Analytical performance evaluation of two automated urine chemistry analyzers using two levels of commercially available quality control material

被引:0
作者
Hokamp, J. A. [1 ]
Freeman, K. [2 ]
Harr, K. E. [3 ]
机构
[1] Ohio State Univ, Coll Vet Med, Dept Vet Biosci, Columbus, OH 43210 USA
[2] Syn Labs VPG Exeter, Exeter, England
[3] URIKA LLC, Mukilteo, WA USA
关键词
ESTERASE TEST-STRIP; CANINE MICROALBUMINURIA; PH-METER; DIPSTICK; ASSURANCE; URINALYSIS; CATS;
D O I
10.1111/jsap.13620
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
Objective: Evaluate two point-of-care urine chemistry analyzers, VetScan SA and VetLab UA using assayed, bilevel (two concentrations) urine quality control material to determine if performance is acceptable for semiquantitative clinical urine chemistry analysis. Materials and Methods: Normal and abnormal urine quality control material sent to 23 veterinary practices was evaluated three times by each clinic on in--clinic automated urinalysis instruments. Accuracy, precision and clinical utility were evaluated. Results: Normal urine quality control material: Results for blood, glucose, ketones and bilirubin were 100% accurate and precise for both analyzers, and pH values were accurately acidic to neutral. However, pH from VetScan SA had clinically significant negative bias. Abnormal urine quality control material: VetScan SA: blood, microalbumin and bilirubin were 100% accurate; glucose, ketones, and protein demonstrated <= 10% inaccuracy; pH demonstrated 34% inaccuracy. VetLab UA: blood, ketones and bilirubin were 100% accurate; glucose and protein demonstrated <= 10% inaccuracy; pH was 100% accurately neutral to alkaline. Clinical Significance: VetScan SA had marked negative pH bias versus VetLab UA resulting in clinically significant, overly acidic results. Specific gravity, nitrite, and leukocyte test pads should not be used. Both instruments had excellent performance in normal quality control material. While blood, glucose, protein and bilirubin are correctly identified as present in abnormal quality control material, exact concentrations cannot be interpreted due to imprecision. Only semiquantitative results, not numerical values implying quantification, should be reported from urine test strips.
引用
收藏
页码:452 / 462
页数:11
相关论文
共 43 条
[1]  
Abaxis, 2017, VETSCAN UA10 UR TEST
[2]  
Abaxis, 2017, VETSCAN UA14 UR TEST
[3]  
[Anonymous], 2019, Manual of Diagnostic Tests and Vaccines for Terrestrial Animals, V8th
[4]  
[Anonymous], 2000, SCAND J CLIN LAB I S, V231, P1, DOI DOI 10.1080/00365513.2000.12056993
[5]   Issue Information [J].
Arnold, Jill E. ;
Camus, Melinda S. ;
Freeman, Kathy P. ;
Giori, Luca ;
Hooijberg, Emma H. ;
Jeffery, Unity ;
Korchia, Jeremie ;
Meindel, Mandy J. ;
Moore, A. Russell ;
Sisson, Sandra C. ;
Vap, Linda M. ;
Cook, Jennifer R. .
VETERINARY CLINICAL PATHOLOGY, 2019, 48 (04) :542-618
[6]   Comparison between the urine dipstick and the pH-meter to assess urine pH in sheep and dogs [J].
Athanasiou, Labrini V. ;
Katsoulos, Panagiotis D. ;
Katsogiannou, Eleni G. ;
Polizopoulou, Zoe S. ;
Diamantaki, Myrto ;
Kamatsos, Constantinos ;
Christodoulopoulos, Georgios .
VETERINARY CLINICAL PATHOLOGY, 2018, 47 (02) :284-288
[7]   Feline Calcium Oxalate Urolithiasis: Risk factors and rational treatment approaches [J].
Bartges, Joseph W. .
JOURNAL OF FELINE MEDICINE AND SURGERY, 2016, 18 (09) :712-722
[8]   Evaluation the Clinitek status™ automated dipstick analysis device for semiquantitative testing of canine urine [J].
Bauer, N. ;
Rettig, S. ;
Moritz, A. .
RESEARCH IN VETERINARY SCIENCE, 2008, 85 (03) :467-472
[9]  
BIO-RAD, 2019, BIOR LIQ UR CONTR LE
[10]   ASVCP quality assurance guidelines: external quality assessment and comparative testing for reference and in-clinic laboratories [J].
Camus, Melinda S. ;
Flatland, Bente ;
Freeman, Kathleen P. ;
Cardona, Janice A. Cruz .
VETERINARY CLINICAL PATHOLOGY, 2015, 44 (04) :477-492