Patient-specific quality assurance strategies for synthetic computed tomography in magnetic resonance-only radiotherapy of the abdomen

被引:8
|
作者
Dal Bello, Riccardo [1 ,2 ,5 ]
Lapaeva, Mariia [1 ,2 ,3 ,4 ]
La Greca Saint-Esteven, Agustina [1 ,2 ,4 ]
Wallimann, Philipp [1 ,2 ]
Gunther, Manuel [3 ]
Konukoglu, Ender [4 ]
Andratschke, Nicolaus [1 ,2 ]
Guckenberger, Matthias [1 ,2 ]
Tanadini-Lang, Stephanie [1 ,2 ]
机构
[1] Univ Hosp Zurich, Dept Radiat Oncol, Zurich, Switzerland
[2] Univ Zurich, Zurich, Switzerland
[3] Univ Zurich, Dept Informat, Artificial Intelligence & Machine Learning Grp, Zurich, Switzerland
[4] Swiss Fed Inst Technol, Comp Vis Lab, Zurich, Switzerland
[5] Univ Spital Zurich USZ, Klin Radioonkol, Ramistr 100, CH-8091 Zurich, Switzerland
来源
PHYSICS & IMAGING IN RADIATION ONCOLOGY | 2023年 / 27卷
关键词
MR-guided radiotherapy; synthetic CT; Neural network; Quality assurance; PSQA; MR-Linac; CT; GENERATION; MRI;
D O I
10.1016/j.phro.2023.100464
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background and purpose: The superior tissue contrast of magnetic resonance (MR) compared to computed tomography (CT) led to an increasing interest towards MR-only radiotherapy. For the latter, the dose calculation should be performed on a synthetic CT (sCT). Patient-specific quality assurance (PSQA) methods have not been established yet and this study aimed to assess several software-based solutions.Materials and methods: A retrospective study was performed on 20 patients treated at an MR-Linac, which were selected to evenly cover four subcategories: (i) standard, (ii) air pockets, (iii) lung and (iv) implant cases. The neural network (NN) CycleGAN was adopted to generate a reference sCT, which was then compared to four PSQA methods: (A) water override of body, (B) five tissue classes with bulk densities, (C) sCT generated by a separate NN (pix2pix) and (D) deformed CT.Results: The evaluation of the dose endpoints demonstrated that while all methods A-D provided statistically equivalent results (p = 0.05) within the 2% level for the standard cases (i), only the methods C-D guaranteed the same result over the whole cohort. The bulk densities override was shown to be a valuable method in absence of lung tissue within the beam path.Conclusion: The observations of this study suggested that the use of an additional sCT generated by a separate NN was an appropriate tool to perform PSQA of a sCT in an MR-only workflow at an MR-Linac. The time and dose endpoints requirements were respected, namely within 10 min and 2%.
引用
收藏
页数:7
相关论文
共 46 条
  • [31] Clinical performance of FractionLab in patient-specific quality assurance for intensity-modulated radiotherapy : a retrospective study
    Oh, Se An
    Kim, Sung Yeop
    Park, Jaehyeon
    Park, Jae Won
    Yea, Ji Woon
    JOURNAL OF YEUNGNAM MEDICAL SCIENCE, 2022, 39 (02): : 108 - 115
  • [32] Quality assurance issues for computed tomography-, ultrasound-, and magnetic resonance imaging-guided brachytherapy
    Cormack, Robert A.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 71 (01): : S136 - S141
  • [33] An Evolutionary Algorithm for Optimization of Affine Transformation Parameters for Dose Matrix Warping in Patient-Specific Quality Assurance of Radiotherapy Dose Distributions
    Witten, Matthew
    Clancey, Owen
    2012 IEEE CONGRESS ON EVOLUTIONARY COMPUTATION (CEC), 2012,
  • [34] Intermodality feature fusion combining unenhanced computed tomography and ferumoxytol-enhanced magnetic resonance angiography for patient-specific vascular mapping in renal impairment
    Yoshida, Takegawa
    Kim-Lien Nguyen
    Shahrouki, Puja
    Quinones-Baldrich, William J.
    Lawrence, Peter F.
    Finn, J. Paul
    JOURNAL OF VASCULAR SURGERY, 2020, 71 (05) : 1674 - 1684
  • [35] Implementing and evaluating a high-resolution diode array for patient-specific quality assurance of robotic brain stereotactic radiosurgery/radiotherapy
    Xu, Qianyi
    Huynh, Kiet
    Nie, Wei
    Rose, Mark S.
    Chawla, Ashish K.
    Choe, Kevin S.
    Kanani, Samir
    Kubicek, Gregory J.
    Fan, Jiajin
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2022, 23 (05):
  • [36] Patient-specific calibration of cone-beam computed tomography data sets for radiotherapy dose calculations and treatment plan assessment
    MacFarlane, Michael
    Wong, Daniel
    Hoover, Douglas A.
    Wong, Eugene
    Johnson, Carol
    Battista, Jerry J.
    Chen, Jeff Z.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2018, 19 (02): : 249 - 257
  • [37] Magnetic resonance-based synthetic computed tomography images generated using generative adversarial networks for nasopharyngeal carcinoma radiotherapy treatment planning
    Peng, Yinglin
    Chen, Shupeng
    Qin, An
    Chen, Meining
    Gao, Xingwang
    Liu, Yimei
    Miao, Jingjing
    Gu, Huikuan
    Zhao, Chong
    Deng, Xiaowu
    Qi, Zhenyu
    RADIOTHERAPY AND ONCOLOGY, 2020, 150 : 211 - 218
  • [38] Magnetic resonance imaging-based bone imaging of the lower limb: Strategies for generating high-resolution synthetic computed tomography
    Florkow, Mateusz C.
    Nguyen, Chien H.
    Sakkers, Ralph J. B.
    Weinans, Harrie
    Jansen, Mylene P.
    Custers, Roel J. H.
    van Stralen, Marijn
    Seevinck, Peter R.
    JOURNAL OF ORTHOPAEDIC RESEARCH, 2024, 42 (04) : 843 - 854
  • [39] Deep learning for patient-specific quality assurance: Identifying errors in radiotherapy delivery by radiomic analysis of gamma images with convolutional neural networks
    Nyflot, Matthew J.
    Thammasorn, Phawis
    Wootton, Landon S.
    Ford, Eric C.
    Chaovalitwongse, W. Art
    MEDICAL PHYSICS, 2019, 46 (02) : 456 - 464
  • [40] A novel dynamic robotic moving phantom system for patient-specific quality assurance in real-time tumor-tracking radiotherapy
    Shiinoki, Takehiro
    Fujii, Fumitake
    Fujimoto, Koya
    Yuasa, Yuki
    Sera, Tatsuhiro
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2020, 21 (07): : 16 - 28