The Feasibility of the Three-Dimensional Footswitch-Operated Robotic Arm Exoscope for Cochlear Implant Surgery

被引:2
|
作者
Karinen, Eerika [1 ]
Iso-Mustajaervi, Matti [2 ]
Dietz, Aarno [2 ,3 ,4 ]
机构
[1] Satasairaala Hosp, Dept Otorhinolaryngol, Pori, Finland
[2] Kuopio Univ Hosp, Dept Otorhinolaryngol, Kuopio, Finland
[3] Univ Eastern Finland, Inst Clin Med, Otorhinolaryngol, Kuopio, Finland
[4] Kuopio Univ Hosp, Dept Otorhinolaryngol, POB 100, FI-70029 Kuopio, Finland
基金
芬兰科学院;
关键词
Cochlear implantation; Complications; Ergonomics; Exoscope; Image quality; Surgical outcomes; Three-dimensional;
D O I
10.1097/MAO.0000000000003952
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
ObjectiveTo compare the three-dimensional (3D) footswitch-operated robotic arm exoscope with the operating microscope (OM) in cochlear implant surgery.Study designMatched case-control study.PatientsCochlear implantation was performed with the exoscope on unselected patients with normal temporal anatomy. The control group that underwent cochlear implantation with the OM was case matched with respect to age, anatomy, surgical technique, and type of anesthesia.InterventionsCochlear implantation performed with the 3D exoscope.Main Outcome measuresSurgical time, occupation of the operation theater, surgical results, and user experience evaluated by a questionnaire.ResultsEleven patients (13 ears) were successfully operated on with the exoscope. In the exoscope group, we observed one minor intraoperative complication, where the middle dura was exposed during mastoidectomy. Although no clear preference was evident for either device in the overall rating, the subdomain rating revealed that the exoscope's image quality was deemed inferior, especially at higher magnifications where pixelation became noticeable. The exoscope received higher scores for usability, particularly excelling in terms of surgeon's ergonomic and comfortability. There was a statistically significant difference in mean surgical time, 146 and 129 min for the exoscope and OM group, respectively.ConclusionsCochlear implant surgery was found to be feasible with a 3D exoscope. However, there is a learning curve to overcome regarding handling and the different quality of the image. The exoscope provides better ergonomics for the surgeon.
引用
收藏
页码:786 / 790
页数:5
相关论文
共 50 条
  • [41] Three-dimensional A-mode ultrasound calibration and registration for robotic orthopaedic knee surgery
    Mozes, Alon
    Chang, Ta-Cheng
    Arata, Louis
    Zhao, Weizhao
    INTERNATIONAL JOURNAL OF MEDICAL ROBOTICS AND COMPUTER ASSISTED SURGERY, 2010, 6 (01): : 91 - 101
  • [42] THREE-DIMENSIONAL IMAGING AND ROBOTIC SURGERY IN A COMPREHENSIVE DEPARTMENT WITHIN AN ACADEMIC INSTITUTIONAL FRAMEWORK
    Masago, Toshihiko
    Yumioka, Tetsuya
    Yamaguchi, Noriya
    Iwamoto, Hideto
    Morizane, Shuichi
    Honda, Masashi
    Sejima, Takehiro
    Takenaka, Atsushi
    JOURNAL OF UROLOGY, 2016, 195 (04): : E475 - E476
  • [43] Robotic dexterity: Evaluation of three-dimensional monitoring system and non-dominant hand maneuverability in robotic surgery
    Ishikawa N.
    Watanabe G.
    Hirano Y.
    Inaki N.
    Kawachi K.
    Oda M.
    Journal of Robotic Surgery, 2007, 1 (3) : 231 - 233
  • [44] Robotic arm three-dimensional printing and modular construction of a meter-scale lattice facade structure
    Leong, Zee
    Chen, Rongsheng
    Xu, Zijie
    Lin, Yangsheng
    Hu, Nan
    ENGINEERING STRUCTURES, 2023, 290
  • [45] Modeling and analysis of the equivalent convex mandrel of the three-dimensional braiding carbon fiber composite robotic arm
    Meng, Zhuo
    Zhang, Rongtao
    Cai, Gaowei
    Zhang, Yujing
    TEXTILE RESEARCH JOURNAL, 2023, 93 (23-24) : 5505 - 5515
  • [46] Feasibility, accuracy, and incremental value of intraoperative three-dimensional transesophageal echocardiography in valve surgery
    Abraham, TP
    Warner, JG
    Kon, ND
    Lantz, PE
    Fowle, KM
    Brooker, RF
    Ge, SP
    Nomeir, AM
    Kitzman, DW
    AMERICAN JOURNAL OF CARDIOLOGY, 1997, 80 (12): : 1577 - 1582
  • [47] Robotic Surgical Approach to the Mesial Temporal Region: A Preliminary Three-Dimensional Cadaveric Study of Technical Feasibility
    Akbas, Ahmet
    Tugcu, Bekir
    Eksi, M. Sakir
    Erkan, Buruc
    Canbolat, Cagri
    Pamir, M. Necmettin
    Gungor, Abuzer
    WORLD NEUROSURGERY, 2020, 144 : E40 - E52
  • [48] Spinal cord compression in thoracolumbar burst fractures: application of high-definition three-dimensional exoscope in minimally invasive lateral surgery
    Pietro Domenico Giorgi
    Maria Ludovica Pallotta
    Simona Legrenzi
    Michele Nardi
    Manzoni Andrea
    Giuseppe Rosario Schirò
    European Journal of Orthopaedic Surgery & Traumatology, 2023, 33 : 2173 - 2177
  • [49] A Simple Method for In Vivo Measurement of Implant Rod Three-Dimensional Geometry During Scoliosis Surgery
    Salmingo, Remel A.
    Tadano, Shigeru
    Fujisaki, Kazuhiro
    Abe, Yuichiro
    Ito, Manabu
    JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2012, 134 (05):
  • [50] Three-dimensional accuracy of partially guided implant surgery based on dental magnetic resonance imaging
    Schwindling, Franz Sebastian
    Juerchott, Alexander
    Boehm, Sophia
    Rues, Stefan
    Kronsteiner, Dorothea
    Heiland, Sabine
    Bendszus, Martin
    Rammelsberg, Peter
    Hilgenfeld, Tim
    CLINICAL ORAL IMPLANTS RESEARCH, 2021, 32 (10) : 1218 - 1227