Design and simulation of control and sampling circuits for an open-path Fourier transform infrared spectrometer based on a rotating mirror

被引:0
|
作者
Zhu, Junfeng [1 ]
Li, Chengrui [1 ]
Chen, Yihai [1 ]
Du, Zhiqiang [1 ]
机构
[1] Sci & Technol Elect Test & Measurement Lab, Qingdao, Peoples R China
基金
中国国家自然科学基金;
关键词
Fourier transform; open-path spectrometer; rotary mirrors; sampling circuits; SPECTROSCOPY; INTERFEROMETERS;
D O I
10.1088/1402-4896/ad1e48
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
We designed a laser pulse sampling circuit to achieve equal optical path difference sampling for an open-path Fourier transform infrared (OP-FTIR) spectrometer based on a rotating mirror. The trigger signals for the equal optical path difference sampling were successfully obtained through circuit simulations, and an intelligent control circuit based on a 52 single-chip microcomputer (SCM) was used to accurately determine the sampling starting point and range for the spectrometer. By processing the grating scale signal and the grating reference signal, the sampling starting point and range of the spectrometer were determined precisely. Finally, the sampling trigger signals required to accomplish equal optical path difference sampling of the infrared spectrum were successfully outputted within the defined sampling range.
引用
收藏
页数:10
相关论文
共 8 条
  • [1] Fourier Transform Spectrometer Based on Rotating Parallel-Mirror-Pair
    Zhao Bao-wei
    Xiangli Bin
    Cai Qi-sheng
    Lu Qun-bo
    Zhou Jin-song
    SPECTROSCOPY AND SPECTRAL ANALYSIS, 2015, 35 (11) : 3209 - 3213
  • [2] AUTOMATED DETECTION OF METHANOL VAPOR BY OPEN-PATH FOURIER-TRANSFORM INFRARED SPECTROMETRY
    BANGALORE, AS
    SMALL, GW
    COMBS, RJ
    KNAPP, RB
    KROUTIL, RT
    ANALYTICA CHIMICA ACTA, 1994, 297 (03) : 387 - 403
  • [3] Advances in Data Processing for Open-Path Fourier Transform Infrared Spectrometry of Greenhouse Gases
    Shao, Limin
    Griffiths, Peter R.
    Leytem, April B.
    ANALYTICAL CHEMISTRY, 2010, 82 (19) : 8027 - 8033
  • [4] Measurements of Stable Isotopes in Atmospheric CO2 and H2O by Open-Path Fourier Transform Infrared Spectrometry
    Wang Wei
    Liu Wen-qing
    Zhang Tian-shu
    SPECTROSCOPY AND SPECTRAL ANALYSIS, 2013, 33 (08) : 2017 - 2023
  • [5] Analysis and Design of Interference Imaging System in Fourier Transform Imaging Spectrometer Based on Multi-Micro-Mirror
    Lu Jin-guang
    Liang Jing-qiu
    Liang Zhong-zhu
    Tian Chao
    Qin Yu-xin
    SPECTROSCOPY AND SPECTRAL ANALYSIS, 2016, 36 (05) : 1554 - 1559
  • [6] Portable open-path optical remote sensing (ORS) Fourier transform infrared (FTIR) instrumentation miniaturization and software for point and click real-time analysis
    Zemek, Peter G.
    Plowman, Steven V.
    NEXT-GENERATION SPECTROSCOPIC TECHNOLOGIES III, 2010, 7680
  • [7] Optical Design and Investigation of a Dual-Interference Channels and Bispectrum Static Fourier-Transform Imaging Spectrometer Based on Stepped Micro-Mirror
    Ren, Jun
    Lu, Jinguang
    Zhao, Baixuan
    Wang, Qiang
    Qin, Yuxin
    Tao, Jin
    Liang, Jingqiu
    Wang, Weibiao
    IEEE ACCESS, 2021, 9 : 81871 - 81881
  • [8] Using Multiple Calibration Sets to Improve the Quantitative Accuracy of Partial Least Squares (PLS) Regression on Open-Path Fourier Transform Infrared (OP/FT-IR) Spectra of Ammonia over Wide Concentration Ranges
    Shao, Limin
    Liu, Bianxia
    Griffiths, Peter R.
    Leytem, April B.
    APPLIED SPECTROSCOPY, 2011, 65 (07) : 820 - 824