High intermediate precision Sm isotope measurements in geological samples by MC-ICP-MS

被引:11
|
作者
Bai, Jiang-Hao [1 ,2 ,4 ]
Lin, Mang [1 ]
Zhong, Song-Xiong [3 ]
Deng, Yi-Nan [2 ]
Zhang, Le [1 ]
Luo, Kai [1 ,4 ]
Wu, Hao [1 ,4 ]
Ma, Jin-Long [1 ]
Wei, Gang-Jian [1 ]
机构
[1] Chinese Acad Sci, Guangzhou Inst Geochem, CAS Ctr Excellence Deep Earth Sci, State Key Lab Isotope Geochem, Guangzhou 510640, Peoples R China
[2] Minist Nat & Resources, Key Lab Marine Mineral Resources, Guangzhou 511458, Peoples R China
[3] Guangdong Acad Sci, Inst Ecoenvironm & Soil Sci, Natl Reg Joint Engn Res Ctr Soil Pollut Control &, Key Lab Integrated Agroenvironm Pollut Control & M, Guangzhou 510650, Guangdong, Peoples R China
[4] Univ Chinese Acad Sci, Beijing 100049, Peoples R China
基金
中国国家自然科学基金;
关键词
RARE-EARTH-ELEMENTS; TODGA RESIN COLUMN; HALF-LIFE; SEPARATION; NEODYMIUM; ND; FRACTIONATION; EXTRACTION; RATIO; EVOLUTION;
D O I
10.1039/d2ja00412g
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The measurement of Sm isotopes in geological materials is challenging, thus limiting their extensive applications in geoscience and environmental science. Herein, we developed a technique for the high intermediate precision analysis of Sm isotopes via multi-collector-inductively coupled plasma-mass spectrometry (MC-ICP-MS). Interference impurities, particularly Nd, Eu, and Gd, were completely removed from Sm using a new two-column [AG50W-X12 (200-400 mesh) and TODGA (50-100 mu m)] procedure, which quantitatively (99.4 +/- 0.3%) extracted Sm. The complete purification of Sm allowed the development of a high intermediate precision calibration method combined with standard-sample bracketing and Eu internal normalization to correct the instrumental mass bias. A long-term intermediate precision better than 0.04 parts per thousand (2SD) for delta Sm-152/149 was obtained. This intermediate precision is significantly better than those of the existing MC-ICP-MS methods (0.15 parts per thousand-0.30 parts per thousand) and double spike-thermal ionization mass spectrometry methods (0.15 parts per thousand-0.20 parts per thousand), as confirmed by multiple purifications and measurements of 11 geological reference materials (GRMs) with different rock types (e.g., basalt, andesite, granodiorite, granite, nepheline, syenite, manganese nodule, and marine sediment). In addition, we observed a detectable variation in delta Sm-152/149 (0.22 parts per thousand) among these 11 GRMs, which could not be identified at the previous intermediate precision. This finding highlights that the high intermediate precision delta Sm-152/149 measurements in natural samples established herein have the potential to provide new constraints on diverse natural processes, ranging from planetary formation to chemical weathering and ocean circulation and create new opportunities for stable Sm isotope studies.
引用
收藏
页码:629 / 637
页数:10
相关论文
共 50 条
  • [21] Determination of europium isotope ratios in natural waters by MC-ICP-MS
    Arantes de Carvalho, Gabriel Gustinelli
    Oliveira, Pedro Vitoriano
    Yang, Lu
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2017, 32 (05) : 987 - 995
  • [22] High-Precision Sn Isotopic Measurement in Geological Materials by Double-Spike MC-ICP-MS
    Song, Wenrui
    Zhu, Chuanwei
    Wen, Hanjie
    Zhang, Yuxu
    Liu, Zerui
    Zhou, Meifu
    GEOSTANDARDS AND GEOANALYTICAL RESEARCH, 2024,
  • [23] Sulfur isotope analysis by MC-ICP-MS and application to small medical samples
    Albalat, Emmanuelle
    Telouk, Philippe
    Balter, Vincent
    Fujii, Toshiyuki
    Bondanese, Victor Packy
    Plissonnier, Marie-Laure
    Vlaeminck-Guillem, Virginie
    Baccheta, Justine
    Thiam, Ndieme
    Miossec, Pierre
    Zoulim, Fabien
    Puisieux, Alain
    Albarede, Francis
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2016, 31 (04) : 1002 - 1011
  • [24] Precise and accurate Ga isotope analysis of solution standards and geological reference materials by MC-ICP-MS
    Li, Dandan
    Zhang, Zida
    Fan, Xuesong
    Wu, Tianhao
    Huang, Zixuan
    Shen, Qianqian
    Liu, Xuefei
    Liu, Sheng-Ao
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2025, 40 (05) : 1383 - 1393
  • [25] High-precision copper and iron isotope analysis of igneous rock standards by MC-ICP-MS
    Liu, Sheng-Ao
    Li, Dandan
    Li, Shuguang
    Teng, Fang-Zhen
    Ke, Shan
    He, Yongsheng
    Lu, Yinghuai
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2014, 29 (01) : 122 - 133
  • [26] High-Precision Cd Isotope Measurements of Soil and Rock Reference Materials by MC-ICP-MS with Double Spike Correction
    Liu, Meng-Shu
    Zhang, Qun
    Zhang, Yingnan
    Zhang, Zhaofeng
    Huang, Fang
    Yu, Hui-Min
    GEOSTANDARDS AND GEOANALYTICAL RESEARCH, 2020, 44 (01) : 169 - 182
  • [27] The impact of MC-ICP-MS plasma conditions on the accuracy and precision of stable isotope measurements evaluated for barium isotopes
    Yu, Yang
    Siebert, Christopher
    Fietzke, Jan
    Goepfert, Tyler
    Hathorne, Ed
    Cao, Zhimian
    Frank, Martin
    CHEMICAL GEOLOGY, 2020, 549
  • [28] A single-stage purification method for the precise determination of zirconium isotopic composition in geological samples by double spike MC-ICP-MS
    Feng, Lanping
    Wang, Chen
    Han, Yan
    Zhang, Wen
    Guo, Jing-Liang
    Zhou, Lian
    Hu, Zhaochu
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2024, 39 (12) : 3035 - 3047
  • [29] High-precision Mg isotope analyses of low-Mg rocks by MC-ICP-MS
    An, Yajun
    Wu, Fei
    Xiang, Yuanxin
    Nan, Xiaoyun
    Yu, Xun
    Yang, Jinhui
    Yu, Huimin
    Xie, Liewen
    Huang, Fang
    CHEMICAL GEOLOGY, 2014, 390 : 9 - 21
  • [30] Efficient Cd separation protocols for high-precision cadmium isotope analyses of diverse samples by double spike MC-ICP-MS
    Huang, Yihang
    Jia, Zijian
    Xu, Hui
    Kreissig, Katharina
    Coles, Barry J.
    Rehkamper, Mark
    Moore, Rebekah E. T.
    TALANTA, 2025, 285