Remote Monitoring of Visual Function in Patients with Maculopathy: The Aphelion Study

被引:2
|
作者
Joseph, Anthony [1 ]
Bullimore, Mark [2 ]
Drawnel, Faye [3 ]
Miranda, Marco [4 ,7 ]
Morgan, Zoe [3 ]
Wang, Yi-Zhong [5 ,6 ]
机构
[1] Retina Associates, Tucson, AZ USA
[2] Univ Houston, Coll Optometry, Houston, TX USA
[3] F Hoffmann La Roche Pharmaceut AG, Basel, Switzerland
[4] Roche Prod Ltd, Welwyn Garden City, England
[5] Retina Fdn Southwest, 9600 N Cent Expressway,Suite 200, Dallas, TX 75321 USA
[6] UT SouthWestern Med Ctr, Dept Ophthalmol, Dallas, TX USA
[7] UCL, Inst Ophthalmol, London, England
关键词
Contrast sensitivity; Hyperacuity; Maculopathy; Remote monitoring; Visual acuity; Smartphone application; COMPACT REDUCED LOGMAR; MACULAR DEGENERATION; SHAPE-DISCRIMINATION; CONTRAST SENSITIVITY; ACUITY CHART; HYPERACUITY; PROGRESSION; VALIDATION; AGREEMENT; VALIDITY;
D O I
10.1007/s40123-023-00854-2
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
IntroductionRemote monitoring of vision, using tools such as the shape discrimination hyperacuity (SDH) test, can detect disease activity in patients with maculopathy. We determined the in-clinic accuracy and repeatability of three myVisionTrack expanded version (mVTx) tests for self-testing of visual acuity (VA) and contrast sensitivity.MethodsAphelion, a single-arm, prospective study conducted at two sites in the USA, included adults with any maculopathy and a baseline VA of 0.7 log of minimum angle of resolution (logMAR) (Snellen 20/100) or better. Participants completed the mVTx tests (tumbling E, Landolt C, contrast sensitivity, and SDH) and standard clinical tests (near and distance Early Treatment Diabetic Retinopathy Study [ETDRS] charts and the Pelli-Robson contrast sensitivity chart). Test-retest repeatability and agreement between the mVTx tests and the corresponding clinical test were assessed by Bland-Altman analyses. Participants also completed a usability survey.ResultsThe mean age of the 122 participants was 67 years. The most common diagnosis was age-related macular degeneration (42% of patients). The tumbling E test had a test-retest 95% limit of agreement (LoA) of +/- 0.18 logMAR; the Landolt C test, +/- 0.23 logMAR; the SDH test, +/- 0.24 logMAR; and the contrast sensitivity test, +/- 0.32 log contrast threshold (logCT). Compared with the distance ETDRS chart, the LoA was +/- 0.35 logMAR for the tumbling E test (mean difference, - 0.07 logMAR) and +/- 0.39 logMAR for the Landolt C test (mean difference, 0.03 logMAR). For the contrast sensitivity test, the LoA compared with the Pelli-Robson chart was +/- 0.30 logCT (mean difference, - 0.25 logCT). Most participants (85%) reported that they learned the tests quickly. The tumbling E test scored the highest on ease of use.ConclusionThe mVTx tests of VA are accurate and repeatable, supporting their potential use alongside the SDH test to detect disease progression remotely between clinic visits.
引用
收藏
页码:409 / 422
页数:14
相关论文
共 50 条
  • [31] Choroidal Thickness and Visual Acuity in High Myopia Without Myopic Maculopathy: Insights From a Chinese Population Study
    Wang, Yueye
    Wang, Decai
    Yin, Qiuxia
    Li, Jiayong
    Li, Zhixi
    He, Mingguang
    TRANSLATIONAL VISION SCIENCE & TECHNOLOGY, 2024, 13 (11):
  • [32] Maculopathy and visual acuity in newly diagnosed type 2 diabetic patients and non-diabetic subjects:: A 10-year follow-up study
    Voutilainen-Kaunisto, R
    Teräsvirta, M
    Uusitupa, M
    Niskanen, L
    ACTA OPHTHALMOLOGICA SCANDINAVICA, 2001, 79 (02): : 163 - 168
  • [33] Home vision monitoring in patients with maculopathy: current and future options for digital technologies
    Balaskas, Konstantinos
    Drawnel, Faye
    Khanani, Arshad M.
    Knox, Paul C.
    Mavromaras, George
    Wang, Yi-Zhong
    EYE, 2023, 37 (15) : 3108 - 3120
  • [34] VISUAL FUNCTION AND RETINAL CHANGES IN PATIENTS WITH BIPOLAR DISORDER
    Garcia-Martin, Elena
    Gavin, Alicia
    Garcia-Campayo, Javier
    Vilades, Elisa
    Orduna, Elvira
    Polo, Vicente
    Larrosa, Jose M.
    Pablo, Luis E.
    Satue, Maria
    RETINA-THE JOURNAL OF RETINAL AND VITREOUS DISEASES, 2019, 39 (10): : 2012 - 2021
  • [35] A study of robot remote monitoring system
    Wang, D
    Chen, YM
    Fang, ML
    He, YY
    PROCEEDINGS OF THE 4TH WORLD CONGRESS ON INTELLIGENT CONTROL AND AUTOMATION, VOLS 1-4, 2002, : 1291 - 1295
  • [36] WEARABLE MULTIPARAMETRIC DEVICE FOR REMOTE MONITORING OF CARDIORESPIRATORY FUNCTION
    Spadotto, Alberto
    Moscato, Serena
    Massaro, Giulia
    Spagni, Stefano
    Chiari, Lorenzo
    Diemberger, Igor
    JOURNAL OF MECHANICS IN MEDICINE AND BIOLOGY, 2023, 23 (06)
  • [37] Continuous Remote Monitoring in Moderate and Severe COVID-19 Patients
    Rajanna, Avinash H.
    Bellary, Vaibhav S.
    Puranic, Sohani Kashi
    Nayana, C.
    Nagaraj, Jatin Raaghava
    Eshanye, D. A.
    Preethi, K.
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2023, 15 (09)
  • [38] On the Remote Monitoring and Controlling Function of Short Message Service
    Gao, Jinglian
    RECENT ADVANCES OF ASIAN LANGUAGE PROCESSING TECHNOLOGIES, 2008, : 223 - 226
  • [39] VALUE OF SCANNING LASER OPHTHALMOSCOPY IN THE EVALUATION OF THE VISUAL FUNCTION OF 47 PATIENTS WITH MODERATE CATARACTS ASSOCIATED WITH MACULOPATHY .1. VALUE OF SCANNING LASER OPHTHALMOSCOPY IN THE EVALUATION OF OPTOTYPE READING CAPACITIES
    LEGARGASSON, JF
    RIGAUDIERE, F
    GUEZ, JE
    SCHMITT, D
    GRALL, Y
    CLINICAL VISION SCIENCES, 1992, 7 (06): : 531 - &
  • [40] Visual function in patients with choroidal neovascularization resulting from age-related macular degeneration: The importance of looking beyond visual acuity