Additive manufactured gyroid-based cell structures under compression: design, testing and simulation for biomedical applications

被引:2
|
作者
Carlos Castro-Sandoval, Juan [1 ]
Chavez, Alfredo [1 ]
Corona-Castuera, Jorge [1 ]
Henao, John [1 ]
Antonio Rivera-Gil, Marco [1 ]
Poblano-Salas, Carlos A. [1 ]
机构
[1] CIATEQ, Adv Mat, AC, Queretaro, Mexico
关键词
Additive manufacturing; gyroid-based cell structures; compression testing; finite element method; BONE; OPTIMIZATION; PERFORMANCE; LATTICES; SCAFFOLD;
D O I
10.1080/10255842.2023.2179366
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
The mechanical behaviour of a DMLS Ti-6Al-4V gyroid-based cellular structure (CS), with potential application in the fabrication of implants, was studied under compressive conditions. The influence of the CS volumetric fraction on the elastic modulus was experimentally evaluated in cubic and cylindrical samples. The experimental results showed that the selected parameters allowed approximating the mechanical behaviour of the CS to that of trabecular bone. Finite element analysis was employed to study the mechanical behaviour of the CS. The model presented a good approximation of the experimental results, being useful to predict the mechanical behaviour of the CS.
引用
收藏
页码:211 / 221
页数:11
相关论文
共 48 条
  • [41] Cold-Formed Steel Equal Angles with Complex Edge Stiffeners under Axial-Compression Laboratory Testing, Numerical Simulation, and Design
    Aruna, G.
    Roy, Krishanu
    Lim, James B. P.
    Kumar, P. Suresh
    PRACTICE PERIODICAL ON STRUCTURAL DESIGN AND CONSTRUCTION, 2022, 27 (02)
  • [42] Design and Simulation of Proposed Low Cost Solar Cell Structures Based on Heavily Doped Silicon Wafers
    Salem, M. S.
    Zekry, A.
    Shaker, A.
    Abouelatta, M.
    2016 IEEE 43RD PHOTOVOLTAIC SPECIALISTS CONFERENCE (PVSC), 2016, : 2393 - 2397
  • [43] Testing, simulation and design of back-to-back built-up cold-formed steel unequal angle sections under axial compression
    Ananthi, G. Beulah Gnana
    Roy, Krishanu
    Chen, Boshan
    Lim, James Bp
    STEEL AND COMPOSITE STRUCTURES, 2019, 33 (04): : 595 - 614
  • [44] Effective design and simulation of surface-based lattice structures featuring volume fraction and cell type grading
    Maskery, I.
    Aremu, A. O.
    Parry, L.
    Wildman, R. D.
    Tuck, C. J.
    Ashcroft, I. A.
    MATERIALS & DESIGN, 2018, 155 : 220 - 232
  • [45] Correlation between microstructures and mechanical properties under tensile and compression tests of heat-treated Ti-6Al-4 V ELI alloy produced by additive manufacturing for biomedical applications
    Longhitano, Guilherme Arthur
    Larosa, Maria Aparecida
    Jardini, Andre Luiz
    de Carvalho Zavaglia, Cecilia Amelia
    Filippini Ierardi, Maria Clara
    JOURNAL OF MATERIALS PROCESSING TECHNOLOGY, 2018, 252 : 202 - 210
  • [46] Nonlinear modeling, strength-based design, and testing of flexible piezoelectric energy harvesters under large dynamic loads for rotorcraft applications
    Leadenham, Stephen
    Erturk, Alper
    ACTIVE AND PASSIVE SMART STRUCTURES AND INTEGRATED SYSTEMS 2014, 2014, 9057
  • [47] Effective design and simulation of surface-based lattice structures featuring volume fraction and cell type grading (vol 155, pg 220, 2018)
    Maskery, I.
    Aremu, A. O.
    Parry, L.
    Wildman, R. D.
    Tuck, C. J.
    Ashcroft, I. A.
    MATERIALS & DESIGN, 2018, 160 : 106 - 107
  • [48] The design of a novel delay cell based 8.3 GHz, low phase noise ring oscillator in C-MOS 180 nm technology for biomedical ultra-wide-band integrated applications
    Raman, Ashish
    Sarin, Rakesh Kumar
    INTERNATIONAL JOURNAL OF BIOMEDICAL ENGINEERING AND TECHNOLOGY, 2013, 13 (01) : 1 - 16