Spatial resolution, noise properties, and detectability index of a deep learning reconstruction algorithm for dual-energy CT of the abdomen

被引:7
作者
Thor, Daniel [1 ,2 ]
Titternes, Rebecca [1 ,3 ]
Poludniowski, Gavin [1 ,3 ,4 ]
机构
[1] Karolinska Univ Hosp, Dept Med Radiat Phys & Nucl Med, Stockholm, Sweden
[2] Karolinska Inst, Dept Oncol Pathol, Stockholm, Sweden
[3] Karolinska Inst, Dept Clin Sci Intervent & Technol, Stockholm, Sweden
[4] Karolinska Univ Hosp, Med Radiat Phys & Nucl Med, SE-17176 Stockholm, Sweden
关键词
DECT; deep learning image reconstruction; image quality; FILTERED BACK-PROJECTION; RADIATION-DOSE REDUCTION; ITERATIVE RECONSTRUCTION; IMAGE-RECONSTRUCTION; PERFORMANCE; CONTRAST; OPTIMIZATION;
D O I
10.1002/mp.16300
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
BackgroundIterative reconstruction (IR) has increasingly replaced traditional reconstruction methods in computed tomography (CT). The next paradigm shift in image reconstruction is likely to come from artificial intelligence, with deep learning reconstruction (DLR) solutions already entering the clinic. An enduring disadvantage to IR has been a change in noise texture, which can affect diagnostic confidence. DLR has demonstrated the potential to overcome this issue and has recently become available for dual-energy CT. PurposeTo evaluate the spatial resolution, noise properties, and detectability index of a commercially available DLR algorithm for dual-energy CT of the abdomen and compare it to single-energy (SE) CT. MethodsAn oval 25 cm x 35 cm custom-made phantom was scanned on a GE Revolution CT scanner (GE Healthcare, Waukesha, WI) at two dose levels (13 and 5 mGy) and two iodine concentrations (8 and 2 mg/mL), using three typical abdominal scan protocols: dual-energy (DE), SE 80 kV (SE-80 kV) and SE 120 kV (SE-120 kV). Reconstructions were performed with three strengths of IR (ASiR-V: AR0%, AR50%, AR100%) and three strengths of DLR (TrueFidelity: low, medium, high). The DE acquisitions were reconstructed as mono-energetic images between 40 and 80 keV. The noise power spectrum (NPS), task transfer function (TTF), and detectability index (d') were determined for the reconstructions following the recommendations of AAPM Task Group 233. ResultsNoise magnitude reductions (relative to AR0%) for the SE protocols were on average (-29%, -21%) for (AR50%, TF-M), while for DE-70 keV were (-28%, -43%). There was less reduction in mean frequency (f(av)) for DLR than for IR, with similar results for SE and DE imaging. There was, however, a substantial change in the NPS shape when using DE with DLR, quantifiable by a marked reduction in the peak frequency (f(peak)) that was absent in SE mode. All protocols and reconstructions (including AR0%) exhibited slight to moderate shifts towards lower spatial frequencies at the lower dose (<12% in f(av)). Spatial resolution was consistently superior for DLR compared to IR for SE but not for DE. All protocols and reconstructions (including AR0%) showed decreased resolution with reduced dose and iodine concentration, with less decrease for DLR compared to IR. DLR displayed a higher d' than IR. The effect of energy was large: d' increased with lower keV, and SE-80 kV had higher d' than SE-120 kV. Using DE with DLR could provide higher d' than SE-80 kV at the higher dose but not at lower dose. ConclusionsDE imaging with DLR maintained spatial resolution and reduced noise magnitude while displaying less change in noise texture than IR. The d' was also higher with DLR than IR, suggesting superiority in detectability of iodinated contrast. Despite these trends being consistent with those previously established for SE imaging, there were some noteworthy differences. For DE imaging there was no improvement in resolution compared to IR and a change in noise texture. DE imaging with low keV and DLR had superior detectability to SE DLR at the high dose but was not better than SE-80 kV at low dose.
引用
收藏
页码:2775 / 2786
页数:12
相关论文
共 50 条
  • [31] A Characterization of Deep Learning Reconstruction Applied to Dual-Energy Computed Tomography Monochromatic and Material Basis Images
    Nikolau, Ethan P.
    Toia, Giuseppe V.
    Nett, Brian
    Tang, Jie
    Szczykutowicz, Timothy P.
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2023, 47 (03) : 437 - 444
  • [32] Deep-learning reconstruction enhances image quality of Adamkiewicz Artery in low-keV dual-energy CT
    Tatsugami, Fuminari
    Higaki, Toru
    Kawashita, Ikuo
    Fujioka, Chikako
    Nakamura, Yuko
    Takahashi, Shinya
    Awai, Kazuo
    ACTA RADIOLOGICA, 2024, : 1569 - 1575
  • [33] Improving lesion conspicuity in abdominal dual-energy CT with deep learning image reconstruction: a prospective study with five readers
    Zhong, Jingyu
    Wang, Lingyun
    Shen, Hailin
    Li, Jianying
    Lu, Wei
    Shi, Xiaomeng
    Xing, Yue
    Hu, Yangfan
    Ge, Xiang
    Ding, Defang
    Yan, Fuhua
    Du, Lianjun
    Yao, Weiwu
    Zhang, Huan
    EUROPEAN RADIOLOGY, 2023, 33 (08) : 5331 - 5343
  • [34] Comparison of virtual monoenergetic imaging between a rapid kilovoltage switching dual-energy computed tomography with deep-learning and four dual-energy CTs with iterative reconstruction
    Greffier, Joel
    Si-Mohamed, Salim
    Guiu, Boris
    Frandon, Julien
    Loisy, Maeliss
    de Oliveira, Fabien
    Douek, Philippe
    Beregi, Jean-Paul
    Dabli, Djamel
    QUANTITATIVE IMAGING IN MEDICINE AND SURGERY, 2022, 12 (02) : 1149 - +
  • [35] Commentary On: Image Quality Evaluation in Dual Energy CT of the Chest, Abdomen and Pelvis in Obese Patients with Deep Learning Image Reconstruction
    Jensen, Corey T.
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2022, 46 (04) : 612 - 613
  • [36] Application of deep-learning reconstruction algorithm for enhanced CT scan of upper abdomen under different radiation doses: focus on noise, contrast-to-noise ratio and image quality
    Xie, Yanrong
    Wang, Yuan-Cheng
    Huang, Shan
    Ju, Shenghong
    CHINESE JOURNAL OF ACADEMIC RADIOLOGY, 2022, 5 (02) : 107 - 121
  • [37] Feasibility of Using Deep Learning to Generate Dual-Energy CT from 120-kV CT
    Tung, Chi-Hsiang
    Liu, Chi-Kuang
    Huang, Hsuan-Ming
    JOURNAL OF MEDICAL AND BIOLOGICAL ENGINEERING, 2023, 43 (01) : 93 - 101
  • [38] Improving spatial resolution and diagnostic confidence with thinner slice and deep learning image reconstruction in contrast-enhanced abdominal CT
    Cao, Le
    Liu, Xiang
    Qu, Tingting
    Cheng, Yannan
    Li, Jianying
    Li, Yanan
    Chen, Lihong
    Niu, Xinyi
    Tian, Qian
    Guo, Jianxin
    EUROPEAN RADIOLOGY, 2023, 33 (03) : 1603 - 1611
  • [39] Deep learning image reconstruction algorithm for carotid dual-energy computed tomography angiography: evaluation of image quality and diagnostic performance
    Chenyu Jiang
    Dan Jin
    Zhuoheng Liu
    Yan Zhang
    Ming Ni
    Huishu Yuan
    Insights into Imaging, 13
  • [40] Comparison of image quality and pancreatic ductal adenocarcinoma conspicuity between the low-kVp and dual-energy CT reconstructed with deep-learning image reconstruction algorithm
    Noda, Yoshifumi
    Takai, Yukiko
    Asano, Masashi
    Yamada, Nao
    Seko, Takuya
    Kawai, Nobuyuki
    Kaga, Tetsuro
    Miyoshi, Toshiharu
    Hyodo, Fuminori
    Kato, Hiroki
    Matsuo, Masayuki
    EUROPEAN JOURNAL OF RADIOLOGY, 2023, 159