Reliability, Concurrent Validity, and Minimal Detectable Change of a Smartphone Application for Measuring Thoracic Kyphosis

被引:1
作者
Sahinoglu, Ertan [1 ]
Ergin, Gulbin [2 ]
Bakirhan, Serkan [3 ]
Unver, Bayram [4 ]
机构
[1] Mavi Physiotherapy, Izmir, Turkiye
[2] Izmir Bakircay Univ, Dept Phys Therapy & Rehabil, Fac Hlth Sci, Izmir, Turkiye
[3] Ege Univ, Dept Phys Therapy & Rehabil, Fac Hlth Sci, Izmir, Turkiye
[4] Dokuz Eylul Univ, Fac Phys Therapy & Rehabil, Izmir, Turkiye
关键词
Reliability; validity; kyphosis; smartphone; spine; INTRARATER RELIABILITY; STATISTICAL-METHODS; MOTION; RANGE; SPINE; OSTEOPOROSIS; INTERRATER; AGREEMENT; MOBILITY; POSTURE;
D O I
10.33808/clinexphealthsci.1038122
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
Objective: To assess the intra - and inter-rater reliability and concurrent validity, and to estimate minimal detectable change of a smartphone application for measuring thoracic kyphosis angle. Methods: A total of 80 healthy university students were evaluated. Two raters measured the thoracic kyphosis angle using a digital inclinometer and the smartphone application. Intra - and inter-rater reliability were assessed using the intraclass correlation coefficient (ICC) with 95% confidence interval. The standard error of measurement (SEM) and the minimal detectable change at the 95% confidence level (MDC95) were also calculated. The concurrent validity between the digital inclinometer and the smartphone application was assessed by the linear regression analysis and Bland and Altman's 95% limits of agreement method. Results: The intra - and inter-rater reliability were excellent for the digital inclinometer and the smartphone application (ICC > 0.75). The SEM values for the digital inclinometer and the smartphone application were close together. The MDC95 values for the smartphone application were 5.11 and 6.30 degrees, and 9.02 degrees for intra - and inter-rater, respectively. The digital inclinometer and the smartphone application showed a positive correlation (R-2 = 0.85). The Bland-Altman plot showed a good agreement between the instruments. Conclusion: The smartphone application used in this study is a cost-effective, practical, reliable, and valid instrument for measuring the thoracic kyphosis angle. More than 9 degrees in the value of the thoracic kyphosis angle measured by the smartphone application can be considered as a true change.
引用
收藏
页码:511 / 516
页数:6
相关论文
共 34 条
[1]  
Azadinia F, 2014, J SPINAL DISORD TECH, V27, pE212
[2]   Is thoracic spine posture associated with shoulder pain, range of motion and function? A systematic review [J].
Barrett, Eva ;
O'Keeffe, Mary ;
O'Sullivan, Kieran ;
Lewis, Jeremy ;
McCreesh, Karen .
MANUAL THERAPY, 2016, 26 :38-46
[3]   Reliability and validity of non-radiographic methods of thoracic kyphosis measurement: A systematic review [J].
Barrett, Eva ;
McCreesh, Karen ;
Lewis, Jeremy .
MANUAL THERAPY, 2014, 19 (01) :10-17
[4]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[5]   Radiographic measures of thoracic kyphosis in osteoporosis: Cobb and vertebral centroid angles [J].
Briggs, A. M. ;
Wrigley, T. V. ;
Tully, E. A. ;
Adams, P. E. ;
Greig, A. M. ;
Bennell, K. L. .
SKELETAL RADIOLOGY, 2007, 36 (08) :761-767
[6]   Reliability and concurrent validity of a Smartphone, bubble inclinometer and motion analysis system for measurement of hip joint range of motion [J].
Charlton, Paula C. ;
Mentiplay, Benjamin F. ;
Pua, Yong-Hao ;
Clark, Ross A. .
JOURNAL OF SCIENCE AND MEDICINE IN SPORT, 2015, 18 (03) :262-267
[7]  
Cicchetti D.V., 1994, Psychological Assessment, V6, P284, DOI [DOI 10.1037/1040-3590.6.4.284, 10.1037/1040-3590.6.4.284]
[8]   Minimal changes in health status questionnaires: distinction between minimally detectable change and minimally important change [J].
de Vet, Henrica C. ;
Terwee, Caroline B. ;
Ostelo, Raymond W. ;
Beckerman, Heleen ;
Knol, Dirk L. ;
Bouter, Lex M. .
HEALTH AND QUALITY OF LIFE OUTCOMES, 2006, 4 (1)
[9]   STATISTICAL METHODOLOGY FOR THE CONCURRENT ASSESSMENT OF INTERRATER AND INTRARATER RELIABILITY - USING GONIOMETRIC MEASUREMENTS AS AN EXAMPLE [J].
ELIASZIW, M ;
YOUNG, SL ;
WOODBURY, MG ;
FRYDAYFIELD, K .
PHYSICAL THERAPY, 1994, 74 (08) :777-788
[10]   Goniometric evaluation of the spinal sagittal curves in children and adolescents: A reliability study [J].
Gravina, Aristide ;
Ferraro, Claudio ;
Poli, Patrizia ;
Barazzuol, Michele ;
Del Felice, Alessandra ;
Masiero, Stefano .
JOURNAL OF BACK AND MUSCULOSKELETAL REHABILITATION, 2017, 30 (02) :325-331