Improvements in image quality after optimization in digital intraoral radiographs

被引:2
作者
Barrineau, Ashley B. [1 ]
Mah, Peter [2 ]
Mallya, Sanjay M. [3 ]
机构
[1] Univ Calif Los Angeles, Sch Dent, Sect Endodont, Los Angeles, CA USA
[2] Dent Imaging Consultants, Box 29214 Med Ctr Stn, San Antonio, TX 78229 USA
[3] Univ Calif Los Angeles, Sch Dent, Sect Maxillofacial Radiol, Los Angeles, CA USA
关键词
Quality control; quality assurance; image optimization; digital radiography; exposure optimization; radiographic phantom; image analysis; ASSURANCE; ACCURACY;
D O I
10.1016/j.adaj.2022.09.008
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Background. Digital intraoral radiographic exposures are optimized largely on the basis of subjective assessment of diagnostic image quality. This study presents an objective approach to optimize radiographic exposure settings for digital intraoral radiographic systems.Methods. Seven size 2 digital intraoral systems were assessed for image quality and determination of optimal exposure following the protocol specified in American National Standard Institute/American Dental Association Standard No. 1094: Quality Assurance for Digital Intra-Oral Radiographic Systems. A ProX radiograph unit (Planmeca) at 63 kVp and 6 mA was used to obtain radiographs of the Dental Digital Quality Assurance phantom. ImageJ software (National Institutes of Health) was used to quantify dynamic range and spatial resolution, and contrast perceptibility was evaluated visually. Optimal exposure is the setting with the maximal contrast perceptibility and spatial resolution while displaying the full dynamic range. After image optimization, a custom phantom consisting of an endodontically prepared tooth was imaged to evaluate the file position relative to the apex for each system. Differences in distances between file position relative to the root apex at the optimal exposure as well as 1 increment above and below were measured.Results. Radiographic images obtained at the optimal exposure yielded better visualization and more accurate measurements of the file tip relative to the apex.Conclusions. Optimizing radiographic exposures improves image quality and accuracy in clinical decisions.Practical Implications. Improvement in image quality and better accuracy in actual distance of the endodontic file to the radiographic apex coupled with complete cleaning, shaping, and obturation of the canal should lead to better endodontic treatment outcomes.
引用
收藏
页码:24 / 31
页数:8
相关论文
共 50 条
  • [21] Subjective image quality of digital panoramic radiographs displayed on monitor and printed on various hardcopy media.
    Gijbels F.
    Sanderink G.
    Pauwels H.
    Jacobs R.
    Clinical Oral Investigations, 2004, 8 (1) : 25 - 29
  • [22] Image quality of digital and film radiographs in applications sent to the Dental Insurance Office in Sweden for treatment approval
    Hellén-Halme, K
    Johansson, PM
    Håkansson, J
    Petersson, A
    SWEDISH DENTAL JOURNAL, 2004, 28 (02) : 77 - 84
  • [23] Digital Image Filtering Optimization Supporting Iberian Ham Quality Prediction
    Peran-Sanchez, Francisco
    Serrano, Salud
    Gutierrez de Rave, Eduardo
    Sanchez-Lopez, Elena
    Cumplido, Ana
    Jimenez-Hornero, Francisco J.
    FOODS, 2020, 9 (01)
  • [24] Reduction of scratch or dirt artifacts on intraoral radiographs using dual imaging plates in image processing
    Imanishi, Yusuke
    Sekiguchi, Takahito
    Kato, Masao
    Kimoto, Hideaki
    Amemiya, Toshihiko
    Dezawa, Ko
    Matsumoto, Kunihito
    Arai, Yoshinori
    ORAL RADIOLOGY, 2023, 39 (02) : 386 - 393
  • [25] The effects of image compression on quantitative measurements of digital panoramic radiographs
    Yasar, Fusun
    Apaydin, Burcu
    Yilmaz, Hasan-Huseyin
    MEDICINA ORAL PATOLOGIA ORAL Y CIRUGIA BUCAL, 2012, 17 (06): : E1074 - E1081
  • [26] Investigation of the Variability in the Assessment of Digital Chest X-ray Image Quality
    Whaley, Jacquelyn S.
    Pressman, Barry D.
    Wilson, Jonathan R.
    Bravo, Lionel
    Sehnert, William J.
    Foos, David H.
    JOURNAL OF DIGITAL IMAGING, 2013, 26 (02) : 217 - 226
  • [27] Investigation of the Variability in the Assessment of Digital Chest X-ray Image Quality
    Jacquelyn S. Whaley
    Barry D. Pressman
    Jonathan R. Wilson
    Lionel Bravo
    William J. Sehnert
    David H. Foos
    Journal of Digital Imaging, 2013, 26 : 217 - 226
  • [28] Image Quality Optimization vs. Patient Dose in Digital Thoracic Radiography
    Perez-Diaz, M.
    Ruiz-Gonzalez, Y.
    Machin-Linares, L.
    Ely-Andrade, M.
    Barros-Saito, J. C.
    Khoury, H. J.
    Ventosa-Almeida, J. C.
    VI LATIN AMERICAN CONGRESS ON BIOMEDICAL ENGINEERING (CLAIB 2014), 2014, 49 : 353 - 356
  • [29] In vitro efficacy of reverse contrast and sharpen enhancement filters for detection of small-size broken files in root canals on digital intraoral radiographs
    Jafarzadeh, Mansour
    Chetin, Negar
    Sharifishoshtari, Sanaz
    Haghighizadeh, Mohammad Hossein
    MEDICAL SCIENCE, 2020, 24 (106) : 4203 - 4210
  • [30] Influence of the Acquisition Parameters on the Image Quality in Digital Mammography
    Al Kattar, Zeina
    El Balaa, Hanna
    Harb, Huda
    Raad, Nisrine
    Hodroj, Noor
    Sleiman, Fatina
    Omayrat, Samia
    2015 INTERNATIONAL CONFERENCE ON ADVANCES IN BIOMEDICAL ENGINEERING (ICABME), 2015, : 33 - 36