Development of a multi-layer quality assurance program to evaluate the uncertainty of deformable dose accumulation in adaptive radiotherapy

被引:6
作者
Zhong, Hualiang [1 ,2 ]
Garcia-Alvarez, Juan A. [1 ]
Kainz, Kristofer [1 ]
Tai, An [1 ]
Ahunbay, Ergun [1 ]
Erickson, Beth [1 ]
Schultz, Christopher J. [1 ]
Li, X. Allen [1 ]
机构
[1] Med Coll Wisconsin, Dept Radiat Oncol, Milwaukee, WI USA
[2] Med Coll Wisconsin, Dept Radiat Oncol, Milwaukee, WI 53226 USA
关键词
adaptive radiotherapy; deformable image registration; dose accumulation; quality assurance; IMAGE REGISTRATION; INVERSE-CONSISTENCY; ALGORITHMS;
D O I
10.1002/mp.16137
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeDeformable dose accumulation (DDA) has uncertainties which impede the implementation of DDA-based adaptive radiotherapy (ART) in clinic. The purpose of this study is to develop a multi-layer quality assurance (MLQA) program to evaluate uncertainties in DDA. MethodsA computer program is developed to generate a pseudo-inverse displacement vector field (DVF) for each deformable image registration (DIR) performed in Accuray's PreciseART. The pseudo-inverse DVF is first used to calculate a pseudo-inverse consistency error (PICE) and then implemented in an energy and mass congruent mapping (EMCM) method to reconstruct a deformed dose. The PICE is taken as a metric to estimate DIR uncertainties. A pseudo-inverse dose agreement rate (PIDAR) is used to evaluate the consequence of the DIR uncertainties in DDA and the principle of energy conservation is used to validate the integrity of dose mappings. The developed MLQA program was tested using the data collected from five representative cancer patients treated with tomotherapy. ResultsDIRs were performed in PreciseART to generate primary DVFs for the five patients. The fidelity index and PICE of these DVFs on average are equal to 0.028 mm and 0.169 mm, respectively. With the criteria of 3 mm/3% and 5 mm/5%, the PIDARs of the PreciseART-reconstructed doses are 73.9 +/- 4.4% and 87.2 +/- 3.3%, respectively. The PreciseART and EMCM-based dose reconstructions have their deposited energy changed by 5.6 +/- 3.9% and 2.6 +/- 1.5% in five GTVs, and by 9.2 +/- 7.8% and 4.7 +/- 3.6% in 30 OARs, respectively. ConclusionsA pseudo-inverse map-based EMCM program has been developed to evaluate DIR and dose mapping uncertainties. This program could also be used as a sanity check tool for DDA-based ART.
引用
收藏
页码:1766 / 1778
页数:13
相关论文
共 26 条
  • [21] Patient-specific quality assurance for proton depth dose distribution using a multi-layer ionization chamber in a single-ring wobbling method
    Kato, Takahiro
    Arai, Kazuhiro
    Sagara, Tatsuhiko
    Kato, Ryohei
    Yamazaki, Yuhei
    Oyama, Sho
    RADIOLOGICAL PHYSICS AND TECHNOLOGY, 2019, 12 (03) : 305 - 311
  • [22] Quality Assurance Decisions with Air Models: A Case Study of Imputation of Missing Input Data Using EPA's Multi-layer Model
    Bowker, George E.
    Schwede, Donna B.
    Lear, Gary G.
    Warren-Hicks, William J.
    Finkelstein, Peter L.
    WATER AIR AND SOIL POLLUTION, 2011, 222 (1-4) : 391 - 402
  • [23] Quality Assurance Decisions with Air Models: A Case Study of Imputation of Missing Input Data Using EPA’s Multi-layer Model
    George E. Bowker
    Donna B. Schwede
    Gary G. Lear
    William J. Warren-Hicks
    Peter L. Finkelstein
    Water, Air, & Soil Pollution, 2011, 222 : 391 - 402
  • [24] Development of a High-Precision, Image-Guided Radiotherapy, Multi-Purpose Radiation Isocenter Quality-Assurance Calibration and Checking System
    Liu, C.
    Yan, G.
    Helmig, R.
    Lebron, S.
    Kahler, D.
    MEDICAL PHYSICS, 2014, 41 (06) : 165 - 166
  • [25] Quality Assurance in the Prospective Multi-institutional Trial on Definitive Radiotherapy Using High-dose-rate Intracavitary Brachytherapy for Uterine Cervical Cancer: The Individual Case Review
    Toita, Takafumi
    Oguchi, Masahiko
    Ohno, Tatsuya
    Kato, Shingo
    Niibe, Yuzuru
    Kodaira, Takeshi
    Kazumoto, Tomoko
    Kataoka, Masaaki
    Shikama, Naoto
    Kenjo, Masahiro
    Teshima, Teruki
    Kagami, Yoshikazu
    JAPANESE JOURNAL OF CLINICAL ONCOLOGY, 2009, 39 (12) : 813 - 819
  • [26] Development and implementation of optimized endogenous contrast sequences for delineation in adaptive radiotherapy on a 1.5T MR-linear-accelerator: a prospective R-IDEAL stage 0-2a quantitative/qualitative evaluation of in vivo site-specific quality-assurance using a 3D T2 fat-suppressed platform for head and neck cancer
    Salzillo, Travis C.
    Dresner, M. Alex
    Way, Ashley
    Wahid, Kareem A.
    Mcdonald, Brigid A.
    Mulder, Sam
    Naser, Mohamed A.
    He, Renjie
    Ding, Yao
    Yoder, Alison
    Ahmed, Sara
    Corrigan, Kelsey L.
    Manzar, Gohar S.
    Andring, Lauren
    Pinnix, Chelsea
    Stafford, R. Jason
    Mohamed, Abdallah S. R.
    Christodouleas, John
    Wang, Jihong
    Fuller, Clifton David
    JOURNAL OF MEDICAL IMAGING, 2023, 10 (06)