A nonparametric measure of noise in x-ray diagnostic images-mammography

被引:1
|
作者
Anton, M. [1 ]
Maeder, U. [2 ]
Schopphoven, S. [3 ]
Reginatto, M. [1 ]
机构
[1] Phys Tech Bundesanstalt Braunschweig & Berlin, Berlin, Germany
[2] Univ Appl Sci, Inst Med Phys & Radiat Protect, Giessen, Germany
[3] Reference Ctr Mammog Screening Southwest Germany, Giessen, Germany
关键词
quality assurance; mammography; noise measure; non-maximum suppression; nonparametric; PATIENT-SPECIFIC QUANTIFICATION; NO-REFERENCE; QUALITY ASSESSMENT; OBSERVER PERFORMANCE; AUTOMATED TECHNIQUE; PARAMETER;
D O I
10.1088/1361-6560/acb485
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Objective. In x-ray diagnostics, modern image reconstruction or image processing methods may render established methods of image quality assessment inadequate. Task specific quality assessment by using model observers has the disadvantage of being very labour-intensive. Therefore, it appears highly desirable to develop novel image quality parameters that neither rely on the linearity and the shift-invariace of the imaging system nor require the acquisition of hundreds of images as is necessary for the application of model observers, and which can be derived directly from diagnostic images. Approach. A new measure for the noise based on non-maximum-suppression images is defined and its properties are explored using simulated images before it is applied to an exposure series of mammograms of a homogeneous phantom and a 3D-printed breast phantom to demonstrate its usefulness under realistic conditions. Main results. The new noise parameter cannot only be derived from images with a homogeneous background but it can be extracted directly from images containing anatomic structures and is proportional to the standard deviation of the noise. At present, the applicability is restricted to mammography, which satisfies the assumption of short covariance length of the noise. Significance. The new measure of the noise is but a first step of the development of a set of parameters that are required to quantify image quality directly from diagnostic images without relying on the assumption of a linear, shift-invariant system, e.g. by providing measures of sharpness, contrast and structural complexity, in addition to the noise measure. For mammography, a convenient method is now available to quantify noise in processed diagnostic images.
引用
收藏
页数:16
相关论文
共 50 条
  • [31] A compliance testing program for diagnostic X-ray equipment
    Hutchinson, DE
    Cobb, BJ
    Jacob, CS
    APPLIED RADIATION AND ISOTOPES, 1999, 50 (01) : 237 - 245
  • [32] Review of challenges in medical diagnostic x-ray sources
    Price, JS
    ADVANCES IN LABORATORY-BASED X-RAY SOURCES AND OPTICS II, 2001, 4502 : 1 - 9
  • [33] Low-dose phase contrast mammography with conventional x-ray sources
    Olivo, A.
    Gkoumas, S.
    Endrizzi, M.
    Hagen, C. K.
    Szafraniec, M. B.
    Diemoz, P. C.
    Munro, P. R. T.
    Ignatyev, K.
    Johnson, B.
    Horrocks, J. A.
    Vinnicombe, S. J.
    Jones, J. L.
    Speller, R. D.
    MEDICAL PHYSICS, 2013, 40 (09)
  • [34] Quasi-monochromatic X-ray source for mammography via crystal array
    Gambaccini, M
    DelGuerra, A
    Taibi, A
    Tuffanelli, A
    PHYSICS OF MEDICAL IMAGING - MEDICAL IMAGING 1997, 1997, 3032 : 154 - 160
  • [35] Uncertainty estimation and statistical comparative methodology for mammography x-ray energy spectra
    Santoro-Fernandes, V
    Santos, J. C.
    Mariano, L.
    Vanin, V. R.
    Costa, P. R.
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2020, 6 (03):
  • [36] Multilayer x-ray detector for contrast-enhanced digital subtraction mammography
    Allec, Nicholas
    Karim, Karim S.
    MEDICAL IMAGING 2010: PHYSICS OF MEDICAL IMAGING, 2010, 7622
  • [37] Classification of compressed breast shapes for the design of equalization filters in x-ray mammography
    Goodsitt, MM
    Chan, HP
    Liu, B
    Guru, SV
    Morton, AR
    Keshavmurthy, S
    Petrick, N
    MEDICAL PHYSICS, 1998, 25 (06) : 937 - 948
  • [38] Investigation of Ghosting Recovery Mechanisms in Selenium X-ray Detector Structures for Mammography
    Mahmood, S. A.
    Kabir, M. Z.
    Tousignant, O.
    Greenspan, J.
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2012, 59 (03) : 597 - 604
  • [39] Laser-based microfocused x-ray source for mammography: Feasibility study
    Krol, A
    Ikhlef, A
    Kieffer, JC
    Bassano, DA
    Chamberlain, CC
    Jiang, Z
    Pepin, H
    Prasad, SC
    MEDICAL PHYSICS, 1997, 24 (05) : 725 - 732
  • [40] Microcontrolled pyro-electric instrument for measuring X-ray intensity in mammography
    de Paula, MH
    de Carvalho, AA
    Brassalotti, AL
    Alter, AJ
    Sakamoto, WK
    Malmonge, JA
    de Almeida, A
    MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2005, 43 (06) : 751 - 755