Volumetric Modulated Arc Therapy Planning for Craniospinal Irradiation With a New O-ring Linac

被引:0
|
作者
Matsumoto, Kenji [1 ]
Monzen, Hajime [2 ]
Kubo, Kazuki [2 ]
Otsuka, Masakazu [1 ]
Nambu, Hidekazu [1 ]
Nishimura, Yasumasa [3 ]
机构
[1] Kindai Univ Hosp, Dept Radiol Serv, Osakasayama, Japan
[2] Kindai Univ, Grad Sch Med Sci, Dept Med Phys, Osaka, Japan
[3] Kindai Univ, Fac Med, Osakasayama, Japan
关键词
auto-feathering; halcyon; rapidarc; csi; supine craniospinal irradiation; RADIOTHERAPY; IMRT;
D O I
10.7759/cureus.36493
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
This study aims to determine the feasibility of using a new O-ring linear accelerator (Halcyon, Varian Medical Systems, CA, USA) to perform treatment planning using volumetric modulated arc therapy (VMAT) for craniospinal irradiation (CSI).A 20-year-old male patient with leukemia was selected. The planning target volume (PTV) was contoured to include the entire contents of the brain and spinal canal. The PTV margin was 10 mm applied to the clinical target volume (CTV). VMAT (RapidArc, Varian Medical Systems, CA, USA) planning was performed using four isocenter with five arcs, two full rotation arcs to cover the brain and upper part of the spinal cord, and one full rotation arc for the lower part of the spinal cord. The plan was created using the auto-feathering photon optimizer calculation of the planning system. The conformity index (CI) and heterogeneity index (HI) as well as dose-volume histograms of organs at risk (OAR) were evaluated. The patient position of +/- 3.0 mm in the craniocaudal direction was moved in to simulate the effect of treatment inaccuracy. The total treatment time was also measured.The CI and HI were 1.09 and 8.44, respectively. The mean dose (PTV) was 105.5%, and the mean dose (OARs) was lower than the planning dose constraints. Simulations with a patient position shift of +/- 3.0 mm resulted in an error of less than +/- 10.0% of the planned dose to the spinal cord. The total treatment time was within 15 minutes.VMAT planning for CSI with Halcyon achieved high conformality, uniform dose distribution, low dose to the surrounding normal tissues, and reduced treatment time.
引用
收藏
页数:8
相关论文
共 50 条
  • [1] Volumetric modulated arc therapy planning method for supine craniospinal irradiation
    Chen, Jianzhou
    Chen, Chuangzhen
    Atwood, Todd F.
    Gibbs, Iris C.
    Soltys, Scott G.
    Fasola, Carolina
    Xing, Lei
    JOURNAL OF RADIATION ONCOLOGY, 2012, 1 (03) : 291 - 297
  • [2] Automating the Treatment Planning Process for Volumetric Modulated Arc Therapy Craniospinal Irradiation
    Romero, Ignacio O.
    Simiele, Eric A.
    Lozko, Yuliia
    Severyn, Yuliia
    Skinner, Lawrie B.
    Yang, Yong
    Wang, Jen-Yeu
    Xing, Lei
    Gibbs, Iris
    Hiniker, Susan M.
    Kovalchuk, Nataliya
    PRACTICAL RADIATION ONCOLOGY, 2024, 14 (02) : e165 - e172
  • [3] Assessment of a conventional volumetric-modulated arc therapy knowledge-based planning model applied to the new Halcyon© O-ring linac in locoregional breast cancer radiotherapy
    Costa, Emilie
    Richir, Thomas
    Robilliard, Magalie
    Bragard, Christel
    Logerot, Christelle
    Kirova, Youlia
    Fourquet, Alain
    De Marzi, Ludovic
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2021, 86 : 32 - 43
  • [4] Volumetric modulated arc therapy of head-and-neck cancer on a fast-rotating O-ring linac: Plan quality and delivery time comparison with a C-arm linac
    Michiels, Steven
    Poels, Kenneth
    Crijns, Wouter
    Delombaerde, Laurence
    De Roover, Robin
    Vanstraelen, Bianca
    Haustermans, Karin
    Nuyts, Sandra
    Depuydt, Tom
    RADIOTHERAPY AND ONCOLOGY, 2018, 128 (03) : 479 - 484
  • [5] Impact of setup errors on multi-isocenter volumetric modulated arc therapy for craniospinal irradiation
    Zhou, Yongqiang
    Ai, Yao
    Han, Ce
    Zheng, Xiaomin
    Yi, Jinling
    Xie, Congying
    Jin, Xiance
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2020, 21 (11): : 115 - 123
  • [6] Intensity-modulated radiation therapy and volumetric-modulated arc therapy for adult craniospinal irradiation-A comparison with traditional techniques
    Studenski, Matthew T.
    Shen, Xinglei
    Yu, Yan
    Xiao, Ying
    Shi, Wenyin
    Biswas, Tithi
    Werner-Wasik, Maria
    Harrison, Amy S.
    MEDICAL DOSIMETRY, 2013, 38 (01) : 48 - 54
  • [7] Treatment planning for volumetric modulated arc therapy
    Bedford, James L.
    MEDICAL PHYSICS, 2009, 36 (11) : 5128 - 5138
  • [8] Acute toxicity of craniospinal irradiation with volumetric-modulated arc therapy in children with solid tumors
    Wong, Kenneth K.
    Ragab, Omar
    Tran, Hung N.
    Pham, Anthony
    All, Sean
    Waxer, Jonathan
    Olch, Arthur J.
    PEDIATRIC BLOOD & CANCER, 2018, 65 (07)
  • [9] Feasibility study of volumetric modulated arc therapy with Halcyon™ linac for total body irradiation
    Uehara, Takuya
    Monzen, Hajime
    Tamura, Mikoto
    Inada, Masahiro
    Otsuka, Masakazu
    Doi, Hiroshi
    Matsumoto, Kenji
    Nishimura, Yasumasa
    RADIATION ONCOLOGY, 2021, 16 (01)
  • [10] Feasibility study of volumetric modulated arc therapy with Halcyon™ linac for total body irradiation
    Takuya Uehara
    Hajime Monzen
    Mikoto Tamura
    Masahiro Inada
    Masakazu Otsuka
    Hiroshi Doi
    Kenji Matsumoto
    Yasumasa Nishimura
    Radiation Oncology, 16