Development of a multi-layer quality assurance program to evaluate the uncertainty of deformable dose accumulation in adaptive radiotherapy

被引:6
|
作者
Zhong, Hualiang [1 ,2 ]
Garcia-Alvarez, Juan A. [1 ]
Kainz, Kristofer [1 ]
Tai, An [1 ]
Ahunbay, Ergun [1 ]
Erickson, Beth [1 ]
Schultz, Christopher J. [1 ]
Li, X. Allen [1 ]
机构
[1] Med Coll Wisconsin, Dept Radiat Oncol, Milwaukee, WI USA
[2] Med Coll Wisconsin, Dept Radiat Oncol, Milwaukee, WI 53226 USA
关键词
adaptive radiotherapy; deformable image registration; dose accumulation; quality assurance; IMAGE REGISTRATION; INVERSE-CONSISTENCY; ALGORITHMS;
D O I
10.1002/mp.16137
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeDeformable dose accumulation (DDA) has uncertainties which impede the implementation of DDA-based adaptive radiotherapy (ART) in clinic. The purpose of this study is to develop a multi-layer quality assurance (MLQA) program to evaluate uncertainties in DDA. MethodsA computer program is developed to generate a pseudo-inverse displacement vector field (DVF) for each deformable image registration (DIR) performed in Accuray's PreciseART. The pseudo-inverse DVF is first used to calculate a pseudo-inverse consistency error (PICE) and then implemented in an energy and mass congruent mapping (EMCM) method to reconstruct a deformed dose. The PICE is taken as a metric to estimate DIR uncertainties. A pseudo-inverse dose agreement rate (PIDAR) is used to evaluate the consequence of the DIR uncertainties in DDA and the principle of energy conservation is used to validate the integrity of dose mappings. The developed MLQA program was tested using the data collected from five representative cancer patients treated with tomotherapy. ResultsDIRs were performed in PreciseART to generate primary DVFs for the five patients. The fidelity index and PICE of these DVFs on average are equal to 0.028 mm and 0.169 mm, respectively. With the criteria of 3 mm/3% and 5 mm/5%, the PIDARs of the PreciseART-reconstructed doses are 73.9 +/- 4.4% and 87.2 +/- 3.3%, respectively. The PreciseART and EMCM-based dose reconstructions have their deposited energy changed by 5.6 +/- 3.9% and 2.6 +/- 1.5% in five GTVs, and by 9.2 +/- 7.8% and 4.7 +/- 3.6% in 30 OARs, respectively. ConclusionsA pseudo-inverse map-based EMCM program has been developed to evaluate DIR and dose mapping uncertainties. This program could also be used as a sanity check tool for DDA-based ART.
引用
收藏
页码:1766 / 1778
页数:13
相关论文
共 26 条
  • [1] Deep learning based uncertainty prediction of deformable image registration for contour propagation and dose accumulation in online adaptive radiotherapy
    Smolders, A.
    Lomax, A.
    Weber, D. C.
    Albertini, F.
    PHYSICS IN MEDICINE AND BIOLOGY, 2023, 68 (24)
  • [2] Development of a deformable dosimetric phantom to verify dose accumulation algorithms for adaptive radiotherapy
    Zhong, Hualiang
    Adams, Jeffrey
    Glide-Hurst, Carri
    Zhang, Hualin
    Li, Haisen
    Chetty, Indrin J.
    JOURNAL OF MEDICAL PHYSICS, 2016, 41 (02) : 106 - 114
  • [3] Deformable dose accumulation is required for adaptive radiotherapy practice
    Zhong, Hualiang
    Pursley, Jennifer M.
    Rong, Yi
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2024, 25 (08):
  • [4] The impact of robustness of deformable image registration on contour propagation and dose accumulation for head and neck adaptive radiotherapy
    Zhang, Lian
    Wang, Zhi
    Shi, Chengyu
    Long, Tengfei
    Xu, X. George
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2018, 19 (04): : 185 - 194
  • [5] An uncertainty metric to evaluate deformation vector fields for dose accumulation in radiotherapy
    Takemura, Akihiro
    Nagano, Akira
    Kojima, Hironori
    Ikeda, Tomohiro
    Yokoyama, Noriomi
    Tsukamoto, Kosuke
    Noto, Kimiya
    Isomura, Naoki
    Ueda, Shinichi
    Kawashima, Hiroki
    PHYSICS & IMAGING IN RADIATION ONCOLOGY, 2018, 6 : 77 - 82
  • [6] Quality assurance of an established online adaptive radiotherapy program: patch and software upgrade
    Bassiri, Nema
    Bayouth, John
    Chuong, Michael D.
    Kotecha, Rupesh
    Weiss, Yonatan
    Mehta, Minesh P.
    Gutierrez, Alonso N.
    Mittauer, Kathryn E.
    FRONTIERS IN ONCOLOGY, 2024, 14
  • [7] Use of a DVH overlay technique for quality assurance of deformable image registration-based dose accumulation
    Kainz, Kristofer
    Garcia Alvarez, Juan
    Zhong, Hualiang
    Lim, Sara
    Ahunbay, Ergun
    Tai, An
    Erickson, Beth
    Lawton, Colleen
    Li, X. Allen
    MEDICAL PHYSICS, 2022, 49 (01) : 611 - 623
  • [8] Validating Dose Uncertainty Estimates Produced by AUTODIRECT: An Automated Program to Evaluate Deformable Image Registration Accuracy
    Kim, Hojin
    Chen, Josephine
    Phillips, Justin
    Pukala, Jason
    Yom, Sue S.
    Kirby, Neil
    TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2017, 16 (06) : 885 - 892
  • [9] End-to-end empirical validation of dose accumulation in MRI-guided adaptive radiotherapy for prostate cancer using an anthropomorphic deformable pelvis phantom
    Bohoudi, Omar
    Lagerwaard, Frank J.
    Bruynzeel, Anna M. E.
    Niebuhr, Nina, I
    Johnen, Wibke
    Senan, Suresh
    Slotman, Ben J.
    Pfaffenberger, Asja
    Palacios, Miguel A.
    RADIOTHERAPY AND ONCOLOGY, 2019, 141 : 200 - 207
  • [10] A deformable head and neck phantom with in-vivo dosimetry for adaptive radiotherapy quality assurance
    Graves, Yan Jiang
    Smith, Arthur-Allen
    Mcilvena, David
    Manilay, Zherrina
    Lai, Yuet Kong
    Rice, Roger
    Mell, Loren
    Jia, Xun
    Jiang, Steve B.
    Cervino, Laura
    MEDICAL PHYSICS, 2015, 42 (04) : 1490 - 1497