Toward 3D printed microfluidic artificial lungs for respiratory support

被引:4
作者
Fleck, Elyse [1 ,2 ]
Keck, Charlise [1 ,2 ]
Ryszka, Karolina [1 ,2 ]
Zhang, Andrew [1 ,2 ]
Atie, Michael [1 ,2 ]
Maddox, Sydney [1 ,2 ]
Potkay, Joseph [1 ,2 ]
机构
[1] Univ Michigan, Dept Surg, ECLS Lab, Ann Arbor, MI 48109 USA
[2] VA Ann Arbor Healthcare Syst, Ann Arbor, MI 48105 USA
基金
美国国家卫生研究院;
关键词
MATHEMATICAL-MODEL; GAS-EXCHANGE; POLY(DIMETHYLSILOXANE); PERFORMANCE; RHEOLOGY; DEVICES; FLOW; AIR;
D O I
10.1039/d3lc00814b
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Microfluidic artificial lungs (mu ALs) are a new class of membrane oxygenators. Compared to traditional hollow-fiber oxygenators, mu ALs closely mimic the alveolar microenvironment due to their size-scale and promise improved gas exchange efficiency, hemocompatibility, biomimetic blood flow networks, and physiologically relevant blood vessel pressures and shear stresses. Clinical translation of mu ALs has been stalled by restrictive microfabrication techniques that limit potential artificial lung geometries, overall device size, and throughput. To address these limitations, a high-resolution Asiga MAX X27 UV digital light processing (DLP) 3D printer and custom photopolymerizable polydimethylsiloxane (PDMS) resin were used to rapidly manufacture small-scale mu ALs via vat photopolymerization (VPP). Devices were designed in SOLIDWORKS with 500 blood channels and 252 gas channels, where gas and blood flow channels were oriented orthogonally and separated by membranes on the top and bottom, permitting two-sided gas exchange. Successful devices were post-processed to remove uncured resin from microchannels and assembled with external tubing in preparation for gas exchange performance testing with ovine whole blood. 3D printed channel dimensions were 172 mu m-tall x 320 mu m-wide, with 62 mu m-thick membranes and 124 mu m-wide support columns. Measured outlet blood oxygen saturation (SO2) agreed with theoretical models and rated flow of the device was 1 mL min-1. Blood side pressure drop was 1.58 mmHg at rated flow. This work presents the highest density of 3D printed microchannels in a single device, one of the highest CO2 transfer efficiencies of any artificial lung to date, and a promising approach to translate mu ALs one step closer to the clinic. Development and testing of a small-scale 3D printed microfluidic artificial lung (mu AL) fabricated via UV light based vat photopolymerization and a custom polydimethylsiloxane photopolymer resin.
引用
收藏
页码:955 / 965
页数:11
相关论文
共 66 条
[41]   Steel reinforced composite silicone membranes and its integration to microfluidic oxygenators for high performance gas exchange [J].
Matharoo, Harpreet ;
Dabaghi, Mohammadhossein ;
Rochow, Niels ;
Fusch, Gerhard ;
Saraei, Neda ;
Tauhiduzzaman, Mohammed ;
Veldhuis, Stephen ;
Brash, John ;
Fusch, Christoph ;
Selvaganapathy, P. Ravi .
BIOMICROFLUIDICS, 2018, 12 (01)
[42]   Poly(dimethylsiloxane) as a material for fabricating microfluidic devices [J].
McDonald, JC ;
Whitesides, GM .
ACCOUNTS OF CHEMICAL RESEARCH, 2002, 35 (07) :491-499
[43]   Tuning of Smart Multifunctional Polymer Coatings Made by Zwitterionic Phosphorylcholines [J].
Muench, Alexander S. ;
Adam, Stefan ;
Fritzsche, Tina ;
Uhlmann, Petra .
ADVANCED MATERIALS INTERFACES, 2020, 7 (01)
[44]  
nanoscribe, NEW IP PDMS PHOTORES
[45]   Spatially and optically tailored 3D printing for highly miniaturized and integrated microfluidics [J].
Noriega, Jose L. Sanchez ;
Chartrand, Nicholas A. ;
Valdoz, Jonard Corpuz ;
Cribbs, Collin G. ;
Jacobs, Dallin A. ;
Poulson, Daniel ;
Viglione, Matthew S. ;
Woolley, Adam T. ;
Van Ry, Pam M. ;
Christensen, Kenneth A. ;
Nordin, Gregory P. .
NATURE COMMUNICATIONS, 2021, 12 (01)
[46]  
olympus-ims, SURFACE ROUGHNESS ME
[47]   Study of separation force in constrained surface projection stereolithography [J].
Pan, Yayue ;
He, Haiyang ;
Xu, Jie ;
Feinerman, Alan .
RAPID PROTOTYPING JOURNAL, 2017, 23 (02) :353-361
[48]   Highly CO2-permeable and -selective membranes derived from crossfinked poly(ethylene glycol) and its nanocomposites [J].
Patel, NP ;
Miller, AC ;
Spontak, RJ .
ADVANCED FUNCTIONAL MATERIALS, 2004, 14 (07) :699-707
[49]   Stability of Polyethylene Glycol and Zwitterionic Surface Modifications in PDMS Microfluidic Flow Chambers [J].
Plegue, Thomas J. ;
Kovach, Kyle M. ;
Thompson, Alex J. ;
Potkay, Joseph A. .
LANGMUIR, 2018, 34 (01) :492-502
[50]   The promise of microfluidic artificial lungs [J].
Potkay, Joseph A. .
LAB ON A CHIP, 2014, 14 (21) :4122-4138