Evaluate the effect of bone density variation on stress distribution at the bone-implant interface using numerical analysis

被引:0
作者
Hindurao, Bhargav [1 ,2 ]
Gujare, Aditya [1 ]
Jadhav, Harshavardhan [1 ]
Dhatrak, Pankaj [1 ]
机构
[1] Dr Vishwanath Karad MIT World Peace Univ, Dept Mech Engn, Pune, Maharashtra, India
[2] Dr Vishwanath Karad MIT World Peace Univ, Dept Mech Engn, S 124 Paud Rd Kothrud, Pune 411038, Maharashtra, India
关键词
Dental implants; cancellous; cortical; bone density; von Mises stress; FINITE-ELEMENT-ANALYSIS; ORTHOTROPIC MATERIAL; PROSTHESES; QUALITY; NUMBERS; MODELS;
D O I
10.1177/09544119241240940
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Graphical abstract The current study aims to comprehend how different bone densities affect stress distribution at the bone-implant interface. This will help understand the behaviour and help predict success rates of the implant planted in different bone densities. The process of implantation involves the removal of bone from a small portion of the jawbone to replace either a lost tooth or an infected one and an implant is inserted in the cavity made as a result. Now the extent of fixation due to osseointegration is largely dependent on the condition of the bone in terms of the density. Generally, the density of the bone is classified into four categories namely D1, D2, D3, and D4; with D1 being purely cortical and D4 having higher percentage of cancellous bordered by cortical bone. A bone model with a form closely resembling the actual bone was made using 3D CAD software and was meshed using Hyper Mesh. The model was subjected to an oblique load of 120 N at 70 degrees to the vertical to replicate occlusal loading. A finite element static analysis was done using Abaqus software. The stress distribution contours at the bone-implant contact zone were studied closely to understand the changes as a result of the varying density. It was revealed that as the quantity of the cancellous bone increased from D1 to D4 the cortical peak stress levels dropped. The bone density and the corresponding change in the material characteristics was also responsible for the variation in the peak stress and displacement values.
引用
收藏
页码:463 / 470
页数:8
相关论文
共 36 条
  • [1] ADELL R, 1985, INT DENT J, V35, P259
  • [2] Al-Khafagy HH., 2010, J INT DENT MED RES, V3, P11
  • [3] Influence of occlusal forces on stress distribution in preloaded dental implant screws
    Alkan, I
    Sertgöz, A
    Ekici, E
    [J]. JOURNAL OF PROSTHETIC DENTISTRY, 2004, 91 (04) : 319 - 325
  • [4] [Anonymous], 2011, ABAQUS ANAL USER MAN, V4
  • [5] Mechanical analysis of a dental implant system under 3 contact conditions and with 2 mechanical factors
    Chang, Hung-Chih
    Li, Hung-Yuan
    Chen, Yen-Nien
    Chang, Chih-Han
    Wang, Chau-Hsiang
    [J]. JOURNAL OF PROSTHETIC DENTISTRY, 2019, 122 (04) : 376 - 382
  • [6] Dhatrak P., 2018, LECT NOTES MECH ENG, V0, P203, DOI DOI 10.1007/978-981-10-6002-1_16
  • [7] A mathematical approach to estimate micro-displacement of a dental implant using electromagnetic Frequency Response Analysis
    Dhatrak, Pankaj
    Bhadri, Karan
    Karnik, Niharika
    Bora, Urvi
    [J]. JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2023, 140
  • [8] Influence of surface coatings on the stress distribution by varying friction contact at implant-bone interface using finite element analysis
    Dhatrak, Pankaj
    Kurup, Alekh
    Khasnis, Neha
    [J]. PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2023, 237 (02) : 233 - 242
  • [9] Influence of cutting flutes on stress distribution for selected dental implants: Numerical studies
    Dhatrak, Pankaj
    Shirsat, Uddhav
    Sumanth, S.
    Deshmukh, Vijay
    [J]. MATERIALS TODAY-PROCEEDINGS, 2021, 38 : 2680 - 2686
  • [10] Significance of Orthotropic Material Models to Predict Stress Around Bone-Implant Interface Using Numerical Simulation
    Dhatrak, Pankaj
    Girme, Vedant
    Shirsat, Uddhav
    Sumanth, S.
    Deshmukh, Vijay
    [J]. BIONANOSCIENCE, 2019, 9 (03) : 652 - 659