No higher tibial post fracture rate with highly cross-linked polyethylene inserts in total knee arthroplasty at a minimum 10-year follow-up

被引:1
作者
Baek, Ji-Hoon [1 ]
Lee, Su Chan [1 ]
Ryu, Suengryol [2 ]
Ahn, Hye Sun [1 ]
Nam, Chang Hyun [1 ]
机构
[1] Himchan Hosp, Dept Orthopaed Surg, Joint & Arthrit Res, 120 Sinmok Ro, Seoul, South Korea
[2] Himnaera Hosp, Dept Orthopaed Surg, Busan, South Korea
关键词
Highly cross-linked polyethylene; Posterior-stabilized; Total knee arthroplasty; Tibial post fracture; WEAR;
D O I
10.1007/s00402-023-04789-y
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
PurposeThe purpose of this study was to assess tibial post fracture between highly cross-linked polyethylene (HXPE) and conventional polyethylene inserts in consecutive posterior-stabilized (PS) total knee arthroplasty (TKA) over a minimum 10-years follow-up period.MethodsBetween July 2008 and December 2011, a consecutive series of 3264 primary TKAs was performed in 2098 patients using a PS total knee prosthesis at a single institution. The final cohort was classified into the HXPE group (792 cases) and the conventional group (2113 cases) according to insert material used during the procedure.ResultsThe mean follow-up period was 11.4 years in the HXPE group and 11.8 years in the conventional group. Both cases of tibial post fracture occurred in the HXPE group, a complication rate of 0.25%. These two cases required replacement with a thicker tibial insert with retention of the femoral and tibial components. There were no cases of tibial post fracture in the conventional group.ConclusionThere was no difference in tibial post fracture rate between the HXPE group and the conventional group in PS TKA on long-term follow-up.
引用
收藏
页码:5807 / 5813
页数:7
相关论文
共 17 条
  • [1] Fifteen-Year to 19-Year Follow-Up of the Insall-Burstein-1 Total Knee Arthroplasty
    Abdeen, Ayesha R.
    Collen, Stacy B.
    Vince, Kelly G.
    [J]. JOURNAL OF ARTHROPLASTY, 2010, 25 (02) : 173 - 178
  • [2] High rate of tibial post-fracture in highly cross-linked polyethylene compared to conventional polyethylene in posterior-stabilized total knee arthroplasty at a minimum 12-year follow-up
    Baek, Ji-Hoon
    Lee, Su Chan
    Choi, Kyungwon
    Nam, Chang Hyun
    [J]. KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2022, 30 (12) : 4072 - 4077
  • [3] Long-term survivorship of total knee arthroplasty with a single-radius, high-flexion posterior stabilized prosthesis
    Baek, Ji-Hoon
    Lee, Su Chan
    Choi, Kyungwon
    Ahn, Hye Sun
    Nam, Chang Hyun
    [J]. KNEE, 2021, 30 : 275 - 282
  • [4] Ten-year results for a single-surgeon series of Scorpio non-restrictive geometry (NRG) posterior stabilised (PS) total knee replacement
    Bajada, Stefan
    Searle, David
    Toms, Andrew D.
    [J]. EUROPEAN JOURNAL OF ORTHOPAEDIC SURGERY AND TRAUMATOLOGY, 2019, 29 (07) : 1533 - 1538
  • [5] Five cases of tibial post fracture in posterior stabilized total knee arthroplasty using Prolong highly cross-linked polyethylene
    Diamond, Owen J.
    Howard, Lisa
    Masri, Bassam
    [J]. KNEE, 2018, 25 (04) : 657 - 662
  • [6] Implant Design Influences Tibial Post Wear Damage in Posterior-stabilized Knees
    Dolan, Mark M.
    Kelly, Natalie H.
    Nguyen, Joseph T.
    Wright, Timothy M.
    Haas, Steven B.
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2011, 469 (01) : 160 - 167
  • [7] Tibial Post Wear in Posterior-stabilized Knee Replacements is Design-dependent
    Furman, Bridgette D.
    Lipman, Joseph
    Kligman, Mordechai
    Wright, Timothy M.
    Haas, Steven B.
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2008, 466 (11) : 2650 - 2655
  • [8] Fracture of a Second-generation Highly Cross-linked UHMWPE Tibial Post in a Posterior-stabilized Scorpio Knee System
    Jung, Kwang Am
    Lee, Su Chan
    Hwang, Seung Hyun
    Kim, Sung Min
    [J]. ORTHOPEDICS, 2008, 31 (11) : 1137 - 1139
  • [9] Jung Kwang Am, 2009, J Knee Surg, V22, P374
  • [10] Knee bearing technology
    Kester, Mark A.
    Herrera, Lizeth
    Aiguo, Wang
    Essner, Aaron