Enhancing flexibility and strength-to-weight ratio of polymeric stents: A new variable-thickness design approach

被引:5
作者
Khatami, Mohamad [1 ]
Doniavi, Ali [1 ]
Abazari, Amir Musa [1 ]
Fotouhi, Mohammad [2 ]
机构
[1] Urmia Univ, Fac Engn, Dept Mech Engn, Orumiyeh, Iran
[2] Delft Univ Technol, Fac Civil Engn & Geosci, Dept Mat Mech Management & Design 3MD, Delft, Netherlands
关键词
Stent; Variable-thickness; Flexibility; 3D printing; Finite element method; DRUG-ELUTING STENTS; CORONARY-ARTERY; PERFORMANCE; OPTIMIZATION; STRATEGIES; BEHAVIOR;
D O I
10.1016/j.jmbbm.2023.106262
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This paper presents a new design strategy to improve the flexibility and strength-to-weight ratio of polymeric stents. The proposed design introduces a variable-thickness (VT) stent that outperforms conventional polymeric stents with constant thickness (CT). While polymeric stents offer benefits like flexibility and bioabsorption, their mechanical strength is lower compared to metal stents. To address this limitation, thicker polymer stents are used, compromising flexibility and clinical performance. Leveraging advancements in 3D printing, a new design approach is introduced in this study and is manufactured by the Liquid Crystal Display (LCD) 3D printing method and PLA resin. The mechanical performance of CT and VT stents is compared using the Finite Element Method (FEM), validated by experimental tests. Results demonstrate that the VT stent offers significant improvements compared to a CT stent in bending stiffness (over 20%), reduced plastic strain distribution of expansion (over 26%), and increased radial strength (over 10%). This research showcases the potential of the VT stent design to enhance clinical outcomes and patient care.
引用
收藏
页数:10
相关论文
共 52 条
[1]   Bioresorbable stents: Current and upcoming bioresorbable technologies [J].
Ang, Hui Ying ;
Bulluck, Heerajnarain ;
Wong, Philip ;
Venkatraman, Subbu S. ;
Huang, Yingying ;
Foin, Nicolas .
INTERNATIONAL JOURNAL OF CARDIOLOGY, 2017, 228 :931-939
[2]   Numerical investigations of the structural behavior of a balloon expandable stent design using finite element method [J].
Azaouzi, M. ;
Makradi, A. ;
Belouettar, S. .
COMPUTATIONAL MATERIALS SCIENCE, 2013, 72 :54-61
[3]   Mechanical properties of the drug-eluting bioresorbable magnesium scaffold compared with polymeric scaffolds and a permanent metallic drug-eluting stent [J].
Barkholt, Trine O. ;
Webber, Bruce ;
Holm, Niels R. ;
Ormiston, John A. .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2020, 96 (07) :E674-E682
[4]   Computational Bench Testing to Evaluate the Short-Term Mechanical Performance of a Polymeric Stent [J].
Bobel, A. C. ;
Petisco, S. ;
Sarasua, J. R. ;
Wang, W. ;
McHugh, P. E. .
CARDIOVASCULAR ENGINEERING AND TECHNOLOGY, 2015, 6 (04) :519-532
[5]   In vivo and in vitro evaluation of a biodegradable magnesium vascular stent designed by shape optimization strategy [J].
Chen, Chenxin ;
Chen, Jiahui ;
Wu, Wei ;
Shi, Yongjuan ;
Jin, Liang ;
Petrini, Lorenza ;
Shen, Li ;
Yuan, Guangyin ;
Ding, Wenjiang ;
Ge, Junbo ;
Edelman, Elazer R. ;
Migliavacca, Francesco .
BIOMATERIALS, 2019, 221
[6]   Flexibility of Biodegradable Polymer Stents with Different Strut Geometries [J].
Chen, Chong ;
Xiong, Yan ;
Li, Zhongyou ;
Chen, Yu .
MATERIALS, 2020, 13 (15)
[7]   Polymer-Free Drug-Eluting Stents: An Overview of Coating Strategies and Comparison with Polymer-Coated Drug-Eluting Stents [J].
Chen, Weiluan ;
Habraken, Tom C. J. ;
Hennink, Wim E. ;
Kok, Robbert J. .
BIOCONJUGATE CHEMISTRY, 2015, 26 (07) :1277-1288
[8]   Finite element analysis of the mechanical performance of a two-layer polymer composite stent structure [J].
Donik, Z. ;
Necemer, B. ;
Glodez, S. ;
Kramberger, J. .
ENGINEERING FAILURE ANALYSIS, 2022, 137
[9]   Temporary scaffolding of coronary arteries with bioabsorbable magnesium stents:: a prospective, non-randomised multicentre trial [J].
Erbel, Raimund ;
Di Mario, Carlo ;
Bartunek, Jozef ;
Bonnier, Johann ;
de Bruyne, Bernard ;
Eberli, Franz R. ;
Erne, Paul ;
Haude, Michael ;
Heublein, Bernd ;
Horrigan, Mark ;
Ilsley, Charles ;
Boese, Dirk ;
Koolen, Jacques ;
Luescher, Thomas F. ;
Weissman, Neil ;
Waksman, Ron .
LANCET, 2007, 369 (9576) :1869-1875
[10]   Stent flexibility versus concertina effect: Mechanism of an unpleasant trade-off in stent design and its implications for stent selection in the cath-lab [J].
Foin, N. ;
Di Mario, C. ;
Francis, D. P. ;
Davies, J. E. .
INTERNATIONAL JOURNAL OF CARDIOLOGY, 2013, 164 (03) :259-261