An empirical method for geometric calibration of a photon counting detector-based cone beam CT system

被引:0
作者
Ghani, Muhammad Usman [1 ]
Makeev, Andrey [1 ]
Manus, Joseph A. [1 ]
Glick, Stephen J. [1 ]
Ghammraoui, Bahaa [1 ]
机构
[1] United States Food & Drug Adm, Div Imaging Diagnost & Software Reliabil, Ctr Devices & Radiol Hlth, Off Sci & Engn Labs, Silver Spring, MD 20993 USA
关键词
Geometric calibration; sphericity; symmetry; photon counting detectors; cone beam CTs; COMPUTED-TOMOGRAPHY; PARAMETERS;
D O I
10.3233/XST-230007
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
BACKGROUND: Geometric calibration is essential in developing a reliable computed tomography (CT) system. It involves estimating the geometry under which the angular projections are acquired. Geometric calibration of cone beam CTs employing small area detectors, such as currently available photon counting detectors (PCDs), is challenging when using traditional-based methods due to detectors' limited areas. OBJECTIVE: This study presented an empirical method for the geometric calibration of small area PCD-based cone beam CT systems. METHODS: Unlike the traditional methods, we developed an iterative optimization procedure to determine geometric parameters using the reconstructed images of small metal ball bearings (BBs) embedded in a custom-built phantom. An objective function incorporating the sphericities and symmetries of the embedded BBs was defined to assess performance of the reconstruction algorithm with the given initial estimated set of geometric parameters. The optimal parameter values were those which minimized the objective function. The TIGRE toolbox was employed for fast tomographic reconstruction. To evaluate the proposed method, computer simulations were carried out using various numbers of spheres placed in various locations. Furthermore, efficacy of the method was experimentally assessed using a custom-made benchtop PCD-based cone beam CT. RESULTS: Computer simulations validated the accuracy and reproducibility of the proposed method. The precise estimation of the geometric parameters of the benchtop revealed high-quality imaging in CT reconstruction of a breast phantom. Within the phantom, the cylindrical holes, fibers, and speck groups were imaged in high fidelity. The CNR analysis further revealed the quantitative improvements of the reconstruction performed with the estimated parameters using the proposed method. CONCLUSION: Apart from the computational cost, we concluded that the method was easy to implement and robust.
引用
收藏
页码:865 / 877
页数:13
相关论文
共 50 条
  • [41] Development of a Spectral Micro-CT System using a Photon Counting Detector with Anti-Coincidence Corrections
    Holbrook, M. D.
    Clark, D. P.
    Badea, C. T.
    MEDICAL IMAGING 2020: PHYSICS OF MEDICAL IMAGING, 2020, 11312
  • [42] Radiation dosimetry of a clinical prototype dedicated cone-beam breast CT system with offset detector
    Tseng, Hsin Wu
    Karellas, Andrew
    Vedantham, Srinivasan
    MEDICAL PHYSICS, 2021, 48 (03) : 1079 - 1088
  • [43] A Novel Method of Cone Beam CT Projection Binning Based on Image Registration
    Park, Seonyeong
    Kim, Siyong
    Yi, Byongyong
    Hugo, Geoffrey
    Gach, H. Michael
    Motai, Yuichi
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2017, 36 (08) : 1733 - 1745
  • [44] Line-based iterative geometric calibration method for a tomosynthesis system
    Choi, Chloe J.
    Vent, Trevor L.
    Acciavatti, Raymond J.
    Maidment, Andrew D. A.
    MEDICAL PHYSICS, 2024, 51 (04) : 2444 - 2460
  • [45] A practical and robust method for beam blocker-based cone beam CT scatter correction
    Cui, Hehe
    Jiang, Xiao
    Tang, Wei
    Lu, Hsiao-Ming
    Yang, Yidong
    PHYSICS IN MEDICINE AND BIOLOGY, 2023, 68 (04)
  • [46] Ultra-low-dose photon-counting CT of paranasal sinus: an in vivo comparison of radiation dose and image quality to cone-beam CT
    Kaatsch, Hanns Leonhard
    Fulisch, Florian
    Dillinger, Daniel
    Kubitscheck, Laura
    Becker, Benjamin V.
    Piechotka, Joel
    Brockmann, Marc A.
    Froelich, Matthias F.
    Schoenberg, Stefan O.
    Overhoff, Daniel
    Waldeck, Stephan
    DENTOMAXILLOFACIAL RADIOLOGY, 2024, 53 (01) : 103 - 108
  • [47] Ultra-low-dose photon-counting CT of paranasal sinus: an in vivo comparison of radiation dose and image quality to cone-beam CT
    Kaatsch, Hanns Leonhard
    Fulisch, Florian
    Dillinger, Daniel
    Kubitscheck, Laura
    Becker, Benjamin, V
    Piechotka, Joel
    Brockmann, Marc A.
    Froelich, Matthias F.
    Schoenberg, Stefan O.
    Overhoff, Daniel
    Waldeck, Stephan
    DENTOMAXILLOFACIAL RADIOLOGY, 2024, 53 (02) : 103 - 108
  • [48] A High-Resolution Imaging Technique using a Whole-body, Research Photon Counting Detector CT System
    Leng, S.
    Yu, Z.
    Halaweish, A.
    Kappler, S.
    Hahn, K.
    Henning, A.
    Li, Z.
    Lane, J.
    Levin, D. L.
    Jorgensen, S.
    Ritman, E.
    McCollough, C.
    MEDICAL IMAGING 2016: PHYSICS OF MEDICAL IMAGING, 2016, 9783
  • [49] Evaluation of conventional imaging performance in a research whole-body CT system with a photon-counting detector array
    Yu, Zhicong
    Leng, Shuai
    Jorgensen, Steven M.
    Li, Zhoubo
    Gutjahr, Ralf
    Chen, Baiyu
    Halaweish, Ahmed F.
    Kappler, Steffen
    Yu, Lifeng
    Ritman, Erik L.
    McCollough, Cynthia H.
    PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (04) : 1572 - 1595
  • [50] Energy Calibration of a Silicon-Strip Detector for Photon-Counting Spectral CT by Direct Usage of the X-ray Tube Spectrum
    Liu, Xuejin
    Chen, Han
    Bornefalk, Hans
    Danielsson, Mats
    Karlsson, Staffan
    Persson, Mats
    Xu, Cheng
    Huber, Ben
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2015, 62 (01) : 68 - 75