Implementation of triggered kilovoltage imaging for stereotactic radiotherapy of the spine for patients with spinal fixation hardware

被引:3
|
作者
Cetnar, Ashley J. [1 ,2 ,3 ]
Degnan, Michael [1 ,2 ]
Pichler, Joseph [1 ,2 ]
Jain, Sagarika [1 ,2 ]
Morelli, Samantha [1 ,2 ]
Thomas, Evan [1 ,2 ]
Elder, J. Bradley [1 ,2 ]
Scharschmidt, Thomas J. [1 ,2 ]
Palmer, Joshua D. [1 ,2 ]
Blakaj, Dukagjin M. [1 ,2 ]
机构
[1] Ohio State Univ, James Canc Hosp, 460 W 10th Ave, Columbus, OH 43210 USA
[2] Ohio State Univ, Solove Res Inst, Dept Radiat Oncol, 460 W 10th Ave, Columbus, OH 43210 USA
[3] Dept Radiat Oncol, A201 Off F,460 W 10th Ave, Columbus, OH 43210 USA
来源
PHYSICS & IMAGING IN RADIATION ONCOLOGY | 2023年 / 25卷
关键词
Intrafraction motion review; Triggered imaging; Advanced imaging; Spine; Stereotactic; BODY RADIATION-THERAPY; INTRAFRACTION MOTION; CONSENSUS GUIDELINES; RESPIRATORY MOTION; METASTASES; PROSTATE; RADIOSURGERY; MANAGEMENT; CANCER; CORD;
D O I
10.1016/j.phro.2023.100422
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background and purpose: Mitigation of intrafraction motion (IM) is valuable in stereotactic radiotherapy (SRT) radiotherapy where submillimeter accuracy is desired. The purpose of this study was to investigate the application of triggered kilovoltage (kV) imaging for spine SRT patients with hardware by correlating kV imaging with patient motion and summarizing implications of tolerance for IM based on calculated dose.Materials and methods: Ten plans (33 fractions) were studied, correlating kV imaging during treatment with preand post-treatment cone beam computed tomography (CBCT). Images were taken at 20-degree gantry angle intervals during the arc-based treatment. The contour of the hardware with a 1 mm expansion was displayed at the treatment console to manually pause treatment delivery if the hardware was visually detected outside the contour. The treatment CBCTs were compared using retrospective image registration to assess the validity of contour-based method for pausing treatment. Finally, plans were generated to estimate dose volume objective differences in case of 1 mm deviation.Results: When kV imaging during treatment was used with the 1 mm contour, 100 % of the post-treatment CBCTs reported consistent results. One patient in the cohort exhibited motion greater than 1 mm during treatment which allowed intervention and re-setup during treatment. The average translational motion was 0.35 mm. Treatment plan comparison at 1 mm deviation showed little differences in calculated dose for the target and cord.Conclusions: Utilizing kV imaging during treatment is an effective method of assessing IM for SRT spine patients with hardware without increasing treatment time.
引用
收藏
页数:5
相关论文
共 45 条
  • [1] Computed Tomography Myelosimulation Versus Magnetic Resonance Imaging Registration to Delineate the Spinal Cord During Spine Stereotactic Radiosurgery
    Beeler, Whitney H.
    Paradis, Kelly C.
    Gemmete, Joseph J.
    Chaudhary, Neeraj
    Kim, Michelle M.
    Smith, Sean R.
    Paradis, Eric
    Matuszak, Martha M.
    Park, Paul
    Archer, Paul G.
    Szerlip, Nicholas J.
    Spratt, Daniel E.
    WORLD NEUROSURGERY, 2019, 122 : E655 - E666
  • [2] Update on Imaging of Spinal Fixation Hardware
    Winegar, Blair A.
    Kay, Matthew D.
    Chadaz, Tyson S.
    Taljanovic, Mihra S.
    Hood, Kenneth A.
    Hunter, Tim B.
    SEMINARS IN MUSCULOSKELETAL RADIOLOGY, 2019, 23 (02) : E56 - E79
  • [3] Early magnetic resonance imaging biomarkers to predict local control after high dose stereotactic body radiotherapy for patients with sarcoma spine metastases
    Spratt, Daniel E.
    Arevalo-Perez, Julio
    Leeman, Jonathan E.
    Gerber, Naamit K.
    Folkert, Michael
    Taunk, Neil K.
    Alektiar, Kaled M.
    Karimi, Sasan
    Lyo, John K.
    Tap, William D.
    Bilsky, Mark H.
    Laufer, Ilya
    Yamada, Yoshiya
    Osborne, Joseph R.
    SPINE JOURNAL, 2016, 16 (03) : 291 - 298
  • [4] SPINAL FIXATION .2. FIXATION TECHNIQUES AND HARDWARE FOR THE THORACIC AND LUMBOSACRAL SPINE
    SLONE, RM
    MACMILLAN, M
    MONTGOMERY, WJ
    HEARE, M
    RADIOGRAPHICS, 1993, 13 (03) : 521 - 543
  • [5] Stereotactic Body Radiotherapy for Spinal Metastases at the Extreme Ends of the Spine: Imaging-Based Outcomes for Cervical and Sacral Metastases
    Zeng, K. Liang
    Myrehaug, Sten
    Soliman, Hany
    Tseng, Chia-Lin
    Atenafu, Eshetu G.
    Campbell, Mikki
    Faruqi, Salman
    Lee, Young K.
    Ruschin, Mark
    da Costa, Leodante
    Yang, Victor
    Spears, Julian
    Heyn, Chris
    Maralani, Pejman Jabehdar
    Whyne, Cari
    Yee, Albert
    Sahgal, Arjun
    NEUROSURGERY, 2019, 85 (05) : 605 - 612
  • [6] Risk-adapted stereotactic body radiotherapy for patients with cervical spinal metastases
    Wang, Huan-Huan
    Tian, Shou-Sen
    Yang, Jia-Min
    Sun, Bing-Sheng
    Chen, Ying
    Song, Yong-Chun
    Dong, Yang
    Wang, Jing-Sheng
    Yuan, Zhi-Yong
    Cui, Yao-Li
    Meng, Mao-Bin
    CANCER SCIENCE, 2022, 113 (12) : 4277 - 4288
  • [7] Commentary: Stereotactic Body Radiotherapy for Spinal Metastases at the Extreme Ends of the Spine: Imaging-Based Outcomes for Cervical and Sacral Metastases
    Vellayappan, Balamurugan
    Tan, Daniel
    Foote, Matthew
    Redmond, Kristin J.
    Lo, Simon S.
    NEUROSURGERY, 2019, 85 (05) : E804 - E805
  • [8] Spine Stereotactic Body Radiotherapy Outcomes in Patients with Concurrent Brain Metastases
    Colaco, Rovel J.
    Park, Henry S.
    Laurans, Maxwell S.
    Chiang, Veronica S.
    Yu, James B.
    Husain, Zain A.
    CUREUS, 2016, 8 (07):
  • [9] SPINAL FIXATION .1. PRINCIPLES, BASIC HARDWARE, AND FIXATION TECHNIQUES FOR THE CERVICAL-SPINE
    SLONE, RM
    MACMILLAN, M
    MONTGOMERY, WJ
    RADIOGRAPHICS, 1993, 13 (02) : 341 - 356
  • [10] Separation Surgery, Fixation With Carbon-Fiber Implants, and Stereotactic Body Radiotherapy for Oligometastatic Spinal Disease
    Galloway, Richard
    Gikas, Nicholas
    Golomohammad, Ramez
    Sherriff, Jenny
    Czyz Sr, Marcin
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2022, 14 (11)