Spatial resolution, noise properties, and detectability index of a deep learning reconstruction algorithm for dual-energy CT of the abdomen

被引:7
作者
Thor, Daniel [1 ,2 ]
Titternes, Rebecca [1 ,3 ]
Poludniowski, Gavin [1 ,3 ,4 ]
机构
[1] Karolinska Univ Hosp, Dept Med Radiat Phys & Nucl Med, Stockholm, Sweden
[2] Karolinska Inst, Dept Oncol Pathol, Stockholm, Sweden
[3] Karolinska Inst, Dept Clin Sci Intervent & Technol, Stockholm, Sweden
[4] Karolinska Univ Hosp, Med Radiat Phys & Nucl Med, SE-17176 Stockholm, Sweden
关键词
DECT; deep learning image reconstruction; image quality; FILTERED BACK-PROJECTION; RADIATION-DOSE REDUCTION; ITERATIVE RECONSTRUCTION; IMAGE-RECONSTRUCTION; PERFORMANCE; CONTRAST; OPTIMIZATION;
D O I
10.1002/mp.16300
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
BackgroundIterative reconstruction (IR) has increasingly replaced traditional reconstruction methods in computed tomography (CT). The next paradigm shift in image reconstruction is likely to come from artificial intelligence, with deep learning reconstruction (DLR) solutions already entering the clinic. An enduring disadvantage to IR has been a change in noise texture, which can affect diagnostic confidence. DLR has demonstrated the potential to overcome this issue and has recently become available for dual-energy CT. PurposeTo evaluate the spatial resolution, noise properties, and detectability index of a commercially available DLR algorithm for dual-energy CT of the abdomen and compare it to single-energy (SE) CT. MethodsAn oval 25 cm x 35 cm custom-made phantom was scanned on a GE Revolution CT scanner (GE Healthcare, Waukesha, WI) at two dose levels (13 and 5 mGy) and two iodine concentrations (8 and 2 mg/mL), using three typical abdominal scan protocols: dual-energy (DE), SE 80 kV (SE-80 kV) and SE 120 kV (SE-120 kV). Reconstructions were performed with three strengths of IR (ASiR-V: AR0%, AR50%, AR100%) and three strengths of DLR (TrueFidelity: low, medium, high). The DE acquisitions were reconstructed as mono-energetic images between 40 and 80 keV. The noise power spectrum (NPS), task transfer function (TTF), and detectability index (d') were determined for the reconstructions following the recommendations of AAPM Task Group 233. ResultsNoise magnitude reductions (relative to AR0%) for the SE protocols were on average (-29%, -21%) for (AR50%, TF-M), while for DE-70 keV were (-28%, -43%). There was less reduction in mean frequency (f(av)) for DLR than for IR, with similar results for SE and DE imaging. There was, however, a substantial change in the NPS shape when using DE with DLR, quantifiable by a marked reduction in the peak frequency (f(peak)) that was absent in SE mode. All protocols and reconstructions (including AR0%) exhibited slight to moderate shifts towards lower spatial frequencies at the lower dose (<12% in f(av)). Spatial resolution was consistently superior for DLR compared to IR for SE but not for DE. All protocols and reconstructions (including AR0%) showed decreased resolution with reduced dose and iodine concentration, with less decrease for DLR compared to IR. DLR displayed a higher d' than IR. The effect of energy was large: d' increased with lower keV, and SE-80 kV had higher d' than SE-120 kV. Using DE with DLR could provide higher d' than SE-80 kV at the higher dose but not at lower dose. ConclusionsDE imaging with DLR maintained spatial resolution and reduced noise magnitude while displaying less change in noise texture than IR. The d' was also higher with DLR than IR, suggesting superiority in detectability of iodinated contrast. Despite these trends being consistent with those previously established for SE imaging, there were some noteworthy differences. For DE imaging there was no improvement in resolution compared to IR and a change in noise texture. DE imaging with low keV and DLR had superior detectability to SE DLR at the high dose but was not better than SE-80 kV at low dose.
引用
收藏
页码:2775 / 2786
页数:12
相关论文
共 50 条
  • [41] A noise-optimized virtual monoenergetic reconstruction algorithm improves the diagnostic accuracy of late hepatic arterial phase dual-energy CT for the detection of hypervascular liver lesions
    De Cecco, Carlo N.
    Caruso, Damiano
    Schoepf, U. Joseph
    De Santis, Domenico
    Muscogiuri, Giuseppe
    Albrecht, Moritz H.
    Meinel, Felix G.
    Wichmann, Julian L.
    Burchett, Philip F.
    Varga-Szemes, Akos
    Sheafor, Douglas H.
    Hardie, Andrew D.
    EUROPEAN RADIOLOGY, 2018, 28 (08) : 3393 - 3404
  • [42] Line Integral Alternating Minimization Algorithm for Dual-Energy X-Ray CT Image Reconstruction
    Chen, Yaqi
    O'Sullivan, Joseph A.
    Politte, David G.
    Evans, Joshua D.
    Han, Dong
    Whiting, Bruce R.
    Williamson, Jeffrey F.
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2016, 35 (02) : 685 - 698
  • [43] Improving diagnostic confidence in low-dose dual-energy CTE with low energy level and deep learning reconstruction
    Lin, Xu
    Gao, Yankun
    Zhu, Chao
    Song, Jian
    Liu, Ling
    Li, Jianying
    Wu, Xingwang
    EUROPEAN JOURNAL OF RADIOLOGY, 2024, 178
  • [44] Comparing two deep learning spectral reconstruction levels for abdominal evaluation using a rapid-kVp-switching dual-energy CT scanner
    Sagdic, Hakki Serdar
    Hosseini-Siyanaki, Mohammadreza
    Raviprasad, Abheek
    Munjerin, Sefat
    Fabri, Daniella
    Grajo, Joseph
    Tonso, Victor Martins
    Magnelli, Laura
    Hochhegger, Daniela
    Anthony, Evelyn
    Hochhegger, Bruno
    Forghani, Reza
    ABDOMINAL RADIOLOGY, 2025,
  • [45] Evaluation of the Imaging Properties of a CT Scanner with the Adaptive Statistical Iterative Reconstruction Algorithm Noise, Contrast and Spatial Resolution Properties of CT Images Reconstructed at Different Blending Levels
    Barca, Patrizio
    Giannelli, Marco
    Fantacci, Maria Evelina
    Caramella, Davide
    PROCEEDINGS OF THE 10TH INTERNATIONAL JOINT CONFERENCE ON BIOMEDICAL ENGINEERING SYSTEMS AND TECHNOLOGIES, VOL 1: BIODEVICES, 2017, : 200 - 206
  • [46] Comparative assessment of noise properties for two deep learning CT image reconstruction techniques and filtered back projection
    Kawashima, Hiroki
    Ichikawa, Katsuhiro
    Takata, Tadanori
    Seto, Issei
    MEDICAL PHYSICS, 2022, 49 (10) : 6359 - 6367
  • [47] Reconstruction of Limited-angle Dual-Energy CT Using Mutual Learning and Cross-estimation (MLCE)
    Zhang, Huayu
    Xing, Yuxiang
    MEDICAL IMAGING 2016: PHYSICS OF MEDICAL IMAGING, 2016, 9783
  • [48] Application of deep-learning reconstruction algorithm for enhanced CT scan of upper abdomen under different radiation doses: focus on noise, contrast-to-noise ratio and image quality
    Yanrong Xie
    Yuan-Cheng Wang
    Shan Huang
    Shenghong Ju
    Chinese Journal of Academic Radiology, 2022, 5 : 107 - 121
  • [49] Raw data consistent deep learning-based field of view extension for dual-source dual-energy CT
    Maier, Joscha
    Erath, Julien
    Sawall, Stefan
    Fournie, Eric
    Stierstorfer, Karl
    Kachelriess, Marc
    MEDICAL PHYSICS, 2024, 51 (03) : 1822 - 1831
  • [50] Deep learning-based image reconstruction of 40-keV virtual monoenergetic images of dual-energy CT for the assessment of hypoenhancing hepatic metastasis
    Lee, Taehee
    Lee, Jeong Min
    Yoon, Jeong Hee
    Joo, Ijin
    Bae, Jae Seok
    Yoo, Jeongin
    Kim, Jae Hyun
    Ahn, Chulkyun
    Kim, Jong Hyo
    EUROPEAN RADIOLOGY, 2022, 32 (09) : 6407 - 6417