A fast response time gas ionization chamber detector with a grid structure

被引:1
|
作者
Chang, Jiahao [1 ,2 ]
Zhu, Chaoyang [1 ,2 ]
Song, Yuanpeng [1 ,2 ]
Wang, Zhentao [1 ,2 ]
机构
[1] Tsinghua Univ, Inst Nucl & New Energy Technol, Beijing, Peoples R China
[2] Beijing Key Lab Nucl Detect Technol, Beijing, Peoples R China
关键词
Time response characteristics; gas ionization chamber; garfield plus; grid detector; detector sensitivity;
D O I
10.3233/XST-230219
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The time response characteristic of the detector is crucial in radiation imaging systems. Unfortunately, existing parallel plate ionization chamber detectors have a slow response time, which leads to blurry radiation images. To enhance imaging quality, the electrode structure of the detector must be modified to reduce the response time. This paper proposes a gas detector with a grid structure that has a fast response time. In this study, the detector electrostatic field was calculated using COMSOL, while Garfield++ was utilized to simulate the detector's output signal. To validate the accuracy of simulation results, the experimental ionization chamber was tested on the experimental platform. The results revealed that the average electric field intensity in the induced region of the grid detector was increased by at least 33%. The detector response time was reduced to 27%-38% of that of the parallel plate detector, while the sensitivity of the detector was only reduced by 10%. Therefore, incorporating a grid structure within the parallel plate detector can significantly improve the time response characteristics of the gas detector, providing an insight for future detector enhancements.
引用
收藏
页码:339 / 354
页数:16
相关论文
共 50 条
  • [1] Design and response function of fast neutron detector using boron-lined ionization chamber
    Zhang, Zi-Xia
    Wei, Zhi-Yong
    Fang, Mei-Hua
    Zhu, Li
    Chen, Guo-Yun
    Shi, Miao
    Yuanzineng Kexue Jishu/Atomic Energy Science and Technology, 2014, 48 (07): : 1274 - 1280
  • [2] The fast-response flame ionization detector
    Cheng, WK
    Summers, T
    Collings, N
    PROGRESS IN ENERGY AND COMBUSTION SCIENCE, 1998, 24 (02) : 89 - 124
  • [3] GAS CHROMATOGRAPHY OF STEROIDS - RELATION OF STRUCTURE TO MOLAR RESPONSE IN AN ARGON IONIZATION DETECTOR
    SWEELEY, CC
    CHANG, TCL
    ANALYTICAL CHEMISTRY, 1961, 33 (13) : 1860 - &
  • [4] Study on time response character for high pressure gas ionization chamber of krypton and xenon
    Tan, Chun-Ming
    Wu, Hai-Feng
    Qing, Shang-Yu
    Wang, Li-Qiang
    Yuanzineng Kexue Jishu/Atomic Energy Science and Technology, 2006, 40 (SUPPL.): : 139 - 141
  • [5] DESIGN AND CONSTRUCTION OF IONIZATION-CHAMBER DETECTOR FOR RADIO GAS CHROMATOGRAPHY
    ARIMOTO, H
    JAPAN ANALYST, 1973, 22 (07): : 837 - 843
  • [6] Gas electron multiplier (GEM) enhanced ionization chamber for fluorescence detector
    Shaban, E. H.
    Siddons, D. P.
    Kuczewski, A.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 582 (01): : 185 - 186
  • [7] Development of an a grid ionization chamber
    China Institute of Atomic Energy, Beijing 102413, China
    Hedianzixue Yu Tance Jishue, 2006, 3 (331-334):
  • [8] GRID IONIZATION CHAMBER MEASUREMENTS OF ELECTRON DRIFT VELOCITIES IN GAS MIXTURES
    ENGLISH, WN
    HANNA, GC
    CANADIAN JOURNAL OF PHYSICS, 1953, 31 (05) : 768 - 797
  • [9] A FAST-RESPONSE FLAME IONIZATION DETECTOR FOR EXHAUST HYDROCARBONS
    GRAIFF, LB
    LEGATE, CE
    ROBINSON, IC
    SAE PROGRESS IN TECHNOLOGY, 1967, 12 : 335 - &
  • [10] FAST IONIZATION CHAMBER FOR FISSION
    KONONOV, VN
    METLEV, AA
    POLETAEV, ED
    PROKOPETS, YS
    INSTRUMENTS AND EXPERIMENTAL TECHNIQUES-USSR, 1969, (06): : 1419 - +