Spatial resolution, noise properties, and detectability index of a deep learning reconstruction algorithm for dual-energy CT of the abdomen

被引:7
作者
Thor, Daniel [1 ,2 ]
Titternes, Rebecca [1 ,3 ]
Poludniowski, Gavin [1 ,3 ,4 ]
机构
[1] Karolinska Univ Hosp, Dept Med Radiat Phys & Nucl Med, Stockholm, Sweden
[2] Karolinska Inst, Dept Oncol Pathol, Stockholm, Sweden
[3] Karolinska Inst, Dept Clin Sci Intervent & Technol, Stockholm, Sweden
[4] Karolinska Univ Hosp, Med Radiat Phys & Nucl Med, SE-17176 Stockholm, Sweden
关键词
DECT; deep learning image reconstruction; image quality; FILTERED BACK-PROJECTION; RADIATION-DOSE REDUCTION; ITERATIVE RECONSTRUCTION; IMAGE-RECONSTRUCTION; PERFORMANCE; CONTRAST; OPTIMIZATION;
D O I
10.1002/mp.16300
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
BackgroundIterative reconstruction (IR) has increasingly replaced traditional reconstruction methods in computed tomography (CT). The next paradigm shift in image reconstruction is likely to come from artificial intelligence, with deep learning reconstruction (DLR) solutions already entering the clinic. An enduring disadvantage to IR has been a change in noise texture, which can affect diagnostic confidence. DLR has demonstrated the potential to overcome this issue and has recently become available for dual-energy CT. PurposeTo evaluate the spatial resolution, noise properties, and detectability index of a commercially available DLR algorithm for dual-energy CT of the abdomen and compare it to single-energy (SE) CT. MethodsAn oval 25 cm x 35 cm custom-made phantom was scanned on a GE Revolution CT scanner (GE Healthcare, Waukesha, WI) at two dose levels (13 and 5 mGy) and two iodine concentrations (8 and 2 mg/mL), using three typical abdominal scan protocols: dual-energy (DE), SE 80 kV (SE-80 kV) and SE 120 kV (SE-120 kV). Reconstructions were performed with three strengths of IR (ASiR-V: AR0%, AR50%, AR100%) and three strengths of DLR (TrueFidelity: low, medium, high). The DE acquisitions were reconstructed as mono-energetic images between 40 and 80 keV. The noise power spectrum (NPS), task transfer function (TTF), and detectability index (d') were determined for the reconstructions following the recommendations of AAPM Task Group 233. ResultsNoise magnitude reductions (relative to AR0%) for the SE protocols were on average (-29%, -21%) for (AR50%, TF-M), while for DE-70 keV were (-28%, -43%). There was less reduction in mean frequency (f(av)) for DLR than for IR, with similar results for SE and DE imaging. There was, however, a substantial change in the NPS shape when using DE with DLR, quantifiable by a marked reduction in the peak frequency (f(peak)) that was absent in SE mode. All protocols and reconstructions (including AR0%) exhibited slight to moderate shifts towards lower spatial frequencies at the lower dose (<12% in f(av)). Spatial resolution was consistently superior for DLR compared to IR for SE but not for DE. All protocols and reconstructions (including AR0%) showed decreased resolution with reduced dose and iodine concentration, with less decrease for DLR compared to IR. DLR displayed a higher d' than IR. The effect of energy was large: d' increased with lower keV, and SE-80 kV had higher d' than SE-120 kV. Using DE with DLR could provide higher d' than SE-80 kV at the higher dose but not at lower dose. ConclusionsDE imaging with DLR maintained spatial resolution and reduced noise magnitude while displaying less change in noise texture than IR. The d' was also higher with DLR than IR, suggesting superiority in detectability of iodinated contrast. Despite these trends being consistent with those previously established for SE imaging, there were some noteworthy differences. For DE imaging there was no improvement in resolution compared to IR and a change in noise texture. DE imaging with low keV and DLR had superior detectability to SE DLR at the high dose but was not better than SE-80 kV at low dose.
引用
收藏
页码:2775 / 2786
页数:12
相关论文
共 50 条
  • [1] Noise and spatial resolution properties of a commercially available deep learning-based CT reconstruction algorithm
    Solomon, Justin
    Lyu, Peijei
    Marin, Daniele
    Samei, Ehsan
    MEDICAL PHYSICS, 2020, 47 (09) : 3961 - 3971
  • [2] Evaluation of Image Quality and Detectability of Deep Learning Image Reconstruction (DLIR) Algorithm in Single- and Dual-energy CT
    Zhong, Jingyu
    Shen, Hailin
    Chen, Yong
    Xia, Yihan
    Shi, Xiaomeng
    Lu, Wei
    Li, Jianying
    Xing, Yue
    Hu, Yangfan
    Ge, Xiang
    Ding, Defang
    Jiang, Zhenming
    Yao, Weiwu
    JOURNAL OF DIGITAL IMAGING, 2023, 36 (04) : 1390 - 1407
  • [3] Performance evaluation of deep learning image reconstruction algorithm for dual-energy spectral CT imaging: A phantom study
    Li, Haoyan
    Li, Zhentao
    Gao, Shuaiyi
    Hu, Jiaqi
    Yang, Zhihao
    Peng, Yun
    Sun, Jihang
    JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, 2024, 32 (03) : 513 - 528
  • [4] Evaluation of Image Quality and Detectability of Deep Learning Image Reconstruction (DLIR) Algorithm in Single- and Dual-energy CT
    Jingyu Zhong
    Hailin Shen
    Yong Chen
    Yihan Xia
    Xiaomeng Shi
    Wei Lu
    Jianying Li
    Yue Xing
    Yangfan Hu
    Xiang Ge
    Defang Ding
    Zhenming Jiang
    Weiwu Yao
    Journal of Digital Imaging, 2023, 36 : 1390 - 1407
  • [5] Cascaded systems analysis of noise and detectability in dual-energy cone-beam CT
    Gang, Grace J.
    Zbijewski, Wojciech
    Stayman, J. Webster
    Siewerdsen, Jeffrey H.
    MEDICAL PHYSICS, 2012, 39 (08) : 5145 - 5156
  • [6] Deep learning image reconstruction for improving image quality of contrast-enhanced dual-energy CT in abdomen
    Sato, Mineka
    Ichikawa, Yasutaka
    Domae, Kensuke
    Yoshikawa, Kazuya
    Kanii, Yoshinori
    Yamazaki, Akio
    Nagasawa, Naoki
    Nagata, Motonori
    Ishida, Masaki
    Sakuma, Hajime
    EUROPEAN RADIOLOGY, 2022, 32 (08) : 5499 - 5507
  • [7] Deep learning image reconstruction for improving image quality of contrast-enhanced dual-energy CT in abdomen
    Mineka Sato
    Yasutaka Ichikawa
    Kensuke Domae
    Kazuya Yoshikawa
    Yoshinori Kanii
    Akio Yamazaki
    Naoki Nagasawa
    Motonori Nagata
    Masaki Ishida
    Hajime Sakuma
    European Radiology, 2022, 32 : 5499 - 5507
  • [8] Impact of deep learning reconstructions on image quality and liver lesion detectability in dual-energy CT: An anthropomorphic phantom study
    Pauthe, Aurelie
    Milliner, Milan
    Pasquier, Hugo
    Campagnolo, Lucie
    Mule, Sebastien
    Luciani, Alain
    MEDICAL PHYSICS, 2025, : 2257 - 2268
  • [9] Image Quality Evaluation in Dual-Energy CT of the Chest, Abdomen, and Pelvis in Obese Patients With Deep Learning Image Reconstruction
    Fair, Eric
    Profio, Mark
    Kulkarni, Naveen
    Laviolette, Peter S.
    Barnes, Bret
    Bobholz, Samuel
    Levenhagen, Maureen
    Ausman, Robin
    Griffin, Michael O.
    Duvnjak, Petar
    Zorn, Adam P.
    Foley, W. Dennis
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2022, 46 (04) : 604 - 611
  • [10] Accuracy of liver metastasis detection and characterization: Dual-energy CT versus single-energy CT with deep learning reconstruction
    Jensen, Corey T.
    Wong, Vincenzo K.
    Wagner-Bartak, Nicolaus A.
    Liu, Xinming
    Sobha, Renjith Padmanabhan Nair
    Sun, Jia
    Likhari, Gauruv S.
    Gupta, Shiva
    EUROPEAN JOURNAL OF RADIOLOGY, 2023, 168