Validation of the fabricated cast nylon head phantom for stereotactic radiosurgery end-to-end test using alanine dosimeter

被引:1
|
作者
Intang, Aungsumalin [1 ,2 ]
Oonsiri, Puntiwa [3 ]
Kingkaew, Sakda [3 ]
Chatchumnan, Nichakan [3 ]
Oonsiri, Sornjarod [3 ]
机构
[1] Chulalongkorn Univ, Fac Med, Dept Radiol, Med Phys Program, Bangkok, Thailand
[2] Off Atoms Peace, Bangkok, Thailand
[3] King Chulalongkorn Mem Hosp, Dept Radiol, Div Radiat Oncol, Bangkok, Thailand
关键词
Alanine dosimeter; fabricated cast nylon head phantom; stereotactic radiosurgery; QUALITY-ASSURANCE PHANTOM; RADIOTHERAPY; ACCURACY;
D O I
10.4103/jmp.jmp_98_22
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: Stereotactic radiosurgery (SRS) is an alternative to surgery as it precisely delivers single-large doses to small tumors. Cast nylon is used in phantom due to its computed tomography (CT) number of about 56-95 HU, which is close to that of the soft tissue. Moreover, cast nylon is also more budget-friendly than the commercial phantoms. Aims: The aim of this study is to design and validate the fabricated cast nylon head phantom for SRS end-to-end test using an alanine dosimeter. Materials and Methods: The phantom was designed using cast nylon. It was initially created by a computer numerical control three-axis vertical machining center. Then, the cast nylon phantom was scanned using a CT simulator. Finally, the validation of the fabricated phantom using alanine dosimeter proficiency with four Varian LINAC machines was performed. Results: The fabricated phantom presented a CT number of 85-90 HU. The outcomes of VMAT SRS plans showed percentage dose differences from 0.24 to 1.55, whereas the percentage dose differences in organ at risk (OAR) were 0.09-10.80 due to the low-dose region. The distance between the target (position 2) and the brainstem (position 3) was 0.88 cm. Conclusions: Variation in dose for OAR is higher, which might be due to a high-dose gradient in the area where measurement was being conducted. The fabricated cast nylon end-to-end test head phantom had been suitably designed to image and irradiate during an end-to-end test for SRS using an alanine dosimeter.
引用
收藏
页码:74 / 79
页数:6
相关论文
共 45 条
  • [1] Development and Validation of End-to-End Test for Stereotactic Radiosurgery
    Jadhav, A. R., Sr.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2019, 105 (01): : E692 - E692
  • [2] Adaptation and validation of a commercial head phantom for cranial radiosurgery dosimetry end-to-end audit
    Dimitriadis, Alexis
    Palmer, Antony L.
    Thomas, Russell A. S.
    Nisbet, Andrew
    Clark, Catharine H.
    BRITISH JOURNAL OF RADIOLOGY, 2017, 90 (1074):
  • [3] End-to-End tests in linac based Stereotactic Radiosurgery and Radiotherapy using a new phantom
    Scheib, S.
    Schmelzer, P.
    Krayenbuehl, J.
    Lang, S.
    RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S112 - S113
  • [4] End-to-end validation of stereotactic radiosurgery delivery for multiple targets
    Gershkevitsh, E.
    Petrovic, B.
    Mahalwar, V.
    Slade-Carter, R.
    Blackmore, A.
    Saplaouras, A.
    RADIOTHERAPY AND ONCOLOGY, 2023, 182 : S469 - S470
  • [5] End-to-End Quality Assurance of Stereotactic Radiation Therapy Using an Anthropomorphic Head Phantom
    Alexander, Km
    Dekker, Kh
    Olding, T.
    Schreiner, Lj
    11TH INTERNATIONAL CONFERENCE ON 3D RADIATION DOSIMETRY, IC3DDOSE, 2022, 2167
  • [6] A Multi-Clinic Validation of An End-To-End Procedure for MLC-Based Stereotactic Radiosurgery with a Novel Phantom
    Brown, T.
    Fagerstrom, J.
    Beck, C.
    Holloway, C.
    Kerns, J.
    Kaurin, D.
    Kielar, K.
    MEDICAL PHYSICS, 2021, 48 (06)
  • [7] Towards An End-To-End Test for Image-Guided Stereotactic Radiotherapy Using the ArcCHECK Phantom
    Sauer, O.
    Groh, C.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [8] Position and dose n end-to-end test audit phantom for stereotactic radiotherapy
    Lehmann, J.
    Jones, R.
    Artschan, R.
    Thwaites, D.
    RADIOTHERAPY AND ONCOLOGY, 2015, 115 : S851 - S851
  • [9] In-house Production of a Head Phantom for Testing the radiotherapy Chain (end-to-end test)
    Kurowski, M.
    Loot, K.
    Waletzko, O.
    Rohn, R.
    Block, A.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2021, 197 (SUPPL 1) : S222 - S223
  • [10] An End-To-End VMAT QA Procedure Using a Customized Anthropomorphic Phantom Head
    Iwai, P.
    Hara, S.
    MEDICAL PHYSICS, 2017, 44 (06) : 2918 - 2918