Reproduction of a conventional anthropomorphic female chest phantom by 3D-printing: Comparison of image contrasts and absorbed doses in CT

被引:5
作者
Kunert, Patrizia [1 ,4 ]
Schlattl, Helmut [1 ]
Trinkl, Sebastian [1 ]
Giussani, Augusto [1 ]
Klein, Lea [2 ]
Janich, Martin [2 ]
Reichert, Detlef [3 ]
Brix, Gunnar [1 ]
机构
[1] Fed Off Radiat Protect, Dept Med & Occupat Radiat Protect, Oberschleissheim, Germany
[2] Martin Luther Univ Halle Wittenberg, Dept Radiat Oncol, Halle, Saale, Germany
[3] Martin Luther Univ Halle Wittenberg, Dept Phys, Halle, Saale, Germany
[4] Bundesamt Strahlenschutz, Abt Med & Beruflichen Strahlenschutz, Externe & Interne Dosimetrie, Biokinet, Ingolstadter Landstr 1, D-85764 Oberschleissheim, Germany
关键词
3D-printing; anthropomorphic phantom; quality assurance; X-ray imaging; 3D; FABRICATION;
D O I
10.1002/mp.16587
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
BackgroundThe production of individualized anthropomorphic phantoms via three-dimensional (3D) printing methods offers promising possibilities to assess and optimize radiation exposures for specifically relevant patient groups (i.e., overweighed or pregnant persons) that are not adequately represented by standardized anthropomorphic phantoms. However, the equivalence of printed phantoms must be demonstrated exemplarily with respect to the resulting image contrasts and dose distributions. PurposeTo reproduce a conventionally produced anthropomorphic phantom of a female chest and breasts and to evaluate their equivalence with respect to image contrasts and absorbed doses at the example of a computed tomography (CT) examination of the chest. MethodsIn a first step, the effect of different print settings on the CT values of printed samples was systematically investigated. Subsequently, a transversal slice and breast add-ons of a conventionally produced female body phantom were reproduced using a multi-material extrusion-based printer, considering six different types of tissues (muscle, lung, adipose, and glandular breast tissue, as well as bone and cartilage). CT images of the printed and conventionally produced phantom parts were evaluated with respect to their geometric correspondence, image contrasts, and absorbed doses measured using thermoluminescent dosimeters. ResultsCT values of printed objects are highly sensitive to the selected print settings. The soft tissues of the conventionally produced phantom could be reproduced with a good agreement. Minor differences in CT values were observed for bone and lung tissue, whereas absorbed doses to the relevant tissues were identical within the measurement uncertainties. Conclusion3D-printed phantoms are with exception of minor contrast differences equivalent to their conventionally manufactured counterparts. When comparing the two production techniques, it is important to note that conventionally manufactured phantoms should not be considered as absolute benchmarks, as they also only approximate the human body in terms of its absorption, and attenuation of x-rays as well as its geometry.
引用
收藏
页码:4734 / 4743
页数:10
相关论文
共 30 条
  • [1] Hounsfield unit for the diagnosis of bone mineral density disease: A proof of concept study
    Batawil, N.
    Sabiq, S.
    [J]. RADIOGRAPHY, 2016, 22 (02) : E93 - E98
  • [2] Assessment of a theoretical formalism for dose estimation in CT: an anthropomorphic phantom study
    Brix, G
    Lechel, U
    Veit, R
    Truckenbrodt, R
    Stamm, G
    Coppenrath, EM
    Griebel, J
    Nagel, HD
    [J]. EUROPEAN RADIOLOGY, 2004, 14 (07) : 1275 - 1284
  • [3] Patient-Specific Quality Assurance Using a 3D-Printed Chest Phantom for Intraoperative Radiotherapy in Breast Cancer
    Choi, Yeonho
    Lee, Ik Jae
    Park, Kwangwoo
    Park, Kyung Ran
    Cho, Yeona
    Kim, Jun Won
    Lee, Ho
    [J]. FRONTIERS IN ONCOLOGY, 2021, 11
  • [4] Evaluation of single material and multi-material patient-specific, 3D-printed radiotherapy phantoms
    Craft, D.
    Burgett, E.
    Howell, R.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2017, 123 : S765 - S766
  • [5] Preparation and fabrication of a full-scale, sagittal-sliced, 3D-printed, patient-specific radiotherapy phantom
    Craft, Daniel F.
    Howell, Rebecca M.
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2017, 18 (05): : 285 - 292
  • [6] Patient specific 3D printed phantom for IMRT quality assurance
    Ehler, Eric D.
    Barney, Brett M.
    Higgins, Patrick D.
    Dusenbery, Kathryn E.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2014, 59 (19) : 5763 - 5773
  • [7] 3D Slicer as an image computing platform for the Quantitative Imaging Network
    Fedorov, Andriy
    Beichel, Reinhard
    Kalpathy-Cramer, Jayashree
    Finet, Julien
    Fillion-Robin, Jean-Christophe
    Pujol, Sonia
    Bauer, Christian
    Jennings, Dominique
    Fennessy, Fiona
    Sonka, Milan
    Buatti, John
    Aylward, Stephen
    Miller, James V.
    Pieper, Steve
    Kikinis, Ron
    [J]. MAGNETIC RESONANCE IMAGING, 2012, 30 (09) : 1323 - 1341
  • [8] 3D printed patient-specific thorax phantom with realistic heterogenous bone radiopacity using filament printer technology
    Hatamikia, Sepideh
    Kronreif, Gernot
    Unger, Alexander
    Oberoi, Gunpreet
    Jaksa, Laszlo
    Unger, Ewald
    Koschitz, Stefan
    Gulyas, Ingo
    Irnstorfer, Nikolaus
    Buschmann, Martin
    Kettenbach, Joachim
    Birkfellner, Wolfgang
    Lorenz, Andrea
    [J]. ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2022, 32 (04): : 438 - 452
  • [9] Using 3D printing techniques to create an anthropomorphic thorax phantom for medical imaging purposes
    Hazelaar, Colien
    van Eijnatten, Maureen
    Dahele, Max
    Wolff, Jan
    Forouzanfar, Tymour
    Slotman, Ben
    Verbakel, Wilko F. A. R.
    [J]. MEDICAL PHYSICS, 2018, 45 (01) : 92 - 100
  • [10] Estimating Age at Death Based on Costal Cartilage Calcification
    Ikeda, Tomoya
    [J]. TOHOKU JOURNAL OF EXPERIMENTAL MEDICINE, 2017, 243 (04) : 237 - 246