Agreement of axial length and anterior segment parameters measured with the MYAH device compared to Pentacam AXL and IOLMaster 700 in myopic children

被引:12
作者
Sabur, Huri [1 ]
Takes, Omer [2 ]
机构
[1] Univ Hlth Sci, Diskapi Yildirim Beyazit Training & Res Hosp, Dept Ophthalmol, Ankara, Turkey
[2] Kaskaloglu Eye Hosp, Izmir, Turkey
关键词
Myopia; Axial length; Optical low coherence interferometry; Myah; PARTIAL-COHERENCE INTERFEROMETRY; SOURCE OPTICAL BIOMETRY; REPEATABILITY; TOMOGRAPHY; CURVATURE; OCT;
D O I
10.1007/s10792-022-02444-w
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose To compare the difference and agreement of axial length (AL) and anterior segment parameters obtained from the MYAH device with Pentacam AXL and IOLMaster 700 in myopic children. Methods The study included 60 eyes of 60 myopic children. AL, keratometry (K), and horizontal corneal diameter (CD) were measured with Pentacam AXL, IOLMaster 700, and MYAH, respectively. The Friedman test was used to assess the differences. The Intraclass Correlation Coefficient (ICC) and Bland-Altman plots were used to assess the consistency of measurements. Results The mean age was 10.2 +/- 1.8 years (7-16 years). No statistically significant difference was determined between the Pentacam AXL, IOLMaster 700, and MYAH devices in terms of mean AL values (23.61 +/- 1.42, 23.62 +/- 1.45, 23.61 +/- 1.42, respectively) (p = 0.06). The difference between devices in the mean steep K, flat K, mean K, and CD was statistically significant but clinically insignificant (steep K; 44.45 +/- 1.25, 44.59 +/- 1.23, 44.51 +/- 1.24, flat K; 43.29 +/- 1.28, 43.43 +/- 1.29, 43.35 +/- 1.30, mean K; 43.85 +/- 1.21, 44.00 +/- 1.19, 43.94 +/- 1.20, and CD; 11.90 +/- 0.34, 12.11 +/- 0.38, 11.96 +/- 0.31, respectively; p < 0.05). ICC and Bland-Altman plot analysis revealed a high correlation between the three devices in AL, steep K, flat K, mean K, and CD measurements. Conclusion There was a quite good agreement between the MYAH, Pentacam AXL, and IOLMaster 700 devices regarding AL and anterior segment parameters. MYAH provides reliable measurements and will be a good option in the diagnosis of and follow-up with myopic children.
引用
收藏
页码:475 / 482
页数:8
相关论文
共 20 条
[1]   Effectiveness and agreement of 3 optical biometers in measuring axial length in the eyes of patients with mature cataracts [J].
Alejandra Henriquez, Maria ;
Zuniga, Raul ;
Camino, Maythe ;
Camargo, Jorge ;
Ruiz-Montenegro, Katia ;
Izquierdo Jr, Luis .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2020, 46 (09) :1222-1228
[2]   Efficiency and measurements agreement between swept-source OCT and low-coherence interferometry biometry systems [J].
Calvo-Sanz, Jorge A. ;
Portero-Benito, Alejandro ;
Arias-Puente, Alfonso .
GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2018, 256 (03) :559-566
[3]   Comparison of Biometry Measurements Using Standard Partial Coherence Interferometry versus New Scheimpflug Tomography with Integrated Axial Length Capability [J].
Haddad, Jorge Selem ;
Barnwell, Eliza ;
Rocha, Karolinne Maia ;
Ambrosio Jr, Renato ;
Waring, George Oral .
CLINICAL OPHTHALMOLOGY, 2020, 14 :353-358
[4]   Comparison of a new optical biometer using swept-source optical coherence tomography and a biometer using optical low-coherence reflectometry [J].
Hoffer, Kenneth J. ;
Hoffmann, Peter C. ;
Savini, Giacomo .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2016, 42 (08) :1165-1172
[5]   Multicenter study of optical low-coherence interferometry and partial-coherence interferometry optical biometers with patients from the United States and China [J].
Hoffer, Kenneth J. ;
Shammas, H. John ;
Savini, Giacomo ;
Huang, Jinhai .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2016, 42 (01) :62-67
[6]   Precision (repeatability and reproducibility) of ocular parameters obtained by the Tomey OA-2000 biometer compared to the IOLMaster in healthy eyes [J].
Hua, Yanjun ;
Qiu, Wei ;
Xiao, Qiuyi ;
Wu, Qiang .
PLOS ONE, 2018, 13 (02)
[7]   The Relationship between Progression in Axial Length/Corneal Radius of Curvature Ratio and Spherical Equivalent Refractive Error in Myopia [J].
Jong, Monica ;
Sankaridurg, Padmaja ;
Naduvilath, Thomas John ;
Li, Wayne ;
He, Mingguang .
OPTOMETRY AND VISION SCIENCE, 2018, 95 (10) :921-929
[8]   Axial length measurement using partial coherence interferometry in myopic children: Repeatability of the measurement and comparison with refractive components [J].
Kimura, Shuhei ;
Hasebe, Satoshi ;
Miyata, Manabu ;
Hamasaki, Ichiro ;
Ohtsuki, Hiroshi .
JAPANESE JOURNAL OF OPHTHALMOLOGY, 2007, 51 (02) :105-110
[9]   Evaluation of 6 biometers based on different optical technologies [J].
Montes-Mico, Robert .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2022, 48 (01) :16-25
[10]   Agreement Between Swept-Source Optical Biometry and Scheimpflug-based Topography Measurements of Anterior Segment Parameters [J].
Ozyol, Pelin ;
Ozyol, Erhan .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2016, 169 :73-78