Improvements in image quality after optimization in digital intraoral radiographs

被引:3
作者
Barrineau, Ashley B. [1 ]
Mah, Peter [2 ]
Mallya, Sanjay M. [3 ]
机构
[1] Univ Calif Los Angeles, Sch Dent, Sect Endodont, Los Angeles, CA USA
[2] Dent Imaging Consultants, Box 29214 Med Ctr Stn, San Antonio, TX 78229 USA
[3] Univ Calif Los Angeles, Sch Dent, Sect Maxillofacial Radiol, Los Angeles, CA USA
关键词
Quality control; quality assurance; image optimization; digital radiography; exposure optimization; radiographic phantom; image analysis; ASSURANCE; ACCURACY;
D O I
10.1016/j.adaj.2022.09.008
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Background. Digital intraoral radiographic exposures are optimized largely on the basis of subjective assessment of diagnostic image quality. This study presents an objective approach to optimize radiographic exposure settings for digital intraoral radiographic systems.Methods. Seven size 2 digital intraoral systems were assessed for image quality and determination of optimal exposure following the protocol specified in American National Standard Institute/American Dental Association Standard No. 1094: Quality Assurance for Digital Intra-Oral Radiographic Systems. A ProX radiograph unit (Planmeca) at 63 kVp and 6 mA was used to obtain radiographs of the Dental Digital Quality Assurance phantom. ImageJ software (National Institutes of Health) was used to quantify dynamic range and spatial resolution, and contrast perceptibility was evaluated visually. Optimal exposure is the setting with the maximal contrast perceptibility and spatial resolution while displaying the full dynamic range. After image optimization, a custom phantom consisting of an endodontically prepared tooth was imaged to evaluate the file position relative to the apex for each system. Differences in distances between file position relative to the root apex at the optimal exposure as well as 1 increment above and below were measured.Results. Radiographic images obtained at the optimal exposure yielded better visualization and more accurate measurements of the file tip relative to the apex.Conclusions. Optimizing radiographic exposures improves image quality and accuracy in clinical decisions.Practical Implications. Improvement in image quality and better accuracy in actual distance of the endodontic file to the radiographic apex coupled with complete cleaning, shaping, and obturation of the canal should lead to better endodontic treatment outcomes.
引用
收藏
页码:24 / 31
页数:8
相关论文
共 50 条
[31]   Use of Texture Feature Maps for the Refinement of Information Derived from Digital Intraoral Radiographs of Lytic and Sclerotic Lesions [J].
Obuchowicz, Rafal ;
Nurzynska, Karolina ;
Obuchowicz, Barbara ;
Urbanik, Andrzej ;
Piorkowski, Adam .
APPLIED SCIENCES-BASEL, 2019, 9 (15)
[32]   Comparison of digital intraoral scanner reproducibility and image trueness considering repetitive experience [J].
Lim, Jung-Hwa ;
Park, Ji-Man ;
Kim, Minji ;
Heo, Seong-Joo ;
Myung, Ji-Yun .
JOURNAL OF PROSTHETIC DENTISTRY, 2018, 119 (02) :225-232
[33]   A Patient Image-based Technique to Assess the Image Quality of Clinical Chest Radiographs [J].
Lin, Yuan ;
Samei, Ehsan ;
Luo, Hui ;
Dobbins, James T., III ;
McAdams, H. Page ;
Wang, Xiaohui ;
Sehnert, William J. ;
Barski, Lori ;
Foos, David H. .
MEDICAL IMAGING 2011: PHYSICS OF MEDICAL IMAGING, 2011, 7961
[34]   Digital radiography: optimization of image quality and dose using multi-frequency software [J].
H. Precht ;
O. Gerke ;
K. Rosendahl ;
A. Tingberg ;
D. Waaler .
Pediatric Radiology, 2012, 42 :1112-1118
[35]   Digital radiography: optimization of image quality and dose using multi-frequency software [J].
Precht, H. ;
Gerke, O. ;
Rosendahl, K. ;
Tingberg, A. ;
Waaler, D. .
PEDIATRIC RADIOLOGY, 2012, 42 (09) :1112-1118
[36]   Image quality evaluation of a highly-integrated digital X-ray imaging sensor for dental intraoral-imaging applications [J].
Cho, H. S. ;
Choi, S. G. ;
Lee, B. S. ;
Kim, S. .
JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2007, 51 (01) :30-34
[37]   Image Quality in Digital Industrial Radiography [J].
Ewert, Uwe ;
Zscherpel, Uwe ;
Heyne, Klara ;
Jechow, Mirko ;
Bavendiek, Klaus .
MATERIALS EVALUATION, 2012, 70 (08) :955-964
[38]   Clinical comparison of intraoral CMOS and PSP detectors in terms of time efficiency, patient comfort, and subjective image quality [J].
Kamburoglu, Kivanc ;
Samunahmetoglu, Ercin ;
Eratam, Nejlan ;
Sonmez, Gul ;
Karahan, Sevilay .
IMAGING SCIENCE IN DENTISTRY, 2022,
[39]   Intraoral digital scans: Part 2-influence of ambient scanning light conditions on the mesh quality of different intraoral scanners [J].
Revilla-Leon, Marta ;
Jiang, Peng ;
Sadeghpour, Mehrad ;
Piedra-Cascon, Wenceslao ;
Zandinejad, Amirali ;
Oezcan, Mutlu ;
Krishnamurthy, Vinayak R. .
JOURNAL OF PROSTHETIC DENTISTRY, 2020, 124 (05) :575-580
[40]   Changes in views on digital intraoral scanners among dental hygienists after training in digital impression taking [J].
Park, Hye-Ran ;
Park, Ji-Man ;
Chun, Youn-Sic ;
Lee, Kkot-Nim ;
Kim, Minji .
BMC ORAL HEALTH, 2015, 15