Needle Aligned Ultrasound Image-Guided Access Through Dual-Segment Array

被引:1
作者
Rahaman, Ashiqur [1 ,2 ]
Tang, Yichuan [1 ]
Gao, Shang [1 ]
Ma, Xihan [1 ]
Sorokin, Igor [3 ]
Zhang, Haichong K. [1 ,2 ,4 ]
机构
[1] Worcester Polytech Inst, Dept Robot Engn, Worcester, MA 01609 USA
[2] Worcester Polytech Inst, Dept Biomed Engn, Worcester, MA 01609 USA
[3] Univ Massachusetts, Med Sch, Dept Urol, Worcester, MA USA
[4] Worcester Polytech Inst, Dept Comp Sci, Worcester, MA 01609 USA
基金
美国国家卫生研究院;
关键词
Needles; Acoustic beams; Probes; Kidney; Imaging; Image segmentation; Visualization; Needle-image alignment; Percutaneous intervention; Ultrasound-guided intervention; Dual-segment array; PERCUTANEOUS NEPHROLITHOTOMY; VISUALIZATION; GUIDANCE; FREEHAND; BIOPSY; SYSTEM;
D O I
10.1109/TBME.2023.3260735
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Ultrasound (US) guided access for percutaneous nephrolithotomy (PCNL) is gaining popularity in the urology community as it reduces radiation risk. The most popular technique involves manual image-needle alignment. A misaligned needle however needs to be retracted and reinserted, resulting in a lengthened operation time and complications such as bleeding. These limitations can be mitigated through the co-registration between the US array and needle. The through-hole array concept provides the primary solution, including a hole at the center of the array. Because of the central opening, the image-needle alignment is achieved inherently. Previous literature has described applications that are limited to superficial and intravascular procedures, suggesting that developing a through-hole array for deeper target applications would be a new breakthrough. Objective. Here, we present a dual-segment array with a central opening. As the prototype development, two segments of 32-element arrays are combined with an open space of 10 mm in length in between them. Method. We conducted phantom and ex-vivo studies considering the target depth of the 80-100 mm range. The image quality and needle visibility are evaluated by comparing the signal-to-noise ratio (SNR), full width at half maximum (FWHM), and contrast-to-noise ratio (CNR) results measured with a no-hole linear array under equivalent conditions. An ex-vivo study is performed using porcine kidneys with ceramic balls embedded to evaluate the needle access accuracy. Results and conclusion.The mean needle access error of 20 trials is found to be 2.94$\pm$1.09 mm, suggesting its potential impact on realizing a simple and intuitive deep US image-guided access.
引用
收藏
页码:2645 / 2654
页数:10
相关论文
共 44 条
  • [1] Automatic needle detection and real-time Bi-planar needle visualization during 3D ultrasound scanning of the liver
    Arif, Muhammad
    Moelker, Adriaan
    van Walsum, Theo
    [J]. MEDICAL IMAGE ANALYSIS, 2019, 53 : 104 - 110
  • [2] RENAL VOLUME ESTIMATION USING FREEHAND ULTRASOUND SCANS: AN EX VIVO DEMONSTRATION
    Benjamin, Alex
    Chen, Melinda
    Li, Qian
    Chen, Lei
    Dong, Yi
    Carrascal, Carolina A.
    Xie, Hua
    Samir, Anthony E.
    Anthony, Brian W.
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 2020, 46 (07) : 1769 - 1782
  • [3] An overview of kidney stone imaging techniques
    Brisbane, Wayne
    Bailey, Michael R.
    Sorensen, Mathew D.
    [J]. NATURE REVIEWS UROLOGY, 2016, 13 (11) : 654 - 662
  • [4] Comparison between 1-3 Piezocomposite and PZT ceramic for High-Intensity Focused Ultrasound Transducer Applications
    Chen, Gin-Shin
    Liu, Hsin-Chih
    Lin, Yu-Li
    [J]. 11TH INTERNATIONAL SYMPOSIUM ON THERAPEUTIC ULTRASOUND, 2012, 1481 : 47 - 52
  • [5] Needle Visualization in Ultrasound-Guided Regional Anesthesia: Challenges and Solutions
    Chin, Ki Ann
    Perlas, Anahi
    Chan, Vincent W. S.
    Brull, Richard
    [J]. REGIONAL ANESTHESIA AND PAIN MEDICINE, 2008, 33 (06) : 532 - 544
  • [6] Design and validation of an ultrasound array optimised for epidural needle guidance
    Cochran, S.
    Corner, G. A.
    Kirk, Y. J.
    Lineo, D. I. A.
    Watson, M. J.
    [J]. 2007 IEEE ULTRASONICS SYMPOSIUM PROCEEDINGS, VOLS 1-6, 2007, : 2255 - +
  • [7] Innovations in Ultrasound Technology in the Management of Kidney Stones
    Dai, Jessica C.
    Bailey, Michael R.
    Sorensen, Mathew D.
    Harper, Jonathan D.
    [J]. UROLOGIC CLINICS OF NORTH AMERICA, 2019, 46 (02) : 273 - +
  • [8] Post-percutaneous nephrolithotomy extensive hemorrhage: A study of risk factors
    El-Nahas, Ahmed R.
    Shokeir, Ahmed A.
    El-Assmy, Ahmed M.
    Mohsen, Tarek
    Shoma, Ahmed M.
    Eraky, Ibrahim
    El-Kenawy, Mahmoud R.
    El-Kappany, Hamdy A.
    [J]. JOURNAL OF UROLOGY, 2007, 177 (02) : 576 - 579
  • [9] Through-needle all-optical ultrasound imaging in vivo: a preclinical swine study
    Finlay, Malcolm C.
    Mosse, Charles A.
    Colchester, Richard J.
    Noimark, Sacha
    Zhang, Edward Z.
    Ourselin, Sebastien
    Beard, Paul C.
    Schilling, Richard J.
    Parkin, Ivan P.
    Papakonstantinou, Ioannis
    Desjardins, Adrien E.
    [J]. LIGHT-SCIENCE & APPLICATIONS, 2017, 6 : e17103 - e17103
  • [10] Percutaneous nephrolithotomy (PCNL) a critical review
    Ganpule, Arvind P.
    Vijayakumar, Mohankumar
    Malpani, Ankur
    Desai, Mahesh R.
    [J]. INTERNATIONAL JOURNAL OF SURGERY, 2016, 36 : 660 - 664