Next generation high resolution perovskite direct conversion detector: Monte Carlo design optimisation and virtual clinical trial

被引:1
作者
O'Connell, J. [1 ]
Kundu, S. [1 ]
Saidaminov, M. [1 ]
Bazalova-Carter, M. [1 ]
机构
[1] Univ Victoria, 3800 Finnerty Rd, Victoria, BC V8P 5C2, Canada
关键词
perovskite; CT; CBCT; VCT; detector; mammography; X-RAY-DETECTORS; AMORPHOUS SELENIUM; IMAGE QUALITY; SYSTEM;
D O I
10.1088/1361-6560/acae15
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We implement virtual clinical integration of next-generation perovskite detectors into common x-ray imaging devices. This was achieved by performing Monte Carlo (MC) optimisation of the design and benchmarking of low cost, high spatial resolution, direct conversion perovskite crystal x-ray flat panel imagers for a next generation of breast-, MV-, and kV-cone beam CT detectors. Semiconductor methylammonium lead bromide perovskite crystals energy deposition efficiencies calculated in TOPAS were benchmarked against four common detector materials for twelve detector crystal thicknesses between 40 to 15 mm and ten beam energies ranging from 20 keV to 6 MeV. Based on these simulations, Koning's dedicated breast CT, and Varian's Truebeam kV- and MV-cone beam CT systems were designated as suitable applications for perovskite detectors. System specific Fastcat hybrid MC cone beam CT image simulation was subsequently used to optimise the perovskite detector design and conduct virtual clinical trials. Device-specific optimal perovskite crystal thicknesses were calculated to be 0.30, 0.86, and 1.99 mm for Koning breast CT and Truebeam kV- and MV-cone beam CT systems, respectively. Replacing the current detectors on these machines with low cost perovskite crystal detectors could be advantageous as it would simultaneously yield 12.1%, 9.5% and 86.1% improvements in detective quantum efficiency as well as increases in contrast to noise ratio in brain, lung, and bone tissues.
引用
收藏
页数:11
相关论文
共 29 条
[1]   Geant4-based Monte Carlo simulations on GPU for medical applications [J].
Bert, Julien ;
Perez-Ponce, Hector ;
El Bitar, Ziad ;
Jan, Sebastien ;
Boursier, Yannick ;
Vintache, Damien ;
Bonissent, Alain ;
Morel, Christian ;
Brasse, David ;
Visvikis, Dimitris .
PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (16) :5593-5611
[2]  
Capote R, 2023, PHSP PHOTON1
[3]   2D perovskite-based high spatial resolution X-ray detectors [J].
Datta, Amlan ;
Fiala, John ;
Motakef, Shariar .
SCIENTIFIC REPORTS, 2021, 11 (01)
[4]   High length-to-width aspect ratio lead bromide microwires via perovskite-induced local concentration gradient for X-ray detection [J].
Dennis, Emma ;
Kundu, Soumya ;
Gangadharan, Deepak Thrithamarassery ;
Huang, Jingjun ;
Burlakov, Victor M. ;
Richtsmeier, Devon ;
Bazalova-Carter, Magdalena ;
Leitch, David C. ;
Saidaminov, Makhsud, I .
CRYSTENGCOMM, 2021, 23 (11) :2215-2221
[5]   Intercomparison of methods for image quality characterization. II. Noise power spectrum [J].
Dobbins, James T., III ;
Samei, Ehsan ;
T Ranger, Nicole ;
Chen, Ying .
MEDICAL PHYSICS, 2006, 33 (05) :1466-1475
[6]  
Dunning CAS, 2020, J MED IMAGING, V7, DOI [10.1117/1.JMI.73.033502, 10.1117/1.JMI.7.3.033502]
[7]   PRACTICAL CONE-BEAM ALGORITHM [J].
FELDKAMP, LA ;
DAVIS, LC ;
KRESS, JW .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1984, 1 (06) :612-619
[8]  
Hern?ndez G., 2016, J OPEN SOURCE SOFTW, V1, DOI [10.21105/joss.00062, DOI 10.21105/JOSS.00062]
[9]   A-site Cation Engineering for Highly Efficient MAPbI3 Single-Crystal X-ray Detector [J].
Huang, Yanmin ;
Qiao, Lu ;
Jiang, Yuanzhi ;
He, Tingwei ;
Long, Run ;
Yang, Fan ;
Wang, Lin ;
Lei, Xiaojuan ;
Yuan, Mingjian ;
Chen, Jun .
ANGEWANDTE CHEMIE-INTERNATIONAL EDITION, 2019, 58 (49) :17834-17842
[10]   Generation of polychromatic projection for dedicated breast computed tomography simulation using anthropomorphic numerical phantom [J].
Jeon, Hosang ;
Youn, Hanbean ;
Kim, Jin Sung ;
Nam, Jiho ;
Lee, Jayoung ;
Lee, Juhye ;
Park, Dahl ;
Kim, Wontaek ;
Ki, Yongkan ;
Kim, Donghyun .
PLOS ONE, 2017, 12 (11)