Validation testing of five home blood pressure monitoring devices for the upper arm according to the ISO 81060-2:2018/AMD 1:2020 protocol

被引:11
作者
Peprah, Yaw Amofa [1 ]
Lee, Ji Young [1 ]
Persell, Stephen D. D. [1 ,2 ]
机构
[1] Northwestern Univ, Feinberg Sch Med, Dept Med, Div Gen Internal Med, Chicago, IL 60611 USA
[2] Northwestern Univ, Inst Publ Hlth & Med, Feinberg Sch Med, Ctr Primary Care Innovat, Chicago, IL USA
基金
美国国家卫生研究院;
关键词
TASK-FORCE; ASSOCIATION; SOCIETY;
D O I
10.1038/s41371-022-00795-6
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
The accuracy of Omron 10 Series BP7450 (HEM-7342T-Z), Omron Platinum BP5450 (HEM-7343T-Z), Walmart Equate Premium 8000 Series UA-8000WM, Walgreens Premium 15+ WGNBPA-960BT, and CVS Series 800 BP3MW1-4YCVS were assessed in an adult general population compared to a mercury sphygmomanometer standard according to the ISO 81060-2:2018/AMD 1:2020 validation procedure. Omron selected the monitors and included three non-Omron monitors because they were from large retail vendors in the United States and these monitors did not have accessible results from validation testing. The BP7450, N = 85, passed both criteria for the standard. Mean (SD) differences in paired SBP and DBP determinations between the test device and reference were 0.5 (7.7) and 2.5 (6.8) mm Hg. The BP5450, N = 86, passed both criteria. Mean (SD) differences in paired SBP and DBP determinations were 1.9 (7.0) and 3.6 (6.4) mm Hg. The UA-8000WM, N = 85, did not meet the first criterion for the standard. Mean (SD) differences in paired SBP and DBP determinations were 2.5 (8.0) and 5.1 (6.4) mm Hg. The WGNBPA-240BT, N = 85, did not meet the first criterion for the standard. Mean (SD) differences in paired SBP and DBP determinations were 7.9 (8.5) and 5.5 (6.7) mm Hg. The BP3MW1-4YCVS, N = 85, did not meet the first criterion for the standard. Mean (SD) differences in paired SBP and DBP determinations were 5.8 (8.7) and 3.1 (5.6) mm Hg. These findings emphasize the importance of verifying the validation status of home blood pressure monitors before use by consumers.
引用
收藏
页码:134 / 140
页数:7
相关论文
共 17 条
[1]   Home sphygmomanometers: validation versus accuracy [J].
Akpolat, Tekin ;
Dilek, Melda ;
Aydogdu, Turkan ;
Adibelli, Zelal ;
Erdem, Dilek Gurgenyatagi ;
Erdem, Emre .
BLOOD PRESSURE MONITORING, 2009, 14 (01) :26-31
[2]  
[Anonymous], 2018, 8106022019 ANSIAAMII
[3]  
[Anonymous], 2020, ISO 81060-2:2018+Amd.1:2020
[4]   Accuracy of blood-pressure monitors owned by patients with hypertension (ACCU-RATE study ): a cross-sectional, observational study in central England [J].
Hodgkinson, James A. ;
Lee, Mei-Man ;
Milner, Siobhan ;
Bradburn, Peter ;
Stevens, Richard ;
Hobbs, Fd Richard ;
Koshiaris, Constantinos ;
Grant, Sabrina ;
Mant, Jonathan ;
McManus, Richard J. .
BRITISH JOURNAL OF GENERAL PRACTICE, 2020, 70 (697) :E548-E554
[5]   Reliability of home blood pressure monitoring: in the context of validation and accuracy [J].
Jung, Mi-Hyang ;
Kim, Gee-Hee ;
Kim, Ji-Hoon ;
Moon, Keon-Woong ;
Yoo, Ki-Dong ;
Rho, Tae-Ho ;
Kim, Chul-Min .
BLOOD PRESSURE MONITORING, 2015, 20 (04) :215-220
[6]   Sources of inaccuracy in themeasurement of adult patients' resting blood pressure in clinical settings: a systematic review [J].
Kallioinen, Noa ;
Hill, Andrew ;
Horswill, Mark S. ;
Ward, Helen E. ;
Watson, Marcus O. .
JOURNAL OF HYPERTENSION, 2017, 35 (03) :421-441
[7]  
Medaval Ltd, 2022, MED VAL MED DEV REG
[8]   Validation Status of Blood Pressure Measuring Devices Sold Globally [J].
Picone, Dean S. ;
Campbell, Norm R. C. ;
Schutte, Aletta E. ;
Olsen, Michael Hecht ;
Ordunez, Pedro ;
Whelton, Paul K. ;
Sharman, James E. .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2022, 327 (07) :680-681
[9]   How to check whether a blood pressure monitor has been properly validated for accuracy [J].
Picone, Dean S. ;
Padwal, Raj ;
Campbell, Norm R. C. ;
Boutouyrie, Pierre ;
Brady, Tammy M. ;
Olsen, Michael Hecht ;
Delles, Christian ;
Lombardi, Cintia ;
Mahmud, Azra ;
Meng, Yaxing ;
Mokwatsi, Gontse G. ;
Ordunez, Pedro ;
Phan, Hoang T. ;
Pucci, Giacomo ;
Schutte, Aletta E. ;
Sung, Ki-Chul ;
Zhang, Xin-Hua ;
Sharman, James E. .
JOURNAL OF CLINICAL HYPERTENSION, 2020, 22 (12) :2167-2174
[10]   Nonvalidated Home Blood Pressure Devices Dominate the Online Marketplace in Australia Major Implications for Cardiovascular Risk Management [J].
Picone, Dean S. ;
Deshpande, Rewati A. ;
Schultz, Martin G. ;
Fonseca, Ricardo ;
Campbell, Norm R. C. ;
Delles, Christian ;
Hecht Olsen, Michael ;
Schutte, Aletta E. ;
Stergiou, George ;
Padwal, Raj ;
Zhang, Xin-Hua ;
Sharman, James E. .
HYPERTENSION, 2020, 75 (06) :1593-1599