Establishing local diagnostic reference levels for computed tomography examinations using size-specific dose estimates

被引:6
|
作者
Alrehily, Faisal A. [1 ]
Alqahtani, Khalid S. [1 ]
Aljohani, Marwan H. [1 ]
Alharbi, Nawaf S. [1 ]
Alharbi, Reyan M. [2 ]
Abdulaal, Osamah M. [1 ]
Alshamrani, Abdullah F. [1 ]
Alsaedi, Abdulrahman S. [3 ]
Al-Murshedi, Sadeq H. [4 ,5 ]
Alhazmi, Fahad H. [1 ]
机构
[1] Taibah Univ, Dept Diag Radiol Technol, Coll Appl Med Sci, Al Madinah Al Munawwarah, Saudi Arabia
[2] King Salman bin Abdulaziz Med City, Dept Radiol, Riyadh, Saudi Arabia
[3] King Fahad Gen Hosp, Dept Nucl Med, Al Madinah Al Munawwarah, Saudi Arabia
[4] AL Zahraa Univ Women, Coll Hlth & Med Technol, Dept Radiol Tech, Karbala, Iraq
[5] Univ Babylon, Coll Educ Pure Sci, Phys Dept, Babylon, Iraq
关键词
CT; DRL; CTDIvol; DLP; SSDE; effective diameter; CT; RADIATION; WEIGHT; CANCER; CHEST;
D O I
10.15537/smj.2023.44.8.20230230
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objectives: To establish local DRL (LDRL) for computed tomography (CT) examinations based on size-specific dose estimates (SSDEs), which consider patient size. The concept of diagnostic reference level (DRL) was introduced to limit patient exposure to unnecessary radiation. However, traditional DRL values do not consider patient size. Methods: Following institutional committee approval, data were collected from CT examinations of adult patients at Madinah General Hospital, Al Madinah Al Munawwarah, Saudi Arabia from January to March 2023. The SSDE was calculated for each patient using the effective diameter (D-eff). Results: The LDRLs of the brain, cervical spine, chest, thoracic spine and kidneys, ureters, and bladder (KUB) examinations were 118 mGy, 12 mGy, 8 mGy, 17 mGy, and 7 mGy, respectively. A strong correlation was observed between SSDEs and the volume computed tomography dose index (CTDIvol) for all examinations except chest scans (p<0.05). Size-specific dose estimates were higher than the CTDIvol, with a greater difference for patients with smaller Deff (p<0.05). Conclusion: The established LDRL was within the international DRL. The use of SSDE has the potential to provide more accurate and relevant data for radiation safety practices; however, widespread adoption of SSDE in new CT scanners is necessary for promoting consistency and standardization methodologies.
引用
收藏
页码:761 / 766
页数:6
相关论文
共 50 条
  • [31] Diagnostic reference levels for common computed tomography (CT) examinations: results from the first Nigerian nationwide dose survey
    Ekpo, Ernest U.
    Adejoh, Thomas
    Akwo, Judith D.
    Emeka, Owujekwe C.
    Modu, Ali A.
    Abba, Mohammed
    Adesina, Kudirat A.
    Omiyi, David O.
    Chiegwu, Uche H.
    JOURNAL OF RADIOLOGICAL PROTECTION, 2018, 38 (02) : 525 - 535
  • [32] Establishment of diagnostic reference levels in cardiac computed tomography
    Rawashdeh, Mohammad
    Saade, Charbel
    Zaitoun, Maha
    Abdelrahman, Mostafa
    Brennan, Patrick
    Alewaidat, Haytham
    McEntee, Mark F.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2019, 20 (10): : 181 - 186
  • [33] Accuracy of patient dose estimation in cone beam computed tomography in breast irradiation by size-specific dose estimates with position correction
    Ueno, Hiroyuki
    Matsubara, Kosuke
    Bou, Sayuri
    Hizume, Masato
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2022, 23 (12):
  • [34] Size-Specific Dose Estimates Based on Water-Equivalent Diameter and Effective Diameter in Computed Tomography Coronary Angiography
    Xu, Jian
    Wang, Xiangquan
    Xiao, Huawei
    Xu, Jianguo
    MEDICAL SCIENCE MONITOR, 2019, 25 : 9299 - 9305
  • [35] Factors influencing size-specific dose estimates of selected computed tomography protocols at two clinical practices in South Africa
    Sebelego, Ida-Keshia
    Acho, Sussan
    van der Merwe, Belinda
    Rae, William I. D.
    RADIATION PROTECTION DOSIMETRY, 2023, 199 (07) : 588 - 602
  • [36] Diagnostic Reference Level of Paediatric Computed Tomography Brain Examinations
    Abdelaziz, Ikhlas
    BIOSCIENCE RESEARCH, 2021, 18 (01): : 477 - 487
  • [37] Size-specific dose estimates calculated using patient size measurements from scanned projection radiograph in high-resolution chest computed tomography
    Brindhaban, Ajit
    JOURNAL OF MEDICAL RADIATION SCIENCES, 2025, 72 (01) : 85 - 92
  • [38] Patient size as a parameter for determining Diagnostic Reference Levels for paediatric Computed Tomography (CT) procedures
    Satharasinghe, Duminda
    Jeyasugiththan, Jeyasingam
    Wanninayake, W. M. N. M. B.
    Pallewatte, A. S.
    Samarasinghe, R. A. N. K. K.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2022, 102 : 55 - 65
  • [39] Diagnostic reference levels of paediatric computed tomography examinations performed at a dedicated Australian paediatric hospital
    Bibbo, Giovanni
    Brown, Scott
    Linke, Rebecca
    JOURNAL OF MEDICAL IMAGING AND RADIATION ONCOLOGY, 2016, 60 (04) : 475 - 484
  • [40] Establishment of diagnostic reference levels for common multi-detector computed tomography examinations in Iran
    Najafi, Mohsen
    Deevband, Mohammad Reza
    Ahmadi, Mohsen
    Kardan, Mohammad Reza
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2015, 38 (04) : 603 - 609