Hemodynamic study of blood flow in the aorta during the interventional robot treatment using fluid-structure interaction

被引:1
|
作者
Zhu, Zongming [1 ]
Ji, Suqiang [2 ]
Liang, Liang [1 ]
Wang, Hao [3 ]
Xia, Haoyu [4 ]
Tang, Puhua [1 ]
机构
[1] Changsha Univ, Coll Electromech Engn, Changsha 410022, Peoples R China
[2] Xiangtan Univ, Sch Mech Engn & Mech, Xiangtan 411105, Peoples R China
[3] Hunan Normal Univ, Coll Engn & Design, Changsha 410081, Peoples R China
[4] Changsha Inst Min Res Co LTD, Changsha 410012, Peoples R China
基金
中国国家自然科学基金;
关键词
Vascular interventional robot; Fluid-structure interaction (FSI); Hemodynamics; Computational fluid dynamics (CFD); Particle image velocimetry (PIV); DISSECTION; STENTS;
D O I
10.1007/s10237-023-01737-y
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
An interventional robot is a means for vascular diagnosis and treatment, and it can perform dredging, releasing drug and operating. Normal hemodynamic indicators are a prerequisite for the application of interventional robots. The current hemodynamic research is limited to the absence of interventional devices or interventional devices in fixed positions. Considering the coupling effect of blood, vessels and robots, based on the bi-directional fluid-structure interaction, using the computational fluid dynamics and particle image velocimetry methods, combined with the sliding and moving mesh technologies, we theoretically and experimentally study the hemodynamic indicators such as blood flow lines, blood pressure, equivalent stress, deformation and wall shear stress of blood vessels when the robot precesses, rotates or does not intervene in the pulsating blood flow. The results show that the intervention of the robot increase the blood flow rate, blood pressure, equivalent stress and deformation of the vessels by 76.4%, 55.4%, 76.5%, and 346%, respectively. The operating mode of the robot during low-speed operation has little impact on the hemodynamic indicators. Using the methyl silicone oil as the experimental fluid, the elastic silicone pipe as the experimental pipe, and the intervention robot having a bioplastic outer shell, the velocity of the fluid around the robot is measured on the developed experimental device for fluid flow field in a pulsating flow when the robot runs. The experimental results are similar to the numerical results. Our work provides an important reference for the hemodynamic study and optimization of the mobile interventional devices.
引用
收藏
页码:1857 / 1872
页数:16
相关论文
共 50 条
  • [31] A Study of the Fluid-Structure Interaction of the Plaque Circumferential Distribution in the Left Coronary Artery
    Du, Yepeng
    Zhang, Lili
    Hou, Zhanju
    Liu, Jian
    APPLIED SCIENCES-BASEL, 2022, 12 (12):
  • [32] Effects of renal artery stenosis on realistic model of abdominal aorta and renal arteries incorporating fluid-structure interaction and pulsatile non-Newtonian blood flow
    Mortazavinia, Z.
    Zare, A.
    Mehdizadeh, A.
    APPLIED MATHEMATICS AND MECHANICS-ENGLISH EDITION, 2012, 33 (02) : 165 - 176
  • [33] Effects of renal artery stenosis on realistic model of abdominal aorta and renal arteries incorporating fluid-structure interaction and pulsatile non-Newtonian blood flow
    Z. Mortazavinia
    A. Zare
    A. Mehdizadeh
    Applied Mathematics and Mechanics, 2012, 33 : 165 - 176
  • [34] Numerical Simulation of Fluid-Structure Interaction Under the Condition of Pulsatile Blood Flow of Renal Artery with Radiofrequency Electrode
    Nan, Qun
    Cheng, Yanyan
    Tian, Zhen
    Dong, Tong
    Gao, Xiang
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING 2018, VOL 1, 2019, 68 (01): : 515 - 518
  • [35] Modeling and evaluation of biomechanics and hemodynamic based on patient-specific small intracranial aneurysm using fluid-structure interaction
    Ma, Zijian
    Zhuang, Yijie
    Long, Xiaoao
    Yu, Bo
    Li, Jiawang
    Yang, Yan
    Yu, Yingxin
    COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 2024, 244
  • [36] Blood vessel tissue prestress modeling for vascular fluid-structure interaction simulation
    Hsu, Ming-Chen
    Bazilevs, Yuri
    FINITE ELEMENTS IN ANALYSIS AND DESIGN, 2011, 47 (06) : 593 - 599
  • [37] A Fluid-Structure Interaction Analysis of Blood Clot Motion in a Branch of Pulmonary Arteries
    Fateme Mirakhorli
    Bahman Vahidi
    Marzieh Pazouki
    Pouria Talebi Barmi
    Cardiovascular Engineering and Technology, 2023, 14 : 79 - 91
  • [38] Noninvasive estimation of central blood pressure through fluid-structure interaction modeling
    Wu, Peishuo
    Zhu, Chi
    BIOMECHANICS AND MODELING IN MECHANOBIOLOGY, 2024, : 423 - 439
  • [39] A Fluid-Structure Interaction Analysis of Blood Clot Motion in a Branch of Pulmonary Arteries
    Mirakhorli, Fateme
    Vahidi, Bahman
    Pazouki, Marzieh
    Barmi, Pouria Talebi
    CARDIOVASCULAR ENGINEERING AND TECHNOLOGY, 2023, 14 (01) : 79 - 91
  • [40] The order of precedence in treatment of multiple intracranial aneurysms: insights from a fluid-structure interaction study
    Klis, Kornelia M.
    Gasowski, Jerzy
    Cierniak, Antoni
    Kwinta, Borys M.
    Stachura, Krzysztof
    Popiela, Tadeusz J.
    Szydlowski, Igor
    Lasocha, Bartlomiej
    Piotrowicz, Karolina
    Grodzicki, Tomasz
    Krzyzewski, Roger M.
    BIOMECHANICS AND MODELING IN MECHANOBIOLOGY, 2025, : 589 - 598