An Improved Noise Reduction Technique for Enhancing the Intelligibility of Sinewave Vocoded Speech: Implication in Cochlear Implants

被引:3
|
作者
Poluboina, Venkateswarlu [1 ]
Pulikala, Aparna [1 ]
Muthu, Arivudai Nambi Pitchai [2 ,3 ]
机构
[1] Natl Inst Technol Karnataka, Dept Elect & Commun Engn, Surathkal 575025, Mangaluru, India
[2] Kasturba Med Coll & Hosp, Dept Audiol & Speech Language Pathol, Mangaluru 575001, India
[3] Manipal Acad Higher Educ, Manipal 576104, India
关键词
Noise measurement; Cochlear implants; Signal to noise ratio; Speech enhancement; Wiener filters; Vocoders; Speech recognition; Noise reduction; speech recognition; cochlear implant; vocoder simulation; PERFORMANCE; ALGORITHMS; MASKING; HEARING; STIMULATION; SUPPRESSION; COMPLEX;
D O I
10.1109/ACCESS.2022.3232549
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
A cochlear implant (CI) is the most suitable option for individuals with severe profound hearing loss. CI restores the audibility to near perfection and offers good speech understanding in quiet. However, the speech perception in noise with CIs is less optimal as most speech coding strategies of CIs encode only the temporal envelope. Besides the current CI signal coding strategies lacks sophisticated pre-processing. In the current study, we proposed a novel pre-processing method to improve speech Intelligibility in noise and tested using the acoustic simulations of cochlear implants. The proposed noise reduction technique aims to minimize the mean square error (MSE) between the temporal envelopes of the enhanced speech and its clean speech. Therefore, the proposed method will be suitable for CI applications. This paper provides an analysis of the theoretical derivation of the noise suppression function and also the performance evaluation using objective and subjective tests. The effectiveness of the proposed method was objectively evaluated using the SRMR-CI and ESTOI. Additionally, speech recognition through the acoustic simulations of the cochlear implant was done for the subjective evaluation. Performance of the proposed method was compared with the Weiner filter (WF) and sigmoidal functions. The sinewave vocoder was used to simulate the cochlear implant perception. Both objective and subjective scores revealed that the performance of the proposed technique is superior to the WF and sigmoidal function.
引用
收藏
页码:787 / 796
页数:10
相关论文
共 50 条
  • [1] Evaluation of Thai Speech Intelligibility Based on Noise Reduction Techniques for Cochlear Implants
    Dachasilaruk, Siriporn
    Luangsawang, Chadarthan
    Jantharamin, Niphat
    Kaewsri, Siraprabha
    Doungta, Patamawadee
    2017 10TH INTERNATIONAL CONGRESS ON IMAGE AND SIGNAL PROCESSING, BIOMEDICAL ENGINEERING AND INFORMATICS (CISP-BMEI), 2017,
  • [2] LEVERAGING AUTOMATIC SPEECH RECOGNITION IN COCHLEAR IMPLANTS FOR IMPROVED SPEECH INTELLIGIBILITY UNDER REVERBERATION
    Hazrati, Oldooz
    Ghaffarzadegan, Shabnam
    Hansen, John H. L.
    2015 IEEE INTERNATIONAL CONFERENCE ON ACOUSTICS, SPEECH, AND SIGNAL PROCESSING (ICASSP), 2015, : 5093 - 5097
  • [3] Effects of envelope filter cutoff frequency on the intelligibility of Mandarin noise-vocoded speech in babble noise: Implications for cochlear implants
    Mai, Guangting
    Minett, James W.
    Wang, William S-Y.
    14TH ANNUAL CONFERENCE OF THE INTERNATIONAL SPEECH COMMUNICATION ASSOCIATION (INTERSPEECH 2013), VOLS 1-5, 2013, : 3613 - 3617
  • [4] An Optimal Envelope-Based Noise Reduction Method for Cochlear Implants: An Upper Bound Performance Investigation
    Chiea, Rafael A.
    Costa, Marcio H.
    Cordioli, Julio A.
    IEEE-ACM TRANSACTIONS ON AUDIO SPEECH AND LANGUAGE PROCESSING, 2021, 29 : 1729 - 1739
  • [5] Deep Learning-Based Noise Reduction Approach to Improve Speech Intelligibility for Cochlear Implant Recipients
    Lai, Ying-Hui
    Tsao, Yu
    Lu, Xugang
    Chen, Fei
    Su, Yu-Ting
    Chen, Kuang-Chao
    Chen, Yu-Hsuan
    Chen, Li-Ching
    Li, Lieber Po-Hung
    Lee, Chin-Hui
    EAR AND HEARING, 2018, 39 (04) : 795 - 809
  • [6] Speech Intelligibility for Cochlear Implant Users with the MMSE Noise-Reduction Time-Frequency Mask
    Mourao, Gustavo Leao
    Costa, Marcio Holsbach
    Paul, Stephan
    BIOMEDICAL SIGNAL PROCESSING AND CONTROL, 2020, 60 (60)
  • [7] The effect of gain thresholds on speech intelligibility for statistical model based noise reduction for cochlear implants: A simulation based verification
    He, Wenzhi
    Zheng, Nengheng
    Meng, Qinglin
    2016 10TH INTERNATIONAL SYMPOSIUM ON CHINESE SPOKEN LANGUAGE PROCESSING (ISCSLP), 2016,
  • [8] A Deep Denoising Autoencoder Approach to Improving the Intelligibility of Vocoded Speech in Cochlear Implant Simulation
    Lai, Ying-Hui
    Chen, Fei
    Wang, Syu-Siang
    Lu, Xugang
    Tsao, Yu
    Lee, Chin-Hui
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2017, 64 (07) : 1568 - 1578
  • [9] Effects of envelope bandwidth on the intelligibility of sine- and noise-vocoded speech
    Souza, Pamela
    Rosen, Stuart
    JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2009, 126 (02) : 792 - 805
  • [10] Effects of electrode separation between speech and noise signals on consonant identification in cochlear implants
    Kwon, Bom Jun
    JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2009, 126 (06) : 3258 - 3267