Design and Testing of a Smart Facemask for Respiratory Monitoring during Cycling Exercise

被引:12
作者
Romano, Chiara [1 ]
Nicolo, Andrea [2 ]
Innocenti, Lorenzo [2 ]
Sacchetti, Massimo [2 ]
Schena, Emiliano [1 ]
Massaroni, Carlo [1 ]
机构
[1] Univ Campus Biomed Roma, Dept Fac Engn, I-00128 Rome, Italy
[2] Univ Rome Foro Italico, Dept Movement Human & Hlth Sci, I-00135 Rome, Italy
来源
BIOSENSORS-BASEL | 2023年 / 13卷 / 03期
关键词
wearable sensors; validity; respiratory frequency; cadence; measurement accuracy; exercise; sport; breathing; respiratory rate; FREQUENCY; RESPONSES; RELIABILITY; VALIDATION; VALIDITY; DYNAMICS;
D O I
10.3390/bios13030369
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Given the importance of respiratory frequency (f(R)) as a valid marker of physical effort, there is a growing interest in developing wearable devices measuring f(R) in applied exercise settings. Biosensors measuring chest wall movements are attracting attention as they can be integrated into textiles, but their susceptibility to motion artefacts may limit their use in some sporting activities. Hence, there is a need to exploit sensors with signals minimally affected by motion artefacts. We present the design and testing of a smart facemask embedding a temperature biosensor for f(R) monitoring during cycling exercise. After laboratory bench tests, the proposed solution was tested on cyclists during a ramp incremental frequency test (RIFT) and high-intensity interval training (HIIT), both indoors and outdoors. A reference flowmeter was used to validate the f(R) extracted from the temperature respiratory signal. The smart facemask showed good performance, both at a breath-by-breath level (MAPE = 2.56% and 1.64% during RIFT and HIIT, respectively) and on 30 s average f(R) values (MAPE = 0.37% and 0.23% during RIFT and HIIT, respectively). Both accuracy and precision (MOD +/- LOAs) were generally superior to those of other devices validated during exercise. These findings have important implications for exercise testing and management in different populations.
引用
收藏
页数:20
相关论文
共 49 条
[1]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[2]  
Borg G., 1998, BORGS PERCEIVED EXER
[3]  
De Tommasi F., 2022, IEEE INT SYM MED MEA
[4]  
DERY R., 1967, CAN ANAESTH SOC J, V14, P287, DOI 10.1007/BF03003698
[5]   VALIDITY AND RELIABILITY OF THE HEXOSKIN WEARABLE BIOMETRIC VEST DURING MAXIMAL AEROBIC POWER TESTING IN ELITE CYCLISTS [J].
Elliot, Catherine A. ;
Hamlin, Michael J. ;
Lizamore, Catherine A. .
JOURNAL OF STRENGTH AND CONDITIONING RESEARCH, 2019, 33 (05) :1437-1444
[6]  
Fabre N, 2007, J SPORT SCI MED, V6, P526
[7]   Neuro-mechanical and chemical influences on locomotor respiratory coupling in humans [J].
Fabre, Nicolas ;
Perrey, Stephane ;
Arbez, Lofc ;
Rouillon, Jean-Denis .
RESPIRATORY PHYSIOLOGY & NEUROBIOLOGY, 2007, 155 (02) :128-136
[8]   The effect of pedalling cadence on respiratory frequency: passive vs. active exercise of different intensities [J].
Girardi, Michele ;
Nicolo, Andrea ;
Bazzucchi, Ilenia ;
Felici, Francesco ;
Sacchetti, Massimo .
EUROPEAN JOURNAL OF APPLIED PHYSIOLOGY, 2021, 121 (02) :583-596
[9]   Reliability and Validity of the ZephyrTM (TM) BioHarness (TM) to Measure Respiratory Responses to Exercise [J].
Hailstone, Jono ;
Kilding, Andrew .
MEASUREMENT IN PHYSICAL EDUCATION AND EXERCISE SCIENCE, 2011, 15 (04) :293-300
[10]   Breath Tools: A Synthesis of Evidence-Based Breathing Strategies to Enhance Human Running [J].
Harbour, Eric ;
Stoeggl, Thomas ;
Schwameder, Hermann ;
Finkenzeller, Thomas .
FRONTIERS IN PHYSIOLOGY, 2022, 13