Towards safer imaging: A comparative study of deep learning-based denoising and iterative reconstruction in intraindividual low-dose CT scans using an in-vivo large animal model

被引:0
作者
Mueck, Jonas [1 ]
Reiter, Elisa [1 ]
Klingert, Wilfried [2 ]
Bertolani, Elisa [2 ]
Schenk, Martin [2 ]
Nikolaou, Konstantin [1 ]
Afat, Saif [1 ]
Brendlin, Andreas S. [1 ]
机构
[1] Eberhard Karls Univ Tubingen, Dept Diagnost & Intervent Radiol, Hoppe Seyler Str 3, D-72076 Tubingen, Germany
[2] Eberhard Karls Univ Tubingen, Dept Gen Visceral & Transplant Surg, Hoppe Seyler Str 3, D-72076 Tubingen, Germany
关键词
COMPUTED-TOMOGRAPHY; LIVER-LESIONS; NOISE; REDUCTION; ALGORITHM; CONTRAST; RISK;
D O I
10.1016/j.ejrad.2023.111267
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Computed tomography (CT) scans are a significant source of medically induced radiation exposure. Novel deep learning -based denoising (DLD) algorithms have been shown to enable diagnostic image quality at lower radiation doses than iterative reconstruction (IR) methods. However, most comparative studies employ low -dose simulations due to ethical constraints. We used real intraindividual animal scans to investigate the dose -reduction capabilities of a DLD algorithm in comparison to IR. Materials and methods: Fourteen veterinarian -sedated alive pigs underwent 2 CT scans on the same 3rd generation dual -source scanner with two months between each scan. Four additional scans ensued each time, with mAs reduced to 50 %, 25 %, 10 %, and 5 %. All scans were reconstructed ADMIRE levels 2 (IR2) and a novel DLD algorithm, resulting in 280 datasets. Objective image quality (CT numbers stability, noise, and contrast -to -noise ratio) was measured via consistent regions of interest. Three radiologists independently rated all possible dataset combinations per time point for subjective image quality (-1 = inferior, 0 = equal, 1 = superior). The points were averaged for a semiquantitative score, and inter -rater agreement was measured using Spearman's correlation coefficient and adequately corrected mixedeffects modeling analyzed objective and subjective image quality. Results: Neither dose -reduction nor reconstruction method negatively impacted CT number stability (p > 0.999). In objective image quality assessment, the lowest radiation dose achievable by DLD when comparing noise (p = 0.544) and CNR (p = 0.115) to 100 % IR2 was 25 %. Overall, inter -rater agreement of the subjective image quality ratings was strong (r >= 0.69, mean 0.93 +/- 0.05, 95 % CI 0.92-0.94; each p < 0.001), and subjective assessments corroborated that DLD at 25 % radiation dose was comparable to 100 % IR2 in image quality, sharpness, and contrast (p >= 0.281). Conclusions: The DLD algorithm can achieve image quality comparable to the standard IR method but with a significant dose reduction of up to 75%. This suggests a promising avenue for lowering patient radiation exposure without sacrificing diagnostic quality.
引用
收藏
页数:12
相关论文
共 47 条
  • [31] Assessment of low-dose paranasal sinus CT imaging using a new deep learning image reconstruction technique in children compared to adaptive statistical iterative reconstruction V (ASiR-V)
    Li, Yang
    Liu, Xia
    Zhuang, Xun-hui
    Wang, Ming-jun
    Song, Xiu-feng
    BMC MEDICAL IMAGING, 2022, 22 (01)
  • [32] Deep Learning-Based Reconstruction vs. Iterative Reconstruction for Quality of Low-Dose Head-and-Neck CT Angiography with Different Tube-Voltage Protocols in Emergency-Department Patients
    Lenfant, Marc
    Comby, Pierre-Olivier
    Guillen, Kevin
    Galissot, Felix
    Haioun, Karim
    Thay, Anthony
    Chevallier, Olivier
    Ricolfi, Frederic
    Loffroy, Romaric
    DIAGNOSTICS, 2022, 12 (05)
  • [33] Diagnosis of lumbar spinal fractures in emergency department: low-dose versus standard-dose CT using model-based iterative reconstruction
    Lee, Sun Hwa
    Yun, Seong Jong
    Jo, Hyeon Hwan
    Song, Jae Gwang
    CLINICAL IMAGING, 2018, 50 : 216 - 222
  • [34] Image quality and radiologists' subjective acceptance using model-based iterative and deep learning reconstructions as adjuncts to ultrahigh-resolution CT in low-dose contrast-enhanced abdominopelvic CT: phantom and clinical pilot studies
    Nishikawa, Makiko
    Machida, Haruhiko
    Shimizu, Yuta
    Kariyasu, Toshiya
    Morisaka, Hiroyuki
    Adachi, Takuya
    Nakai, Takehiro
    Sakaguchi, Kosuke
    Saito, Shun
    Matsumoto, Saki
    Koyanagi, Masamichi
    Yokoyama, Kenichi
    ABDOMINAL RADIOLOGY, 2022, 47 (02) : 891 - 902
  • [35] Radiation Dose Reduction for 80-kVp Pediatric CT Using Deep Learning-Based Reconstruction: A Clinical and Phantom Study
    Nagayama, Yasunori
    Goto, Makoto
    Sakabe, Daisuke
    Emoto, Takafumi
    Shigematsu, Shinsuke
    Oda, Seitaro
    Tanoue, Shota
    Kidoh, Masafumi
    Nakaura, Takeshi
    Funama, Yoshinori
    Uchimura, Ryutaro
    Takada, Sentaro
    Hayashi, Hidetaka
    Hatemura, Masahiro
    Hirai, Toshinori
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2022, 219 (02) : 315 - 324
  • [36] Impact of Deep Learning-based Optimization Algorithm on Image Quality of Low-dose Coronary CT Angiography with Noise Reduction: A Prospective Study
    Liu, Peijun
    Wang, Man
    Wang, Yining
    Yu, Min
    Wang, Yun
    Liu, Zhuoheng
    Li, Yumei
    Jin, Zhengyu
    ACADEMIC RADIOLOGY, 2020, 27 (09) : 1241 - 1248
  • [37] Usefulness of model-based iterative reconstruction in low-dose lumbar spine CT after spine surgery with metal implant: a phantom study
    Fukushima, Yasuhiro
    Matsuda, Akira
    Koizumi, Koji
    Honda, Maya
    Masamoto, Kazutaka
    Fujibayashi, Shunsuke
    ACTA RADIOLOGICA, 2021, 62 (10) : 1333 - 1340
  • [38] Low-contrast-dose liver CT using low monoenergetic images with deep learning-based denoising for assessing hepatocellular carcinoma: a randomized controlled noninferiority trial
    Bae, Jae Seok
    Lee, Jeong Min
    Kim, Se Woo
    Park, Sungeun
    Han, Seungchul
    Yoon, Jeong Hee
    Joo, Ijin
    Hong, Hyunsook
    EUROPEAN RADIOLOGY, 2023, 33 (06) : 4344 - 4354
  • [39] Feasibility of ultra-low radiation dose reduction for renal stone CT using model-based iterative reconstruction: prospective pilot study
    Kriegshauser, J. Scott
    Naidu, Sailen G.
    Paden, Robert G.
    He, Miao
    Wu, Qing
    Hara, Amy K.
    CLINICAL IMAGING, 2015, 39 (01) : 99 - 103
  • [40] Ultra-low-dose CT for left atrium and pulmonary veins imaging using new model-based iterative reconstruction algorithm
    Annoni, A. D.
    Andreini, D.
    Pontone, G.
    Formenti, A.
    Petulla, M.
    Consiglio, E.
    Nobili, E.
    Baggiano, A.
    Conte, E.
    Mushtaq, S.
    Bertella, E.
    Billi, F.
    Bartorelli, A. L.
    Montorsi, P.
    Pepi, M.
    EUROPEAN HEART JOURNAL-CARDIOVASCULAR IMAGING, 2015, 16 (12) : 1366 - 1373