X-ray analyses, being relatively low cost while maintaining good spatial and temporal resolution and capability to sustain high counting rates (up to MHz/mm2). The development of these new and improved detectors is thus of interest, especially in the research about diagnostic and control of machines for fusion energy. In this work, the performance of a new triple-GEM detector, characterized by an aluminum metallic coating on both layers of the GEM foils, is presented and a comparison is made with the more conventional design employing copper coating. The performances of an aluminum-coated GEM (Al-GEM) detector and of a standard copper-coated GEM (Cu-GEM) detector in revealing quasi-monochromatic X-ray beams coming from different fluorescence materials are compared. The Al-GEM detector is shown to less suffer the issue of unwanted background on the signal caused by the presence of copper inside the detector itself. The suppression of this noise source encourages the use of Al-GEM detectors to perform spectroscopy in harsh environments, such as tokamak machines, where soft X-rays are useful probes to control different plasma properties and parameters.
引用
收藏
页数:8
相关论文
共 50 条
  • [1] Performance of the triple GEM detector built using commercially manufactured GEM foils in India
    Gola, Mohit
    Malhotra, Shivali
    Shah, Aashaq
    Ahmed, Asar
    Kumar, Ashok
    Naimuddin, Md
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2020, 951
  • [2] Characterization of a scintillating GEM detector with low energy x-rays
    Seravalli, E.
    de Boer, M. R.
    Geurink, F.
    Huizenga, J.
    Kreuger, R.
    Schippers, J. M.
    van Eijk, C. W. E.
    PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (21): : 6195 - 6209
  • [3] Measurement of GEM parameters with X-rays
    Bencivenni, G
    Bonivento, W
    Cardini, A
    Deplano, C
    De Simone, P
    Murtas, F
    Pinci, D
    Poli-Lener, M
    Raspino, D
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2003, 50 (05) : 1297 - 1302
  • [4] Performance of a GEM detector with copper-less foils
    Mindur, B.
    Fiutowski, T.
    Koperny, S.
    Wiacek, P.
    Dabrowski, W.
    JOURNAL OF INSTRUMENTATION, 2017, 12
  • [5] Impact of the hole orientation of asymmetric GEM foils on the performance of single and triple GEM detectors
    Bouhali, O.
    Hoepfner, K.
    Ivone, F.
    Kamon, T.
    Keller, H.
    Malhotra, S.
    Al Rashdi, B.
    JOURNAL OF INSTRUMENTATION, 2021, 16 (10)
  • [6] Stability tests performed on the triple GEM detector built using commercially manufactured GEM foils in India
    Gola, M.
    Malhotra, S.
    Kumar, A.
    Naimuddin, Md
    JOURNAL OF INSTRUMENTATION, 2019, 14 (08)
  • [7] Discriminating cosmic muons and X-rays based on rise time using a GEM detector
    吴会寅
    赵圣鹰
    王晓冬
    章先鸣
    祁辉荣
    张伟
    吴柯岩
    胡碧涛
    张毅
    Chinese Physics C, 2016, 40 (08) : 76 - 79
  • [8] Discriminating cosmic muons and X-rays based on rise time using a GEM detector
    吴会寅
    赵圣鹰
    王晓冬
    章先鸣
    祁辉荣
    张伟
    吴柯岩
    胡碧涛
    张毅
    Chinese Physics C, 2016, (08) : 76 - 79
  • [9] Results from the measurement of GEM-based detector parameters performed with X-rays
    Bencivenni, G
    Bonivento, W
    Cardini, A
    Deplano, C
    De Simone, P
    Murtas, F
    Pinci, D
    Poli-Lener, M
    Raspino, D
    2002 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORD, VOLS 1-3, 2003, : 595 - 599
  • [10] Discriminating cosmic muons and X-rays based on rise time using a GEM detector
    Wu, Hui-Yin
    Zhao, Sheng-Ying
    Wang, Xiao-Dong
    Zhang, Xian-Ming
    Qi, Hui-Rong
    Zhang, Wei
    Wu, Ke-Yan
    Hu, Bi-Tao
    Zhang, Yi
    CHINESE PHYSICS C, 2016, 40 (08)