Optical design of micro-focus X-ray photon correlation spectroscopy beamline at Shenzhen Light Source

被引:0
|
作者
Lei, Tao [1 ,2 ]
Tan, Yonggen [1 ,2 ]
Zhang, Zheng [1 ,2 ]
Chen, Siyuan [1 ,2 ]
Feng, Jun [1 ,2 ]
机构
[1] Southern Univ Sci & Technol SUSTech, Dept Mat Sci & Engn, Shenzhen, Guangdong, Peoples R China
[2] Southern Univ Sci & Technol, Guangdong Prov Key Lab Funct Oxide Mat & Devices, Shenzhen 518055, Guangdong, Peoples R China
基金
中国博士后科学基金;
关键词
Beamline; XPCS; Synchrotron radiation; Tender X-ray; Coherence; Diffraction-limited source; UNDULATOR; MONOCHROMATOR; DIFFRACTION; PERFORMANCE; RADIATION;
D O I
10.1016/j.nima.2022.168006
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
An optical design study of a beamline proposed for the 3 GeV synchrotron radiation facility, the Shenzhen Light Source, is described. The beamline was designed to cover an energy range from 2.0 to 20 keV with two experimental station, one for X-ray absorption spectroscopy (XAS) experiments and the other for X-ray photon correlation spectroscopy (XPCS) experiments. A 4 m planar undulator with a relatively short magnetic period (20 mm) was used as the radiation source. Horizontally and vertically focusing mirrors separately focus the source along the horizontal and vertical directions, respectively, thus enhancing the flexibility of the beamline. The focus point was employed as the secondary source, and a pair of KB mirrors were used to achieve a near-circular focus spot of about 4.5 mu m x 4.5 mu m. A Si(111) DCM was selected to produce the monochromatic light that covers the 2.0-20 keV energy range. When the photon energy is lower than 10 keV, the energy resolution can be better than 1.47 x 10-4 and the total flux is larger than 1 x 1014 photons/s. The longitudinal coherence length is in a few mu m range. With aperture size adjustment, the transverse coherence length at the sample position is approximately 3 mu m and the coherent flux is up to 1012 phs/s for the photon energy around 2.5 keV. This excellent coherence performance can be applied for the purpose of X-ray photon correlation spectroscopy related experiments.
引用
收藏
页数:7
相关论文
共 50 条
  • [1] Design of the Soft X-ray Tomography Beamline at Taiwan Photon Source
    Su, Yi-Jr
    Fu, Huang-Wen
    Chung, Shih-Chun
    Fung, Hok-Sum
    Liu, Din-Goa
    Huang, Liang-Jen
    Yan, Hong-Yi
    Chou, Yu-Ching
    Yin, Gung-Chian
    Lai, Lee-Jene
    PROCEEDINGS OF THE 12TH INTERNATIONAL CONFERENCE ON SYNCHROTRON RADIATION INSTRUMENTATION (SRI2015), 2016, 1741
  • [2] Optical design of the Short Pulse Soft X-ray Spectroscopy beamline at the Advanced Photon Source
    Reininger, R.
    Keavney, D. J.
    Borland, M.
    Young, L.
    JOURNAL OF SYNCHROTRON RADIATION, 2013, 20 : 654 - 659
  • [3] Characterization of the X-ray coherence properties of an undulator beamline at the Advanced Photon Source
    Ju, Guangxu
    Highland, Matthew J.
    Thompson, Carol
    Eastman, Jeffrey A.
    Fuoss, Paul H.
    Zhou, Hua
    Dejus, Roger
    Stephenson, G. Brian
    JOURNAL OF SYNCHROTRON RADIATION, 2018, 25 : 1036 - 1047
  • [4] X-ray photon correlation spectroscopy
    Shpyrko, Oleg G.
    JOURNAL OF SYNCHROTRON RADIATION, 2014, 21 : 1057 - 1064
  • [5] The X-ray Correlation Spectroscopy instrument at the Linac Coherent Light Source
    Alonso-Mori, Roberto
    Caronna, Chiara
    Chollet, Matthieu
    Curtis, Robin
    Damiani, Daniel S.
    Defever, Jim
    Feng, Yiping
    Flath, Daniel L.
    Glownia, James M.
    Lee, Sooheyong
    Lemke, Henrik T.
    Nelson, Silke
    Bong, Eric
    Sikorski, Marcin
    Song, Sanghoon
    Srinivasan, Venkat
    Stefanescu, Daniel
    Zhu, Diling
    Robert, Aymeric
    JOURNAL OF SYNCHROTRON RADIATION, 2015, 22 : 508 - 513
  • [6] Commissioning of a laser-plasma X-Ray micro-focus source for phase contrast imaging
    Martin, L.
    Benlliure, J.
    Cortina, D.
    Gonzalez, D.
    Llerena, J. J.
    Pena, J.
    Ruiz, C.
    RELATIVISTIC PLASMA WAVES AND PARTICLE BEAMS AS COHERENT AND INCOHERENT RADIATION SOURCES III, 2019, 11036
  • [7] A soft X-ray free-electron laser beamline at SACLA: the light source, photon beamline and experimental station
    Owada, Shigeki
    Togawa, Kazuaki
    Inagaki, Takahiro
    Hara, Toru
    Tanaka, Takashi
    Joti, Yasumasa
    Koyama, Takahisa
    Nakajima, Kyo
    Ohashi, Haruhiko
    Senba, Yasunori
    Togashi, Tadashi
    Tono, Kensuke
    Yamaga, Mitsuhiro
    Yumoto, Hirokatsu
    Yabashi, Makina
    Tanaka, Hitoshi
    Ishikawa, Tetsuya
    JOURNAL OF SYNCHROTRON RADIATION, 2018, 25 : 282 - 288
  • [8] The versatile X-ray beamline of the Munich Compact Light Source: design, instrumentation and applications
    Guenther, Benedikt
    Gradl, Regine
    Jud, Christoph
    Eggl, Elena
    Huang, Juanjuan
    Kulpe, Stephanie
    Achterhold, Klaus
    Gleich, Bernhard
    Dierolf, Martin
    Pfeiffer, Franz
    JOURNAL OF SYNCHROTRON RADIATION, 2020, 27 : 1395 - 1414
  • [9] Optimization of the soft X-ray transmission microscopy beamline at the ALBA light source
    Sorrentino, Andrea
    Pereiro, Eva
    Valcarcel, Ricardo
    Ferrer, Salvador
    Nicolas, Josep
    ADVANCES IN X-RAY/EUV OPTICS AND COMPONENTS VIII, 2013, 8848
  • [10] The X-ray Correlation Spectroscopy instrument at the Linac Coherent Light Source
    Robert, A.
    Curtis, R.
    Flath, D.
    Gray, A.
    Sikorski, M.
    Song, S.
    Srinivasan, V.
    Stefanescu, D.
    11TH INTERNATIONAL CONFERENCE ON SYNCHROTRON RADIATION INSTRUMENTATION (SRI 2012), 2013, 425