Image quality improvement in bowtie-filter-equipped cone-beam CT using a dual-domain neural network

被引:1
作者
Yun, Sungho [1 ]
Jeong, Uijin [1 ]
Lee, Donghyeon [1 ]
Kim, Hyeongseok [2 ]
Cho, Seungryong [1 ,2 ,3 ,4 ]
机构
[1] Korea Adv Inst Sci & Technol KAIST, Dept Nucl & Quantum Engn, Daejeon 34141, South Korea
[2] Korea Adv Inst Sci & Technol KAIST, KAIST Inst Artificial Intelligence, Daejeon, South Korea
[3] Korea Adv Inst Sci & Technol KAIST, KAIST Inst Hlth Sci & Technol, Daejeon, South Korea
[4] Korea Adv Inst Sci & Technol KAIST, KAIST Inst IT Convergence, Daejeon, South Korea
基金
新加坡国家研究基金会;
关键词
bowtie-filter; cone-beam CT; dual-domain network; eclipse artifacts; MONTE-CARLO SIMULATIONS; CRESCENT ARTIFACTS; RADIATION; OPTIMIZATION; DETECTOR; CBCT; TOOL;
D O I
10.1002/mp.16693
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
BackgroundThe bowtie-filter in cone-beam CT (CBCT) causes spatially nonuniform x-ray beam often leading to eclipse artifacts in the reconstructed image. The artifacts are further confounded by the patient scatter, which is therefore patient-dependent as well as system-specific.PurposeIn this study, we propose a dual-domain network for reducing the bowtie-filter-induced artifacts in CBCT images.MethodsIn the projection domain, the network compensates for the filter-induced beam-hardening that are highly related to the eclipse artifacts. The output of the projection-domain network was used for image reconstruction and the reconstructed images were fed into the image-domain network. In the image domain, the network further reduces the remaining cupping artifacts that are associated with the scatter. A single image-domain-only network was also implemented for comparison.ResultsThe proposed approach successfully enhanced soft-tissue contrast with much-reduced image artifacts. In the numerical study, the proposed method decreased perceptual loss and root-mean-square-error (RMSE) of the images by 84.5% and 84.9%, respectively, and increased the structure similarity index measure (SSIM) by 0.26 compared to the original input images on average. In the experimental study, the proposed method decreased perceptual loss and RMSE of the images by 87.2% and 92.1%, respectively, and increased SSIM by 0.58 compared to the original input images on average.ConclusionsWe have proposed a deep-learning-based dual-domain framework to reduce the bowtie-filter artifacts and to increase the soft-tissue contrast in CBCT images. The performance of the proposed method has been successfully demonstrated in both numerical and experimental studies.
引用
收藏
页码:7498 / 7512
页数:15
相关论文
共 36 条
  • [1] Fluence field optimization for noise and dose objectives in CT
    Bartolac, Steven
    Graham, Sean
    Siewerdsen, Jeff
    Jaffray, David
    [J]. MEDICAL PHYSICS, 2011, 38 : S2 - S17
  • [2] Geant4-based Monte Carlo simulations on GPU for medical applications
    Bert, Julien
    Perez-Ponce, Hector
    El Bitar, Ziad
    Jan, Sebastien
    Boursier, Yannick
    Vintache, Damien
    Bonissent, Alain
    Morel, Christian
    Brasse, David
    Visvikis, Dimitris
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (16) : 5593 - 5611
  • [3] Radiation dose response of normal lung assessed by Cone Beam CT - A potential tool for biologically adaptive radiation therapy
    Bertelsen, Anders
    Schytte, Tine
    Bentzen, Soren M.
    Hansen, Olfred
    Nielsen, Morten
    Brink, Carsten
    [J]. RADIOTHERAPY AND ONCOLOGY, 2011, 100 (03) : 351 - 355
  • [4] Method for evaluating bow tie filter angle-dependent attenuation in CT: Theory and simulation results
    Boone, John M.
    [J]. MEDICAL PHYSICS, 2010, 37 (01) : 40 - 48
  • [5] Decoupling of bowtie and object effects for beam hardening and scatter artefact reduction in iterative cone-beam CT
    Cai, Meng
    Byrne, Mikel
    Archibald-Heeren, Ben
    Metcalfe, Peter
    Rosenfeld, Anatoly
    Wang, Yang
    [J]. PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2020, 43 (04) : 1161 - 1170
  • [6] Image artifacts caused by incorrect bowtie filters in cone-beam CT image-guided radiotherapy
    Cao, Yanan
    Ma, Tianjun
    de Boer, Steven F.
    Wang, Iris Z.
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2020, 21 (07): : 153 - 159
  • [7] CONE-BEAM CT FOR RADIOTHERAPY APPLICATIONS
    CHO, PS
    JOHNSON, RH
    GRIFFIN, TW
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (11) : 1863 - 1883
  • [8] Enhancement of soft-tissue contrast in cone-beam CT using an anti-scatter grid with a sparse sampling approach
    Cho, Sanghoon
    Lim, Sunho
    Kim, Changhwan
    Wi, Sunhee
    Kwon, Taejin
    Youn, Won Sik
    Lee, Sang Hyun
    Kang, Bo Sun
    Cho, Seungryong
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2020, 70 : 1 - 9
  • [9] Deep Learning for Low-Dose CT Denoising Using Perceptual Loss and Edge Detection Layer
    Gholizadeh-Ansari, Maryam
    Alirezaie, Javad
    Babyn, Paul
    [J]. JOURNAL OF DIGITAL IMAGING, 2020, 33 (02) : 504 - 515
  • [10] Crescent artifacts in cone-beam CT
    Giles, William
    Bowsher, James
    Li, Hao
    Yin, Fang-Fang