Carotid flow as a surrogate of the left ventricular stroke volume

被引:8
作者
Cheong, Issac [1 ,2 ]
Castro, Victoria Otero [1 ]
Sosa, Fernando Ariel [2 ,3 ]
Oribe, Barbara Tort [3 ]
Merlo, Pablo Martin [2 ,4 ]
Tamagnone, Francisco Marcelo [2 ,5 ]
机构
[1] Sanatorio Los Arcos, Dept Crit Care Med, Buenos Aires, DF, Argentina
[2] Argentinian Crit Care Ultrasonog Assoc ASARUC, Buenos Aires, DF, Argentina
[3] Hosp Aleman, Dept Crit Care Med, Buenos Aires, DF, Argentina
[4] Hosp Agudos Bernardino Rivadavia, Buenos Aires, DF, Argentina
[5] Hosp Mil Campo Mayo, Buenos Aires, DF, Argentina
关键词
Cardiac output; Stroke volume; Carotid flow; POCUS; Hemodynamic monitoring; FLUID RESPONSIVENESS; BLOOD-FLOW;
D O I
10.1007/s10877-022-00938-7
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Transthoracic echocardiography (TTE) is a fundamental tool for hemodynamic monitoring in critical patients. It allows evaluating the left ventricle's stroke volume based on the measurement of the velocity-time integral (VTI) of the left ventricle outflow tract (LVOT). However, in the intensive care unit obtaining adequate echocardiographic views may present a challenge. We propose to measure, as a surrogate of the stroke volume, the carotid flow with a novel technique. This is an observational, prospective, and simple blind study, conducted in the intensive care unit of Sanatorio de los Arcos and Hospital Aleman, in Buenos Aires, Argentina. We measured the carotid systodiastolic flow (CSD) VTI and the carotid systolic flow (CS) VTI at the level of the left supraclavicular fossa and we compared it with the LVOT VTI obtained by TTE. We evaluated 43 subjects. Spearman's correlation coefficient between LVOT VTI and CS VTI was 0.81 (95% CI 0.67-0.89) and between LVOT VTI and CSD VTI was 0.89 (95% CI 0.81-0.94). The Bland-Altman method analysis of the 5-chamber apical window LVOT VTI compared to the CSD VTI showed a bias of - 0.2 (95% CI - 0.82 to 0.43), with a concordance interval between - 4.2 (95% CI - 5.2 to - 3.1) and 3.8 cm (95% CI 2.7 to 4.9). The percentage error was 37.9%. Almost 100% of the values fell within the concordance limits, and no trend was observed in bias across the spectrum of mean variables. Although the CSD VTI could not be interchangeable with the LVOT VTI, it could be considered as its surrogate.
引用
收藏
页码:661 / 667
页数:7
相关论文
共 18 条
[1]   Reporting of method comparison studies: a review of advice, an assessment of current practice, and specific suggestions for future reports [J].
Abu-Arafeh, A. ;
Jordan, H. ;
Drummond, G. .
BRITISH JOURNAL OF ANAESTHESIA, 2016, 117 (05) :569-575
[2]   Changes in stroke volume induced by passive leg raising in spontaneously breathing patients: comparison between echocardiography and Vigileo™/FloTrac™ device [J].
Biais, Matthieu ;
Vidil, Lionel ;
Sarrabay, Philippe ;
Cottenceau, Vincent ;
Revel, Philippe ;
Sztark, Francois .
CRITICAL CARE, 2009, 13 (06) :R195
[3]   Rationale for using the velocity-time integral and the minute distance for assessing the stroke volume and cardiac output in point-of-care settings [J].
Blanco, Pablo .
ULTRASOUND JOURNAL, 2020, 12 (01)
[4]   Consensus on circulatory shock and hemodynamic monitoring. Task force of the European Society of Intensive Care Medicine [J].
Cecconi, Maurizio ;
De Backer, Daniel ;
Antonelli, Massimo ;
Beale, Richard ;
Bakker, Jan ;
Hofer, Christoph ;
Jaeschke, Roman ;
Mebazaa, Alexandre ;
Pinsky, Michael R. ;
Teboul, Jean Louis ;
Vincent, Jean Louis ;
Rhodes, Andrew .
INTENSIVE CARE MEDICINE, 2014, 40 (12) :1795-1815
[5]   A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques [J].
Critchley, LAH ;
Critchley, JAJH .
JOURNAL OF CLINICAL MONITORING AND COMPUTING, 1999, 15 (02) :85-91
[6]  
Eicke BM, 2001, J ULTRAS MED, V20, P1293
[7]   Respiratory changes in aortic blood velocity as an indicator of fluid responsiveness in ventilated patients with septic shock [J].
Feissel, M ;
Michard, F ;
Mangin, I ;
Ruyer, O ;
Faller, JP ;
Teboul, JL .
CHEST, 2001, 119 (03) :867-873
[8]   Reporting Standards for a Bland-Altman Agreement Analysis: A Review of Methodological Reviews [J].
Gerke, Oke .
DIAGNOSTICS, 2020, 10 (05)
[9]   Ventriculoarterial decoupling in human septic shock [J].
Guarracino, Fabio ;
Ferro, Baldassare ;
Morelli, Andrea ;
Bertini, Pietro ;
Baldassarri, Rubia ;
Pinsky, Michael R. .
CRITICAL CARE, 2014, 18 (02)
[10]   Predicting Fluid Responsiveness in Critically Ill Patients by Using Combined End-Expiratory and End-Inspiratory Occlusions With Echocardiography [J].
Jozwiak, Mathieu ;
Depret, Francois ;
Teboul, Jean-Louis ;
Alphonsine, Jean-Emmanuel ;
Lai, Christopher ;
Richard, Christian ;
Monnet, Xavier .
CRITICAL CARE MEDICINE, 2017, 45 (11) :E1131-E1138