A case study for in-house method validation of gas chromatography technique using class-1 calibration gas mixtures for greenhouse gases monitoring

被引:3
作者
Komal [1 ,2 ]
Soni, Daya [1 ,2 ]
Aggarwal, Shankar G. [1 ,2 ]
机构
[1] CSIR Natl Phys Lab, Dr KS Krishnan Marg, New Delhi 110012, India
[2] Acad Sci & Innovat Res AcSIR, Ghaziabad 201002, India
关键词
Greenhouse gases; GC-FID; Method validation; Calibration gas mixture; CARBON-DIOXIDE; ANALYTICAL-CHEMISTRY; LINEAR-REGRESSION; METHANE; PERFORMANCE; SELECTIVITY; CH4;
D O I
10.1007/s00769-023-01552-z
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
In this study, analytical method validation has been done for the measurement of carbon dioxide/nitrogen (CO2/N-2) and methane/nitrogen (CH4/N-2) calibration gas mixtures using gas chromatography with flame ionization detector (GC-FID). Class-I calibration gas mixtures (CGMs) of CO2 (500 mu mol mol-1 to 1100 mu mol mol(-1)) and CH4 (2 mu mol mol(-1) to 130 mu mol mol(-1)) used in method validation process has been prepared gravimetrically following ISO 6142-1. All prepared gas mixtures have expanded uncertainty 1 % at coverage factor (k) of 2 with 95 % confidence. The following parameters are chosen for this case study which include selectivity, accuracy, precision, linearity, limit of detection (LOD), limit of quantification (LOQ), robustness, stability, and uncertainty. Different statistical approaches are taken into consideration for each parameter assessment. The results indicate that GC-FID is selective for CO2 and CH4. CGMs represent good repeatability and reproducibility having percentage relative deviation < 1 % among measurements. A good linear behaviour was observed for CGMs of CH4 and CO2 on basis of least square regression with R-2 value 0.9995 and 1, respectively. LOD and LOQ for CH4 are calculated 0.47 and 1.59 mu mol mol(-1) based on signal-to-noise ratio by taking its lowest concentration of 2.9 mu mol mol-1. The in-house validated method for GC-FID using CGMs for the measurement of greenhouse gases (CO2 and CH4) is found to be precise, accurate and fit for purpose.
引用
收藏
页码:209 / 220
页数:12
相关论文
共 49 条
  • [1] A heuristic derivation of the Horwitz curve
    Albert, R
    Horwitz, W
    [J]. ANALYTICAL CHEMISTRY, 1997, 69 (04) : 789 - 790
  • [2] Notes on the use of Mandel's test to check for nonlinearity in laboratory calibrations
    Andrade, J. M.
    Gomez-Carracedo, M. P.
    [J]. ANALYTICAL METHODS, 2013, 5 (05) : 1145 - 1149
  • [3] [Anonymous], 2023, ISO 5725-1
  • [4] [Anonymous], 2001, 6143 ISO
  • [5] [Anonymous], 2014, The Fitness for Purpose of Analytical Methods: A Laboratory Guide to Method Validation and Related Topics
  • [6] [Anonymous], 2008, 1002008 BIPM JCGM
  • [7] [Anonymous], 2015, ISO 6142-1
  • [8] [Anonymous], 2017, ISO GUIDE 35
  • [9] [Anonymous], 1994, INT C HARM TECHN REQ
  • [10] [Anonymous], 2007, ISO/IEC Guide 99