Clinical Characterization of a Table Mounted Range Shifter Board for Synchrotron-Based Intensity Modulated Proton Therapy for Pediatric Craniospinal Irradiation

被引:0
|
作者
Hrinivich, William T. [1 ,2 ]
Li, Heng [1 ,2 ]
Tran, Anh [1 ,2 ]
Acharya, Sahaja [1 ,2 ]
Ladra, Matthew M. [1 ,2 ]
Sheikh, Khadija [1 ,2 ]
机构
[1] Johns Hopkins Univ, Sch Med, Dept Radiat Oncol & Mol Radiat Sci, Baltimore, MD 21287 USA
[2] Johns Hopkins Univ, Johns Hopkins Proton Therapy Ctr, Sch Med, Washington, DC 20016 USA
关键词
craniospinal irradiation; pediatric radiotherapy; proton therapy; pencil beam scanning; image-guided radiotherapy; 2ND CANCERS; BEAM; MEDULLOBLASTOMA; REDUCTION; DELIVERY; SYSTEM; PHOTON; BOLUS; RISK;
D O I
10.3390/cancers15112882
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Intensity modulated proton therapy provides unparalleled normal tissue sparing for the treatment of pediatric cancers such as medulloblastoma using craniospinal irradiation. However, proton delivery systems have a minimum deliverable energy and depth, requiring range shifter devices for the treatment of shallow targets and/or small anatomy. Standard range shifters are mounted to the treatment nozzle and introduce lateral scatter which, unfortunately, increases normal tissue dose. In this study we designed, manufactured, and characterized a range shifter board which rests directly under the patient during treatment to maintain a shallow treatment depth while minimizing lateral scatter, similar to previously proposed designs but optimized for our synchrotronbased delivery system. We present the dosimetric measurements we performed to characterize the device for treatment planning including severity of artifacts in on-board images. We compared the standard and range shifter board planning approaches in an anthropomorphic phantom and delivered the treatment plans to confirm the dosimetric differences. Finally, we compared treatment plans from two previously treated pediatric patients, demonstrating the ability to significantly reduce lung and kidney dose using this device compared to a standard range shifter. This device is now in routine use at our clinic. Purpose: To report our design, manufacturing, commissioning and initial clinical experience with a table-mounted range shifter board (RSB) intended to replace the machine-mounted range shifter (MRS) in a synchrotron-based pencil beam scanning (PBS) system to reduce penumbra and normal tissue dose for image-guided pediatric craniospinal irradiation (CSI). Methods: A custom RSB was designed and manufactured from a 3.5 cm thick slab of polymethyl methacrylate (PMMA) to be placed directly under patients, on top of our existing couch top. The relative linear stopping power (RLSP) of the RSB was measured using a multi-layer ionization chamber, and output constancy was measured using an ion chamber. End-to-end tests were performed using the MRS and RSB approaches using an anthropomorphic phantom and radiochromic film measurements. Cone beam CT (CBCT) and 2D planar kV X-ray image quality were compared with and without the RSB present using image quality phantoms. CSI plans were produced using MRS and RSB approaches for two retrospective pediatric patients, and the resultant normal tissue doses were compared. Results: The RLSP of the RSB was found to be 1.163 and provided computed penumbra of 6.9 mm in the phantom compared to 11.8 mm using the MRS. Phantom measurements using the RSB demonstrated errors in output constancy, range, and penumbra of 0.3%, 0.8%, and 0.6 mm, respectively. The RSB reduced mean kidney and lung dose compared to the MRS by 57.7% and 46.3%, respectively. The RSB decreased mean CBCT image intensities by 86.8 HU but did not significantly impact CBCT or kV spatial resolution providing acceptable image quality for patient setup. Conclusions: A custom RSB for pediatric proton CSI was designed, manufactured, modeled in our TPS, and found to significantly reduce lateral proton beam penumbra compared to a standard MRS while maintaining CBCT and kV image-quality and is in routine use at our center.
引用
收藏
页数:13
相关论文
共 16 条
  • [1] Robustly optimized hybrid intensity-modulated proton therapy for craniospinal irradiation
    Noufal, Manthala Padannayil
    Sharma, Shamurailatpam Dayananda
    Krishnan, Ganapathy
    Sawant, Mayur
    Gaikwad, Utpal
    Jalali, Rakesh
    JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2022, 18 (06) : 1597 - 1603
  • [2] Early in vivo Radiation Damage Quantification for Pediatric Craniospinal Irradiation Using Longitudinal MRI for Intensity Modulated Proton Therapy
    Chang, Chih-Wei
    Goette, Matt
    Kadom, Nadja
    Wang, Yinan
    Wynne, Jacob
    Wang, Tonghe
    Liu, Tian
    Esiashvili, Natia
    Zhou, Jun
    Eaton, Bree R.
    Yang, Xiaofeng
    ADVANCES IN RADIATION ONCOLOGY, 2023, 8 (05)
  • [3] Use of Intensity Modulated Proton Therapy (IMPT) for Reducing the Dose to Cochlea in Craniospinal Irradiation (CSI) of Pediatric Patients
    Dormer, J.
    Kassaee, A.
    Lin, H.
    Ding, X.
    Lustig, R.
    MEDICAL PHYSICS, 2014, 41 (06) : 302 - 302
  • [4] Evaluating Intensity Modulated Proton Therapy Relative to Passive Scattering Proton Therapy for Increased Vertebral Column Sparing in Craniospinal Irradiation in Growing Pediatric Patients
    Giantsoudi, Drosoula
    Seco, Joao
    Eaton, Bree R.
    Simeone, F. Joseph
    Kooy, Hanne
    Yock, Torunn I.
    Tarbell, Nancy J.
    DeLaney, Thomas F.
    Adams, Judith
    Paganetti, Harald
    MacDonald, Shannon M.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2017, 98 (01): : 37 - 46
  • [5] Redefine the role of range shifter in treating bilateral head and neck cancer in the era of Intensity Modulated Proton Therapy
    Ding, Xuanfeng
    Li, Xiaoqiang
    Qin, An
    Zhou, Jun
    Yan, Di
    Chen, Peter
    Prakash, Chinnaiyan
    Stevens, Craig
    Deraniyagala, Rohan
    Kabolizadeh, Peyman
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2018, 19 (05): : 749 - 755
  • [6] Impact of spot positional errors in robustly optimized intensity-modulated proton therapy plan of craniospinal irradiation
    Noufal, Manthala Padannayil
    Sharma, Shamurailatpam Dayananda
    Patro, Katikeshwar
    Arjunan, Manikandan
    Krishnan, Ganapathy
    Tyagarajan, Rajesh
    Rana, Suresh
    Chillukuri, Srinivas
    Jalali, Rakesh
    RADIOLOGICAL PHYSICS AND TECHNOLOGY, 2021, 14 (03) : 271 - 278
  • [7] Impact of spot positional errors in robustly optimized intensity-modulated proton therapy plan of craniospinal irradiation
    Manthala Padannayil Noufal
    Shamurailatpam Dayananda Sharma
    Katikeshwar Patro
    Manikandan Arjunan
    Ganapathy Krishnan
    Rajesh Tyagarajan
    Suresh Rana
    Srinivas Chillukuri
    Rakesh Jalali
    Radiological Physics and Technology, 2021, 14 : 271 - 278
  • [8] Intensity Modulated Proton Therapy for Craniospinal Irradiation: Organ-at-Risk Exposure and a Low-Gradient Junctioning Technique
    Stoker, Joshua B.
    Grant, Jonathan
    Zhu, X. Ronald
    Pidikiti, Rajesh
    Mahajan, Anita
    Grosshans, David R.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2014, 90 (03): : 637 - 644
  • [9] A Dosimetric Comparison of Intensity-Modulated Proton Therapy Optimization Techniques for Pediatric Craniopharyngiomas: A Clinical Case Study
    Yeung, Daniel
    McKenzie, Craig
    Indelicato, Daniel J.
    PEDIATRIC BLOOD & CANCER, 2014, 61 (01) : 89 - 94
  • [10] Dosimetric characterization of a silicon diode detector in cyclotron-based passively scattered and synchrotron-based scanning clinical proton beams
    Raffaele, Luigi
    Ciocca, Mario
    Mirandola, Alfredo
    Spampinato, Sofia
    Cirrone, G. A. Pablo
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2018, 891 : 125 - 132