Efficacy of poller screw in addition to lag screw in the treatment of intertrochanteric fractures with proximal femoral nail: a biomechanical evaluation

被引:2
|
作者
Horoz, Levent [1 ]
Cakmak, Mehmet Fevzi [1 ]
Cici, Hakan [2 ]
机构
[1] Kirsehir Ahi Evran Univ, Kirsehir, Turkiye
[2] Demokrasi Univ, Izmir, Turkiye
关键词
Intertrochanteric fracture; Poller screw; Biomechanical study; Proximal femoral nail; SLIDING HIP SCREW; PERTROCHANTERIC FRACTURES; CEMENT AUGMENTATION; NECK FRACTURES; RISK-FACTORS; FIXATION; ANTIROTATION; FAILURE; PLATE;
D O I
10.1007/s00068-024-02477-z
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Purpose The most common type of failure in treating intertrochanteric fractures with proximal femoral nails is cut-out due to varus collapse. We aim to evaluate the effect of the poller screw applied to the proximal fragment and the lag screw on varus collapse and stability in intertrochanteric fractures. Methods An unstable intertrochanteric fracture model without medial support was simulated in 20 synthetic femur models. In the poller screw group, in addition to the lag screw, pole screws were applied to the proximal fragment superior and inferior to the lag screw. In the progressive cyclic loading test, starting from 100 N, the loading was increased by 50 N in each cycle, and the test was continued until the maximum load at which failure occurred as a result of conditioning cycles and progressive cyclic loading tests, stiffness, type of failure, force at failure, lag screw displacement, and varus collapse were recorded. Result The average stiffness was found to be 124.705 N/mm in the poller screw group and 102.77 N/mm in the control group (P < 0.001). The maximum load to failure was 1897.10 N in the poller screw group and 1475.20 N in the control group (P < 0.001). The average displacement of the lag screw within the femoral head was 0.85 mm in the poller screw group and 3.60 mm in the control group (P < 0.001). Conclusion As a result, it has been shown that poller screws applied around the lag screw increase fixation stiffness and reduce varus collapse.
引用
收藏
页码:1591 / 1598
页数:8
相关论文
共 50 条
  • [21] Treatment options for unstable trochanteric fractures: Screw or helical proximal femoral nail
    Bajpai Jeetendra
    Maheshwari Rajesh
    Bajpai Akansha
    Saini Sumit
    中华创伤杂志英文版, 2015, 18 (06) : 342 - 346
  • [22] Minimally invasive screw plates for surgery of unstable intertrochanteric femoral fractures: A biomechanical comparative study
    Ropars, Mickael
    Mitton, David
    Skalli, Wafa
    CLINICAL BIOMECHANICS, 2008, 23 (08) : 1012 - 1017
  • [23] Biomechanical Investigation of an Integrated 2-Screw Cephalomedullary Nail Versus a Sliding Hip Screw in Unstable Intertrochanteric Fractures
    Santoni, Brandon G.
    Diaz, Miguel A.
    Stoops, T. Kyle
    Lannon, Sean
    Ali, Ashley
    Sanders, Roy W.
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2019, 33 (02) : 82 - 87
  • [24] The nail-shaft-axis of the of proximal femoral nail antirotation (PFNA) is an important prognostic factor in the operative treatment of intertrochanteric fractures
    Jiamton, Chittawee
    Boernert, Katja
    Babst, Reto
    Beeres, Frank J. P.
    Link, Bjorn-Christian
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2018, 138 (03) : 339 - 349
  • [25] Proximal femoral locking plate versus dynamic hip screw for unstable intertrochanteric femoral fractures
    Upadhyay, Sachin
    Raza, H. K. T.
    JOURNAL OF ORTHOPAEDIC SURGERY, 2014, 22 (01) : 130 - 131
  • [26] Clinical Outcomes of the Proximal Femoral Nails in the Treatment of Intertrochanteric Fractures
    Cetin, Engin
    Kilic, Mehmet
    Kilic, Feyzi
    Aybar, Ahmet
    Ozkazanli, Gokhan
    GAZI MEDICAL JOURNAL, 2019, 30 (01): : 39 - 42
  • [27] In silico biomechanical analysis of poller screw-assisted small-diameter intramedullary nail in the treatment of distal tibial fractures
    Tan, Jinchuan
    Yang, Yang
    Wang, Mian
    Huang, Xuecheng
    Ouyang, Hanbin
    Zhao, Dongliang
    Huang, Gang
    Deng, Yuping
    Huang, Wenhua
    FRONTIERS IN BIOENGINEERING AND BIOTECHNOLOGY, 2023, 11
  • [28] Comparison of two methods of locating proximal femoral nail anti-rotation in the treatment of femoral intertrochanteric fractures
    Tian, Rong-hua
    Zhang, Qin-ming
    Chu, Feng-long
    Li, Xiao-yan
    Jiang, Zhen
    Han, Liang
    Sun, Peng
    Wang, Hai-bin
    Chi, Yu-lei
    Wu, Bin
    JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2020, 15 (01)
  • [29] Proximal femoral nail antirotation versus dynamic hip screw fixation for treatment of osteoporotic type 31-A1 intertrochanteric femoral fractures in elderly patients
    Zeng, Xianshang
    Zhang, Nan
    Zeng, Dan
    Zhang, Lili
    Xu, Ping
    Cao, Lei
    Yu, Weiguang
    Zhan, Ke
    Zhang, Xinchao
    JOURNAL OF INTERNATIONAL MEDICAL RESEARCH, 2017, 45 (03) : 1109 - 1123
  • [30] Dynamic Hip Screw in Comparison with Proximal Femoral Nail Technique in Intertrochanteric Femur Fracture Patients
    Khan, Mudir
    Siraj, Muhammad
    Ali, Abbas