Efficacy of poller screw in addition to lag screw in the treatment of intertrochanteric fractures with proximal femoral nail: a biomechanical evaluation

被引:2
|
作者
Horoz, Levent [1 ]
Cakmak, Mehmet Fevzi [1 ]
Cici, Hakan [2 ]
机构
[1] Kirsehir Ahi Evran Univ, Kirsehir, Turkiye
[2] Demokrasi Univ, Izmir, Turkiye
关键词
Intertrochanteric fracture; Poller screw; Biomechanical study; Proximal femoral nail; SLIDING HIP SCREW; PERTROCHANTERIC FRACTURES; CEMENT AUGMENTATION; NECK FRACTURES; RISK-FACTORS; FIXATION; ANTIROTATION; FAILURE; PLATE;
D O I
10.1007/s00068-024-02477-z
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Purpose The most common type of failure in treating intertrochanteric fractures with proximal femoral nails is cut-out due to varus collapse. We aim to evaluate the effect of the poller screw applied to the proximal fragment and the lag screw on varus collapse and stability in intertrochanteric fractures. Methods An unstable intertrochanteric fracture model without medial support was simulated in 20 synthetic femur models. In the poller screw group, in addition to the lag screw, pole screws were applied to the proximal fragment superior and inferior to the lag screw. In the progressive cyclic loading test, starting from 100 N, the loading was increased by 50 N in each cycle, and the test was continued until the maximum load at which failure occurred as a result of conditioning cycles and progressive cyclic loading tests, stiffness, type of failure, force at failure, lag screw displacement, and varus collapse were recorded. Result The average stiffness was found to be 124.705 N/mm in the poller screw group and 102.77 N/mm in the control group (P < 0.001). The maximum load to failure was 1897.10 N in the poller screw group and 1475.20 N in the control group (P < 0.001). The average displacement of the lag screw within the femoral head was 0.85 mm in the poller screw group and 3.60 mm in the control group (P < 0.001). Conclusion As a result, it has been shown that poller screws applied around the lag screw increase fixation stiffness and reduce varus collapse.
引用
收藏
页码:1591 / 1598
页数:8
相关论文
共 50 条
  • [31] Biomechanical evaluation of intramedullary retrograde nail and dynamic condylar screw used in unstable distal femoral fractures
    Dizdar, Ayberk
    Celik, Talip
    Ozkan, Arif
    PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2023, 46 (04) : 1755 - 1764
  • [32] Importance of screw position in intertrochanteric femoral fractures treated by dynamic hip screw
    Guven, M.
    Yavuz, U.
    Kadioglu, B.
    Akman, B.
    Kilincoglu, V.
    Unay, K.
    Altintas, F.
    ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2010, 96 (01) : 20 - 26
  • [33] Biomechanical comparison of dynamic hip screw and Gamma nail for the treatment of unstable trochanteric fractures: a finite element study
    Huang, Xiaowei
    Yu, Baoqing
    Gu, Yong
    Li, Zexiang
    INTERNATIONAL JOURNAL OF CLINICAL AND EXPERIMENTAL MEDICINE, 2017, 10 (05): : 7867 - 7874
  • [34] Prospective randomised controlled trial of an intramedullary nail versus a sliding hip screw for intertrochanteric fractures of the femur
    Aktselis, Ioannis
    Kokoroghiannis, Constantine
    Fragkomichalos, Evaggelos
    Koundis, Georgios
    Deligeorgis, Anastasios
    Daskalakis, Emmanouil
    Vlamis, John
    Papaioannou, Nikolaos
    INTERNATIONAL ORTHOPAEDICS, 2014, 38 (01) : 155 - 161
  • [35] Comparison of cutout resistance of dynamic condylar screw and proximal femoral nail in reverse oblique trochanteric fractures: A biomechanical study
    Cheema, Gursimrat Singh
    Rastogi, Amit
    Singh, Vakil
    Goel, Satish Chandra
    Mishra, Diwakar
    Arora, Sumit
    INDIAN JOURNAL OF ORTHOPAEDICS, 2012, 46 (03) : 259 - 265
  • [36] The new technique of precise insertion of lag screw in an operative treatment of trochanteric femoral fractures with a short intramedullary nail
    Nishiura, Takashi
    Nozawa, Masahiko
    Morio, Hidenori
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2009, 40 (10): : 1077 - 1083
  • [37] A Comparison of Proximal Femoral Nail Antirotation and Dynamic Hip Screw Devices in Trochanteric Fractures
    Zou, J.
    Xu, Y.
    Yang, H.
    JOURNAL OF INTERNATIONAL MEDICAL RESEARCH, 2009, 37 (04) : 1057 - 1064
  • [38] The Comparison of Dynamic Condylar Screw Plate to Proximal Femoral Nail in Reverse Oblique and Transverse Intertrochanteric Fractures: A Retrospective Study on 61 Patients
    Sensoez, Ersin
    Ergun, Selim
    Kayaalp, Mahmut Enes
    Eceviz, Engin
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2023, 15 (03)
  • [39] Biopolymer augmentation of the lag screw in the treatment of femoral neck fractures - a biomechanical in-vitro study
    A Paech
    E Wilde
    AP Schulz
    G Heinrichs
    R Wendlandt
    C Queitsch
    B Kienast
    Ch Jürgens
    European Journal of Medical Research, 15
  • [40] Proximal femoral locking plate versus dynamic hip screw for unstable intertrochanteric femoral fractures
    Dhamangaonkar, Anoop C.
    Joshi, Deepak
    Goregaonkar, Arvind B.
    Tawari, Akhil A.
    JOURNAL OF ORTHOPAEDIC SURGERY, 2013, 21 (03) : 317 - 322