The novel magnesium-titanium hybrid cannulated screws for the treatment of vertical femoral neck fractures: Biomechanical evaluation

被引:6
|
作者
Zhang, Qida [1 ,6 ]
Chen, Zhenxian [2 ]
Peng, Yinghu [3 ]
Jin, Zhongmin [4 ,5 ]
Qin, Ling [1 ,6 ]
机构
[1] Chinese Univ Hong Kong, Dept Orthopaed & Traumatol, Musculoskeletal Res Lab, Hong Kong, Peoples R China
[2] Changan Univ, Sch Mech Engn, Key Lab Rd Construction Technol & Equipment, Minist Educ, Xian, Peoples R China
[3] Chinese Acad Sci, Shenzhen Inst Adv Technol, CAS Key Lab Human Machine Intelligence Synergy Sys, Shenzhen, Peoples R China
[4] Southwest Jiaotong Univ, Tribol Res Inst, Sch Mech Engn, Chengdu, Peoples R China
[5] Univ Leeds, Inst Med & Biol Engn, Sch Mech Engn, Leeds, England
[6] Chinese Univ Hong Kong, Prince Wales Hosp, Dept Orthopaed & Traumatol, Musculoskeletal Res Lab,Shatin, 30-32 Ngan Shing St, Hong Kong, Peoples R China
基金
中国国家自然科学基金;
关键词
Biomechanics; Cannulated screws; Finite element analysis; Musculoskeletal multibody model; Vertical femoral neck fractures; MECHANICAL-PROPERTIES; INTERNAL-FIXATION; TRABECULAR BONE; PREDICTION; DENSITY; LOAD; TRANSLATION; FORCE;
D O I
10.1016/j.jot.2023.08.003
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Conventional cannulated screws are commonly used for internal fixation in the treatment of vertical femoral neck fractures. However, the noticeably high rates of undesirable outcomes such as nonunion, malunion, avascular necrosis, and fixation failure still troubled the patients and surgeons. It is urgent to develop new cannulated screws to improve the above clinical problems. The purpose of this study was to design a novel magnesium-titanium hybrid cannulated screw and to further evaluate its biomechanical performance for the treatment of vertical femoral neck fractures. Methods: A novel magnesium-titanium hybrid cannulated screw was designed, and the conventional titanium cannulated screw was also modeled. The finite element models for vertical femoral neck fractures with magnesium-titanium hybrid cannulated screws and conventional cannulated screws were respectively established. The hip joint contact force during walking gait calculated by a subject-specific musculoskeletal multibody dynamics model, was used as loads and boundary conditions for both finite element models. The stress and displacement distributions of the cannulated screws and the femur, the micromotion of the fracture surfaces of the femoral neck, and the overall stiffness were calculated and analyzed using finite element models. The biomechanical performance of the Magnesium-Titanium hybrid cannulated screws was evaluated. Results: The maximum stresses of the magnesium-titanium hybrid cannulated screws and the conventional cannulated screws were 451.5 MPa and 476.8 MPa, respectively. The maximum stresses of the femur with the above different cannulated screws were 140.3 MPa and 164.8 MPa, respectively. The maximum displacement of the femur with the hybrid cannulated screws was 6.260 mm, lower than the femur with the conventional cannulated screws, which was 7.125 mm. The tangential micromotions in the two orthogonal directions at the fracture surface of the femoral neck with the magnesium-titanium hybrid cannulated screws were comparable to those with the conventional cannulated screws. The overall stiffness of the magnesium-titanium hybrid cannulated screw system was 490.17 N/mm, higher than that of the conventional cannulated screw system, which was 433.92 N/mm. Conclusion: The magnesium-titanium hybrid cannulated screw had superior mechanical strength and fixation stability for the treatment of the vertical femoral neck fractures, compared with those of the conventional cannulated screw, indicating that the magnesium-titanium hybrid cannulated screw has great potential as a new fixation strategy in future clinical applications. The translational potential of this article: This study highlights an innovative design of the magnesium-titanium hybrid cannulated screw for the treatment of vertical femoral neck fractures. The novel magnesium-titanium
引用
收藏
页码:127 / 136
页数:10
相关论文
共 50 条
  • [1] Titanium alloy cannulated screws and biodegradable magnesium alloy bionic cannulated screws for treatment of femoral neck fractures: a finite element analysis
    Kai Ding
    Weijie Yang
    Jian Zhu
    Xiaodong Cheng
    Haicheng Wang
    Du Hao
    Song Yinuo
    Yanbin Zhu
    Yingze Zhang
    Wei Chen
    Qi Zhang
    Journal of Orthopaedic Surgery and Research, 16
  • [2] Titanium alloy cannulated screws and biodegradable magnesium alloy bionic cannulated screws for treatment of femoral neck fractures: a finite element analysis
    Ding, Kai
    Yang, Weijie
    Zhu, Jian
    Cheng, Xiaodong
    Wang, Haicheng
    Hao, Du
    Song Yinuo
    Zhu, Yanbin
    Zhang, Yingze
    Chen, Wei
    Zhang, Qi
    JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2021, 16 (01)
  • [3] Biomechanical Evaluation of the Femoral Neck System in Unstable Pauwels III Femoral Neck Fractures: A Comparison with the Dynamic Hip Screw and Cannulated Screws
    Stoffel, Karl
    Zderic, Ivan
    Gras, Florian
    Sommer, Christoph
    Eberli, Ursula
    Mueller, David
    Oswald, Martin
    Gueorguiev, Boyko
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2017, 31 (03) : 131 - 137
  • [4] Titanium alloy cannulated screws and biodegrade ceramic nails for treatment of femoral neck fractures: A finite element analysis
    Liu, Yang
    Ren, Zhiyuan
    Shao, Huifeng
    Wang, Xueding
    Ma, Yongsheng
    Song, Wenjie
    Wu, Xiaogang
    Zhang, Xiangyu
    Li, Pengcui
    He, Yong
    Wei, Xiaochun
    Duan, Wangping
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2023, 54 (08):
  • [5] Biomechanical and clinical evaluation of interlocking hip screw in Pauwels III femoral neck fractures: A comparison with inverted triangle cannulated screws
    Zhang, Jian
    Jiang, Haozheng
    Dai, Wei
    Hersi, Salad Abdirahman
    Cheong, John Chun Tien Chui Wan
    Chu, Zhenchen
    Lou, Zhiyuan
    Zhang, Deqiang
    Liu, Changjian
    Tian, Kang
    Tang, Xin
    FRONTIERS IN BIOENGINEERING AND BIOTECHNOLOGY, 2022, 10
  • [6] Proximal Femoral Locking Plate With Cannulated Screws for the Treatment of Femoral Neck Fractures
    Lin, Dasheng
    Lian, Kejian
    Ding, Zhenqi
    Zhai, Wenliang
    Hong, Jiayuan
    ORTHOPEDICS, 2012, 35 (01) : E1 - E5
  • [7] Comparison of femoral neck system and three cannulated cancellous screws in the treatment of vertical femoral neck fractures: clinical observation and finite element analysis
    Huang, Shaolong
    Zhang, Yazhong
    Zhang, Xu
    Zhou, Chengqiang
    Li, Wenbo
    Wang, Yunqing
    Wang, Bin
    Zhu, Ziqiang
    BIOMEDICAL ENGINEERING ONLINE, 2023, 22 (01)
  • [8] Biomechanical comparison of femoral neck system and cannulated screws coupled with medial plate for treating Pauwels III femoral neck fractures
    Nan, Chong
    Li, Yanjun
    Liu, Yuxiu
    Ma, Liang
    Ma, Zhanbei
    TECHNOLOGY AND HEALTH CARE, 2023, 31 (04) : 1161 - 1170
  • [9] Cannulated Screws or Hemiarthroplasty for Femoral Neck Fractures: Is There a Mortality Difference?
    Thompson, Austen L.
    Kuttner, Nicolas P.
    Greenberg, Marc
    Hidden, Krystin A.
    Yuan, Brandon J.
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2024, 38 (11) : 596 - 601
  • [10] Biomechanical analysis of a new cannulated screw for unstable femoral neck fractures
    Chen, Zhigang
    Chen, Feiyang
    Xu, Xinbao
    Li, Xin
    Cui, Haidong
    Zhang, Wen
    Jiang, Dong
    Zhang, Feixiang
    Chen, Yinbing
    Zhou, Shiran
    Lyu, Shujun
    FRONTIERS IN BIOENGINEERING AND BIOTECHNOLOGY, 2024, 12